Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems

被引:4
|
作者
Hariri, Sanaz [1 ]
Shahriari, Majid [2 ]
机构
[1] Shahid Beheshti Univ, Dept Radiat Med Engn, Tehran, Iran
[2] Shahid Beheshti Univ, Dept Radiat Applicat, Tehran, Iran
来源
关键词
Monte Carlo; MCNP4C; multileaf collimator (MLC); intensity-modulated radiation therapy (IMRT); LINEAR-ACCELERATOR; MODEL; TONGUE; BEAM;
D O I
10.1120/jacmp.v11i3.3101
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Due to intensive use of multileaf collimators (MLCs) in clinics, finding an optimum design for the leaves becomes essential. There are several studies which deal with comparison of MLC systems, but there is no article with a focus on offering an optimum design using accurate methods like Monte Carlo. In this study, we describe some characteristics of MLC systems including the leaf tip transmission, beam hardening, leakage radiation and penumbra width for Varian and Elekta 80-leaf MLCs using MCNP4C code. The complex geometry of leaves in these two common MLC systems was simulated. It was assumed that all of the MLC systems were mounted on a Varian accelerator and with a similar thickness as Varian's and the same distance from the source. Considering the obtained results from Varian and Elekta leaf designs, an optimum design was suggested combining the advantages of three common MLC systems and the simulation results of this proposed one were compared with the Varian and the Elekta. The leakage from suggested design is 29.7% and 31.5% of the Varian and Elekta MLCs. In addition, other calculated parameters of the proposed MLC leaf design were better than those two commercial ones. Although it shows a wider penumbra in comparison with Varian and Elekta MLCs, taking into account the curved motion path of the leaves, providing a double focusing design will solve the problem. The suggested leaf design is a combination of advantages from three common vendors (Varian, Elekta and Siemens) which can show better results than each one. Using the results of this theoretical study may bring about superior practical outcomes.
引用
收藏
页码:173 / 185
页数:13
相关论文
共 50 条
  • [41] A Monte Carlo simulation for the radiation imaging technique based on the Hemispherical Rotational Modulation Collimator (H-RMC)
    Le Bao, V.
    Kim, G.
    [J]. JOURNAL OF INSTRUMENTATION, 2017, 12
  • [42] Monte Carlo simulation of CZT-based cardiac SPECT systems
    Liu, Zemei
    Ljungberg, Michael
    Chan, Chung
    Liu, Chi
    Holstensson, Maria
    Hutton, Brian
    Pretorius, P.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [43] Monte Carlo design and simulation of a grid-type multi-layer pixel collimator for radiotherapy: Feasibility study
    Yoon, Do-Kun
    Jung, Joo-Young
    Suh, Tae Suk
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (09) : 1385 - 1394
  • [44] Combining dynamic process simulation and Monte Carlo simulation for risk-based design
    Noh, Yeelyong
    Chang, Daejun
    [J]. SAFETY, RELIABILITY AND RISK ANALYSIS: BEYOND THE HORIZON, 2014, : 1903 - 1906
  • [45] A new variable parallel holes collimator for scintigraphic device with validation method based on Monte Carlo simulations
    Trinci, G.
    Massari, R.
    Scandellari, M.
    Boccalini, S.
    Costantini, S.
    Di Sero, R.
    Basso, A.
    Sala, R.
    Scopinaro, F.
    Soluri, A.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3): : 406 - 412
  • [46] A New Algorithm for the Risk of Project Time Based on Monte Carlo Simulation
    Bi, Xing
    Chen, Zhiyuan
    Li, Lei
    [J]. PROCEEDINGS OF THE 21ST INTERNATIONAL CONFERENCE ON INDUSTRIAL ENGINEERING AND ENGINEERING MANAGEMENT 2014, 2015, : 463 - 466
  • [47] Two and Three Dimensional Monte Carlo Simulation of Magnetite Nanoparticle Based Ferrofluids
    Montazeri, R.
    Zamankhan, P.
    [J]. INTERNATIONAL JOURNAL OF ENGINEERING, 2013, 26 (01): : 99 - 104
  • [48] A simple Monte Carlo simulation method for geotechnical reliability-based design
    Li, Bin
    Zhang, Lianyu
    Yuan, Jinquan
    [J]. GEORISK-ASSESSMENT AND MANAGEMENT OF RISK FOR ENGINEERED SYSTEMS AND GEOHAZARDS, 2024, 18 (03) : 558 - 569
  • [49] Reliability-based robust geotechnical design using Monte Carlo simulation
    Peng, Xing
    Li, Dian-Qing
    Cao, Zi-Jun
    Gong, Wenping
    Juang, C. Hsein
    [J]. BULLETIN OF ENGINEERING GEOLOGY AND THE ENVIRONMENT, 2017, 76 (03) : 1217 - 1227
  • [50] Optimal design for μBq/kg gamma spectrometer based on Monte Carlo simulation
    Pan, Xingyu
    Ma, Hao
    Zeng, Zhi
    Zeng, Weihe
    Zeng, Ming
    Cheng, Jianping
    Yue, Qian
    Li, Junli
    Zhang, Hui
    [J]. APPLIED RADIATION AND ISOTOPES, 2020, 157