The calibration of targeting errors for an ultrasound-guided high-intensity focused ultrasound system

被引:0
|
作者
Li, Ke [1 ,2 ]
Bai, Jingfeng [1 ,2 ]
Chen, Yazhu [1 ,2 ]
Ji, Xiang [1 ,2 ]
机构
[1] Shanghai Jiao Tong Univ, Sch Biomed Engn, Biomed Instrument Inst, Shanghai, Peoples R China
[2] Shanghai Jiao Tong Univ, Shanghai Med X Engn Res Ctr, Shanghai, Peoples R China
基金
中国国家自然科学基金;
关键词
Targeting accuracy; Ultrasound-guided high-intensity focused ultrasound (USgHIFU); Calibration; Phasedarray; ACCURACY;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Accurate targeting is one indispensable feature of image-guided high-intensity focused ultrasound (HIFU) systems for treatment safety and efficacy. In our previously developed ultrasound-guided phased-array HIFU system, a rotatable imaging probe was mounted into the central hole of applicator for targeting and monitoring. Two-dimensional image sequence of different imaging planes can be obtained by rotation of the probe. The misalignment between the spots predetermined in the image and the spots sonicated in the tissue can result in the ablation of normal tissue outside the targeting volume, and thus targeting error is unavoidable. An acrylic plate internally placing two flat-head bolts was constructed to measure and calibrate the targeting error. The imaging planes were switched from -90 degrees to 90 degrees with a 30 degrees step, and the targeting errors were measured 12 times for each of these planes before and after calibration. The targeting errors in other imaging planes could be estimated by linear interpolation using the measured errors in the nearest two imaging planes. The coordinates of the spots to be sonicated were corrected in consideration of the targeting errors in the selected imaging plane. After calibration, the mean targeting errors were reduced to 0.30 similar to 0.68 mm from 0.86 similar to 1.74 mm. Besides, in the ex vivo experiment the needle-thermocouple tip was used as the target which could be identified in the image. The temperature rise measured by the thermocouple during sonication was in accordance with the theoretical result. In conclusion, the calibration of targeting errors is effective for our system, and the targeting accuracy is also capable to ensure safe sonication.
引用
收藏
页码:10 / 14
页数:5
相关论文
共 50 条
  • [21] Treatment of Breast Fibroadenoma with Ultrasound-Guided High-Intensity Focused Ultrasound Ablation: A Feasibility Study
    Brenin, David R.
    Patrie, James
    Nguyen, Jonathan
    Rochman, Carrie M.
    JOURNAL OF BREAST IMAGING, 2019, 1 (04) : 316 - 323
  • [22] Ultrasound-Guided High-Intensity Focused Ultrasound vs Laparoscopic Myomectomy for Symptomatic Uterine Myomas
    Wang, Fulan
    Tang, Liangdan
    Wang, Li
    Wang, Xiaoyan
    Chen, Jinyun
    Liu, Xiaohua
    Gong, Yao
    JOURNAL OF MINIMALLY INVASIVE GYNECOLOGY, 2014, 21 (02) : 279 - 284
  • [23] Pregnancy outcomes in patients with uterine fibroids treated with ultrasound-guided high-intensity focused ultrasound
    Zou, M.
    Chen, L.
    Wu, C.
    Hu, C.
    Xiong, Y.
    BJOG-AN INTERNATIONAL JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 2017, 124 : 30 - 35
  • [24] Long-term Efficacy of Fibroid Devascularization with Ultrasound-Guided High-Intensity Focused Ultrasound
    Zhou, Yun
    Chen, Ping
    Ji, Xiang
    Sun, Taotao
    Li, Yuhong
    Yuan, Jiangjing
    Chen, Youguo
    Xiong, Yuanhuan
    Wang, Yudong
    ACADEMIC RADIOLOGY, 2024, 31 (05) : 1931 - 1939
  • [25] A Prediction of NPVR ≥ 80% of Ultrasound-Guided High-Intensity Focused Ultrasound Ablation for Uterine Fibroids
    Yang, Mei-Jie
    Yu, Ren-Qiang
    Chen, Wen-Zhi
    Chen, Jin-Yun
    Wang, Zhi-Biao
    FRONTIERS IN SURGERY, 2021, 8
  • [26] An Integrated Ultrasound-guided High Intensity Focused Ultrasound System for In-vivo Experiment
    Liu, Dalong
    Ebbini, Emad S.
    PROCEEDINGS FROM THE 14TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2017, 1821
  • [27] Novel treatment system using endoscopic ultrasound-guided high-intensity focused ultrasound: A proof-of-concept study
    Ashida, Reiko
    Kawabata, Ken-ichi
    Asami, Rei
    Kitano, Masayuki
    PANCREATOLOGY, 2024, 24 (01) : 88 - 92
  • [28] Safety of ultrasound-guided high-intensity focused ultrasound ablation for diffuse adenomyosis: A retrospective cohort study
    Feng, Yujie
    Hu, Liang
    Chen, Wenzhi
    Zhang, Rong
    Wang, Xi
    Chen, Jinyun
    ULTRASONICS SONOCHEMISTRY, 2017, 36 : 139 - 145
  • [29] Abdominal wall endometriosis treatment by ultrasound-guided high-intensity focused ultrasound ablation: a case report
    Lee, Jae-Seong
    Kim, Ye-Jin
    Hong, Gi-Youn
    Nam, Su-Kyung
    Kim, Tae-Eung
    GYNECOLOGICAL ENDOCRINOLOGY, 2019, 35 (02) : 109 - 111
  • [30] Pregnancy outcomes following ultrasound-guided high-intensity focused ultrasound in submucous leiomyomas: a retrospective study
    Qu, Dacheng
    Liu, Yue
    Jiang, Jing
    Shi, Qiuling
    Zhou, Honggui
    Wang, Zhibiao
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2023, 40 (01)