Monte Carlo simulation of the dynamic micro-multileaf collimator of a LINAC Elekta Precise using PENELOPE

被引:7
|
作者
Gonzalez, W. [1 ]
Lallena, A. M. [1 ]
Alfonso, R. [2 ]
机构
[1] Univ Granada, Dept Fis Atom Mol & Nucl, E-18071 Granada, Spain
[2] INOR, Dept Radioterapia, Havana, Cuba
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 11期
关键词
STEREOTACTIC RADIOSURGERY; OUTPUT FACTORS; ELECTRON-BEAM; FOCAL SPOT; PHOTON; MODEL; MLC; ACCELERATOR; DOSIMETRY; QUALITY;
D O I
10.1088/0031-9155/56/11/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Micro-multileaf collimators are devices that are added to LINAC heads for stereotactic radiosurgery. In this work, the performance of an Elekta Precise LINAC with a dynamic micro-multileaf collimator manufactured by 3D-line has been studied. Monte Carlo simulations based on PENELOPE code and measurements with three different detectors (PTW Semiflex 31010 chamber, PTW PinPoint 31016 chamber and PTW Diode 60008) have been carried out. Simulations were tuned by reproducing the experimental TPR20,10 quality index, providing a nice description of both the PDD curve and the transverse profiles at the two depths measured. The geometry of the micro-multileaf collimator was tested by calculating the transmission through it, and it was needed to significantly reduce the leaf separation indicated by the manufacturer to reproduce the experimental results. An approximate simulation in which the transport of the particles traversing the dynamic micro-multileaf collimator was described in a simplified way was analyzed, providing good agreement with the full simulations. With the MC model fixed, output factors for various field sizes were calculated and compared to the experimental ones, obtaining good agreement. Percentage depth doses (PDDs) and transverse profiles at two depths measured with the diode for small fields were well reproduced by the simulation, while the measurements performed with the PinPoint chamber showed differences in the PDDs, at large depths, and transverse profiles, at the penumbra. Monte Carlo simulations and Semiflex and diode measurements, performed for a 7.0 cm x 7.0 cm field, were in nice agreement, while those obtained with the PinPoint chamber showed differences that increased with the depth in water. At the phantom entrance, all measurements showed non-negligible differences that made Monte Carlo a good option to estimate the absorbed dose in this region.
引用
收藏
页码:3417 / 3431
页数:15
相关论文
共 50 条
  • [1] Monte Carlo Simulation of a Micro-multileaf Collimator
    Gonzalez, W.
    Lallena, A. M.
    Alfonso, R.
    [J]. 5TH LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2011): SUSTAINABLE TECHNOLOGIES FOR THE HEALTH OF ALL, PTS 1 AND 2, 2013, 33 (1-2): : 1252 - +
  • [2] Effects of collimator backscatter in an Elekta linac by Monte Carlo simulation
    Kairn, T.
    Crowe, S. B.
    Poole, C. M.
    Fielding, A. L.
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (03) : 129 - 135
  • [3] Effects of collimator backscatter in an Elekta linac by Monte Carlo simulation
    T. Kairn
    S. B. Crowe
    C. M. Poole
    A. L. Fielding
    [J]. Australasian Physics & Engineering Sciences in Medicine, 2009, 32 : 129 - 135
  • [4] CONFORMITY OF LINAC-BASED STEREOTACTIC RADIOSURGERY USING DYNAMIC CONFORMAL ARCS AND MICRO-MULTILEAF COLLIMATOR
    Hazard, Lisa J.
    Wang, Brian
    Skidmore, Thomas B.
    Chern, Shyh-Shi
    Salter, Bill J.
    Jensen, Randy L.
    Shrieve, Dennis C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02): : 562 - 570
  • [5] Gamma Knife, CyberKnife or micro-multileaf collimator LINAC for intracranial radiosurgery?
    Mindermann, Thomas
    [J]. ACTA NEUROCHIRURGICA, 2015, 157 (04) : 557 - 558
  • [6] Gamma Knife, CyberKnife or micro-multileaf collimator LINAC for intracranial radiosurgery?
    Thomas Mindermann
    [J]. Acta Neurochirurgica, 2015, 157 : 557 - 558
  • [7] Study on the tongue and groove effect of the Elekta multileaf collimator using Monte Carlo simulation and film dosimetry
    Haryanto, F
    Fippel, M
    Bakai, A
    Nüsslin, F
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2004, 180 (01) : 57 - 61
  • [8] Monte Carlo modeling of a micro multileaf collimator
    Vieira, A. M.
    Cruz, J. C.
    Lopes, C.
    Sakuraba, R.
    Caprioglio, L.
    Rodrigues, L. N.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S228 - S228
  • [9] Monte Carlo simulation of the Elekta VersaHD linac
    Borzov, E.
    Nevelsky, A.
    Daniel, S.
    Bar-Deroma, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S717 - S717
  • [10] Linac-Based Fractionated Stereotactic Radiotherapy with a Micro-Multileaf Collimator for Brainstem Metastasis
    Sugimoto, Tadashi
    Matsuda, Ryosuke
    Tamamoto, Tetsuro
    Hontsu, Shigeto
    Yamaki, Kaori
    Miura, Sachiko
    Park, Young-Su
    Nakase, Hiroyuki
    Hasegawa, Masatoshi
    [J]. WORLD NEUROSURGERY, 2019, 132 : E680 - E686