Prostate Image-Guided Radiotherapy by Megavolt Cone-Beam CT

被引:1
|
作者
Zucca, Sergio
Carau, Barbara
Solla, Ignazio
Garibaldi, Elisabetta
Farace, Paolo [1 ]
Lay, Giancarlo
Meleddu, Gianfranco
Gabriele, Pietro [1 ]
机构
[1] Reg Oncol Hosp, Dept Radiooncol, I-09121 Cagliari, Italy
关键词
Megavolt cone-beam CT; IGRT; Prostate; Radiotherapy; INTENSITY-MODULATED RADIOTHERAPY; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; FIDUCIAL MARKERS; GOLD MARKERS; BONY ANATOMY; LOCALIZATION; ONLINE; CANCER; SETUP;
D O I
10.1007/s00066-011-2241-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To test megavolt cone-beam CT (MV-CBCT) in order to evaluate setup errors in prostate radiotherapy. Patients and Methods: The setup of 9 patients was verified weekly by electronic portal imaging (EPI) and MV-CBCT, both performed in the same treatment session. EPI were compared with digitally reconstructed radiographies (DRRs). MV-CBCTs were matched to simulation CTs by manual registration based on bone markers (BMR), by manual registration based on soft tissues (STR) - rectum, bladder, and seminal vesicles - and by automatic registration (AR) performed by a mutual information algorithm. Shifts were evaluated along the three main axes: anteroposterior (AP), craniocaudal (CC), and laterolateral (LL). Finally, in 4 additional patients showing intraprostatic calcifications, the calcification mismatch error was used to evaluate the three MV-CBCT matching methods. Results: A total of 50 pairs of orthogonal EPIs and 50 MV-CBCTs were analyzed. Assuming an overall tolerance of 2 mm, no significant differences were observed comparing EPI vs BMR in any axis. A significant difference (p<0.001) was observed along the AP axis comparing EPI vs AR and EPI vs STR. On the calcification data set (22 measures), the calcification mismatch along the AP direction was significantly lower (p<0.05) after STR than after BMR or AR. Conclusion: Bone markers were not an effective surrogate of the target position and significant differences were observed comparing EPI or BMR vs STR, supporting the assessment of soft tissue position by MVCBs to verify and correct patient setup in prostate radiotherapy.
引用
收藏
页码:473 / 478
页数:6
相关论文
共 50 条
  • [31] Translational and rotational localization errors in cone-beam CT based image-guided lung stereotactic radiotherapy
    Garibaldi, Cristina
    Piperno, Gaia
    Ferrari, Annamaria
    Surgo, Alessia
    Muto, Matteo
    Ronchi, Sara
    Bazani, Alessia
    Pansini, Floriana
    Cremonesi, Marta
    Jereczek-Fossa, Barbara Alicja
    Orecchia, Roberto
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (07): : 859 - 865
  • [32] Clinical introduction of image-guided radiotherapy, IGRT with cone beam CT
    Sorcini, B
    Odh, R
    Näfstadius, P
    Näslund, I
    Wersäll, P
    Tilikidis, A
    Severin, AC
    Isoz, A
    Fjällman, P
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S103 - S103
  • [33] Fast kilovoltage/megavoltage (kVMV) breathhold cone-beam CT for image-guided radiotherapy of lung cancer
    Wertz, Hansjoerg
    Stsepankou, Dzmitry
    Blessing, Manuel
    Rossi, Michael
    Knox, Chris
    Brown, Kevin
    Gros, Uwe
    Boda-Heggemann, Judit
    Walter, Cornelia
    Hesser, Juergen
    Lohr, Frank
    Wenz, Frederik
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (15): : 4203 - 4217
  • [34] Theoretical aspects of implementation of kilovoltage cone-beam CT onboard linear accelerator for image-guided radiotherapy
    Rodriguez Cordon, Marta
    Ferrer Albiach, Carlos
    CLINICAL & TRANSLATIONAL ONCOLOGY, 2009, 11 (08): : 511 - 517
  • [35] The influence of a diet on prostate movement and automatic prostate localization on cone-beam CT scans for image guided radiotherapy
    Pos, F.
    Smitsmans, M.
    de Bois, J.
    Heemsbergen, W.
    Jaffray, D.
    Lebesque, J.
    van Herk, M.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S51 - S51
  • [36] The influence of a diet on prostate movement and automatic prostate localization on cone-beam CT scans for image guided radiotherapy
    Smitsmans, M. P. H.
    Pos, F. J.
    de Bois, J.
    Heemsbergen, W.
    Sonke, J. J.
    Jaffray, D. A.
    Lebesque, J. V.
    van Herk, M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S367 - S367
  • [37] Investigation of uncertainties in image registration of cone beam CT to CT on an image-guided radiotherapy system
    Sykes, J. R.
    Brettle, D. S.
    Magee, D. R.
    Thwaites, D. I.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (24): : 7263 - 7283
  • [38] Automatic prostate localization for high precision image-guided radiotherapy (IGRT) - First results using cone-beam CT (CBCT) scans
    Smitsmans, MHP
    De Bois, J
    Sonke, JJ
    Jaffray, DA
    Lebesque, JV
    Van Herk, M
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S50 - S51
  • [39] Intraoperative cone-beam CT for image-guided tibial plateau fracture reduction
    Khoury, A.
    Siewerdsen, J. H.
    Whyne, C. M.
    Daly, M. J.
    Kreder, H. J.
    Moseley, D. J.
    Jaffray, D. A.
    COMPUTER AIDED SURGERY, 2007, 12 (04) : 195 - 207
  • [40] Comparison of implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate
    Moseley, DJ
    White, E
    Haycocks, T
    Sharpe, M
    Siewerdsen, J
    Bissonnette, J
    Alasti, H
    Gospodarowicz, M
    Warde, P
    Catton, C
    Jaffray, D
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S330 - S331