Use of single photon counting detector arrays in combined PET/MR: Characterization of LYSO-SiPM detector modules and comparison with a LSO-APD detector

被引:44
|
作者
Spanoudaki, V. C. [1 ,2 ]
Mann, A. B. [2 ]
Otte, A. N. [3 ]
Konorov, I. [2 ]
Torres-Espallardo, I. [1 ]
Paul, S. [2 ]
Ziegler, S. I. [1 ]
机构
[1] Tech Univ Munich, Klinikum Rechts Isar, Nukl Med Klin & Poliklin, D-81675 Munich, Germany
[2] Tech Univ Munich, Dept Phys, Lehrstuhl Expt Phys E18, D-85748 Munich, Germany
[3] Humboldt Univ, D-12489 Berlin, Germany
来源
关键词
gamma camera; SPECT; PET PET/CT; coronary CT angiography (CTA); multi-modality systems;
D O I
10.1088/1748-0221/2/12/P12002
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We propose in this study a novel PET detector concept as insert for simultaneous PET/MR imaging, using arrays of Silicon Photomultipliers (SiPMs) as photodetectors, read out by a data acquisition system based on sampling ADCs. A 2 x 2 LSO-SiPM detector array and four single channel LYSO-SiPM detectors have been evaluated and compared to a LSO-APD detector. A 17.9% energy resolution and a 1.4 ns time resolution have been measured. No degradation of these values could be detected when simultaneous MR acquisitions were performed. The nonlinear detector behaviour due to the limited dynamic range and recovery time effects has been studied. In addition, the contribution of dark counts and optical crosstalk for PET applications was also addressed. The feasibility for position localization of the incident light to a SiPM array using Anger logic has been investigated.
引用
收藏
页数:17
相关论文
共 3 条
  • [1] Validated simulation for LYSO:Ce scintillator based PET detector modules built on fully digital SiPM arrays
    Jatekos, B.
    Gasparini, L.
    Lorincz, E.
    Erdei, G.
    [J]. JOURNAL OF INSTRUMENTATION, 2016, 11
  • [2] A prototype detector module for combined PET/CT or combined photon counting/standard CT based on SiPM technology
    Persson, Andreas
    Khaplanov, Anton
    Cederwall, Bo
    Bohm, Christian
    [J]. 2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3503 - +
  • [3] Experimental characterization of novel small animal PET detector modules based on scintillation crystal blocks read out by APD arrays
    Maas, MC
    van der Laan, DJ
    Schaart, DR
    Huizenga, J
    Brouwer, JC
    Bruyndonckx, P
    Léonard, S
    Lemaître, C
    van Eijk, CWE
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2942 - 2946