Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems

被引:4
|
作者
Hariri, Sanaz [1 ]
Shahriari, Majid [2 ]
机构
[1] Shahid Beheshti Univ, Dept Radiat Med Engn, Tehran, Iran
[2] Shahid Beheshti Univ, Dept Radiat Applicat, Tehran, Iran
来源
关键词
Monte Carlo; MCNP4C; multileaf collimator (MLC); intensity-modulated radiation therapy (IMRT); LINEAR-ACCELERATOR; MODEL; TONGUE; BEAM;
D O I
10.1120/jacmp.v11i3.3101
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Due to intensive use of multileaf collimators (MLCs) in clinics, finding an optimum design for the leaves becomes essential. There are several studies which deal with comparison of MLC systems, but there is no article with a focus on offering an optimum design using accurate methods like Monte Carlo. In this study, we describe some characteristics of MLC systems including the leaf tip transmission, beam hardening, leakage radiation and penumbra width for Varian and Elekta 80-leaf MLCs using MCNP4C code. The complex geometry of leaves in these two common MLC systems was simulated. It was assumed that all of the MLC systems were mounted on a Varian accelerator and with a similar thickness as Varian's and the same distance from the source. Considering the obtained results from Varian and Elekta leaf designs, an optimum design was suggested combining the advantages of three common MLC systems and the simulation results of this proposed one were compared with the Varian and the Elekta. The leakage from suggested design is 29.7% and 31.5% of the Varian and Elekta MLCs. In addition, other calculated parameters of the proposed MLC leaf design were better than those two commercial ones. Although it shows a wider penumbra in comparison with Varian and Elekta MLCs, taking into account the curved motion path of the leaves, providing a double focusing design will solve the problem. The suggested leaf design is a combination of advantages from three common vendors (Varian, Elekta and Siemens) which can show better results than each one. Using the results of this theoretical study may bring about superior practical outcomes.
引用
收藏
页码:173 / 185
页数:13
相关论文
共 50 条
  • [1] Monte Carlo Simulation of a Micro-multileaf Collimator
    Gonzalez, W.
    Lallena, A. M.
    Alfonso, R.
    [J]. 5TH LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2011): SUSTAINABLE TECHNOLOGIES FOR THE HEALTH OF ALL, PTS 1 AND 2, 2013, 33 (1-2): : 1252 - +
  • [2] Design of Diagnostic Multileaf Collimator Based On the Monte Carlo Simulation for Dose Reduction of Diagnostic Radiography
    Han, S.
    Ji, Y.
    Kim, K.
    Choi, S.
    Jung, H.
    Park, S.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3405 - 3405
  • [3] Monte Carlo Simulation and Development of a Multileaf Collimator for Proton Therapy
    Ainsley, C.
    Scheuermann, R.
    Avery, S.
    Dolney, D.
    Maughan, R.
    McDonough, J.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [4] Monte Carlo simulation based study of a proposed multileaf collimator for a telecobalt machine
    Sahani, G.
    Sharma, Sunil Dutt
    Sharma, P. K. Dash
    Sharma, D. N.
    Hussain, S. A.
    [J]. MEDICAL PHYSICS, 2013, 40 (02)
  • [5] Monte Carlo simulations of the dosimetric characteristics of a new multileaf collimator
    Tacke, M
    Szymanowski, H
    Schulze, C
    Nuss, S
    Wehrwein, E
    Leidenberger, S
    Oelfke, U
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2018 - 2018
  • [6] Assessment of a new multileaf collimator concept using GEANT4 Monte Carlo simulations
    Tacke, MB
    Szymanowski, H
    Oelfke, U
    Schulze, C
    Nuss, S
    Wehrwein, E
    Leidenberger, S
    [J]. MEDICAL PHYSICS, 2006, 33 (04) : 1125 - 1132
  • [7] Monte Carlo simulation of the dynamic micro-multileaf collimator of a LINAC Elekta Precise using PENELOPE
    Gonzalez, W.
    Lallena, A. M.
    Alfonso, R.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (11): : 3417 - 3431
  • [8] Study on the tongue and groove effect of the Elekta multileaf collimator using Monte Carlo simulation and film dosimetry
    Haryanto, F
    Fippel, M
    Bakai, A
    Nüsslin, F
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2004, 180 (01) : 57 - 61
  • [9] Determining PGAA collimator plug design using Monte Carlo simulation
    Jalil, A.
    Chetaine, A.
    Amsil, H.
    Embarch, K.
    Benchrif, A.
    Laraki, K.
    Marah, H.
    [J]. NUCLEAR ENGINEERING AND TECHNOLOGY, 2021, 53 (03) : 942 - 948
  • [10] Collimator design for neutron radiography station using Monte Carlo simulation
    Wonglee, S.
    Liamsuwan, T.
    Pornroongruengchok, W.
    Channuie, J.
    Picha, R.
    Khaweerat, S.
    [J]. SIAM PHYSICS CONGRESS 2018 (SPC2018): A CREATIVE PATH TO SUSTAINABLE INNOVATION, 2018, 1144