Correction of Shearing Distortions in Echo-Planar Imaging

被引:2
|
作者
Liu, Guoxiang [1 ,2 ]
Oshio, Koichi [3 ]
Ogawa, Seiji [4 ]
Murata, Tsutomu [1 ]
机构
[1] Natl Inst Informat & Commun Technol, Biol ICT Grp, Kobe, Hyogo, Japan
[2] JST, CREST, Tokyo, Japan
[3] Keio Univ, Sch Med, Dept Diagnost Radiol, Tokyo, Japan
[4] Tohoku Fukushi Univ, Sendai, Miyagi, Japan
关键词
Distortion; echo-planar imaging (EPI); eddy currents; magnetic resonance imaging; EPI; REDUCTION; FMRI;
D O I
10.1109/TMAG.2010.2041785
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this study, the existence of linear geometric distortions due to readout (RO) train-induced background gradients was investigated under homogeneity; this paper was aimed at deriving and implementing a simple method for correcting such linear geometric distortions without carrying out the reference measurements of field errors within the echo-planar imaging (EPI) acquisition window. The background gradient in the EPI acquisition window in the RO direction was estimated from the autocorrelation phase, which was calculated from the complex-valued RO1-D inverse-Fourier-transformed k-space (x-ky space). All the correction procedures were performed in the complex domain in the hybrid x-ky space after acquiring the data. No prior adjustments or additional measurements are required, and this correction algorithm can be applied to most EPI techniques. The echo-planar images of several phantoms and a human brain were acquired with one 1.5-T and two 3-T MRI systems to validate the effectiveness of the correction method. The severe global spatial shearing distortion caused by the linear background RO gradient in the images was corrected, and improvements were observed in all cases. The results indicate that distortions caused by the background RO gradient under homogeneity can be corrected by online processing, even in high-resolution echo-planar images such as 128 X 128 pixel imaging.
引用
收藏
页码:2628 / 2634
页数:7
相关论文
共 50 条
  • [41] Correction of shear artifact caused by unbalanced alternating readout gradient in echo-planar Imaging
    Chiou, JY
    Ahn, CB
    Muftuler, LT
    Nalcioglu, O
    PHYSICA MEDICA, 2003, 19 (02) : 175 - 179
  • [42] Distortion correction for a double inversion-recovery sequence with an echo-planar imaging readout
    Meara, Simon J. P.
    Ernbleton, Karl V.
    Parker, Geoffrey J. M.
    MAGNETIC RESONANCE IMAGING, 2008, 26 (07) : 943 - 953
  • [43] A flexible two-dimensional phase correction for interleaved echo-planar imaging reconstruction
    Liu, HY
    JOURNAL OF APPLIED PHYSICS, 1998, 83 (11) : 6846 - 6848
  • [44] Flexible two-dimensional phase correction for interleaved echo-planar imaging reconstruction
    Liu, Haiying
    Journal of Applied Physics, 1998, 83 (11 pt 2):
  • [45] Overdiscrete echo-planar spectroscopic imaging with correlated higher-order phase correction
    Coello, Eduardo
    Hafalir, Fatih S.
    Noeske, Ralph
    Menzel, Marion
    Haase, Axel
    Menze, Bjoern
    Schulte, Rolf F.
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (01) : 11 - 24
  • [46] FUNCTIONAL MAGNETIC-RESONANCE-IMAGING WITH ECHO-PLANAR IMAGING
    KWONG, KK
    MAGNETIC RESONANCE QUARTERLY, 1995, 11 (01) : 1 - 20
  • [47] Center-out echo-planar spectroscopic imaging with correction of gradient-echo phase and time shifts
    Labadie, Christian
    Hetzer, Stefan
    Schulz, Jessica
    Mildner, Toralf
    Aubert-Frecon, Monique
    Moeller, Harald E.
    MAGNETIC RESONANCE IN MEDICINE, 2013, 70 (01) : 16 - 24
  • [48] Three dimensional echo-planar imaging at 7 Tesla
    Poser, B. A.
    Koopmans, P. J.
    Witzel, T.
    Wald, L. L.
    Barth, M.
    NEUROIMAGE, 2010, 51 (01) : 261 - 266
  • [49] REAL-TIME ECHO-PLANAR IMAGING BY NMR
    MANSFIELD, P
    BRITISH MEDICAL BULLETIN, 1984, 40 (02) : 187 - &
  • [50] The development of echo-planar imaging (EPI): 1977–1982
    R. Ordidge
    Magnetic Resonance Materials in Physics, Biology and Medicine, 1999, 9 (3) : 117 - 121