Design and validation of a CT-guided robotic system for lung cancer brachytherapy

被引:18
|
作者
Dou, Huaisu [1 ]
Jiang, Shan [1 ,2 ]
Yang, Zhiyong [1 ]
Sun, Luqing [1 ]
Ma, Xiaodong [1 ]
Huo, Bin [3 ]
机构
[1] Tianjin Univ, Sch Mech Engn, Tianjin 300350, Peoples R China
[2] Tianjin Univ, Ctr Adv Mech & Robot, Tianjin 300350, Peoples R China
[3] Tianjin Med Univ, Dept Oncol, Hosp 2, Tianjin 300211, Peoples R China
基金
中国国家自然科学基金;
关键词
CT-guided; lung cancer brachytherapy; surgical robot; treatment planning; PUNCTURE ROBOT; ACCURACY; BIOPSIES; SAFETY;
D O I
10.1002/mp.12435
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Currently, lung brachytherapy in clinical setting is a complex procedure. Operation accuracy depends on accurate positioning of the template; however, it is difficult to guarantee the positioning accuracy manually. Application of robotic-assisted systems can simplify the procedure and improve the manual positioning accuracy. Therefore, a novel CT-guided robotic system was developed to assist the lung cancer brachytherapy. Methods: A four degree-of-freedom (DOF) robot, controlled by a lung brachytherapy treatment planning system (TPS) software, was designed and manufactured to assist the template positioning. Target position of the template can be obtained from the treatment plan, thus the robot is driven to the target position automatically. The robotic system was validated in both the laboratory and the CT environment. In laboratory environment, a 3D laser tracker and an inertial measurement unit (IMU) were used to measure the mechanical accuracy in air, which includes positioning accuracy and position repeatability. Working reliability was also validated in this procedure by observing the response reliability and calculating the position repeatability. Imaging artifacts and accuracy of the robot registration were validated in the CT environment by using an artificial phantom with fiducial markers. CT images were obtained and used to test the image artifact and calculate the registration accuracy. Phantom experiments were conducted to test the accuracy of needle insertion by using a transparent hydrogel phantom with a high imitation artificial phantom. Also, the efficiency was validated in this procedure by comparing time costs in manual positioning with robotic positioning under the same experimental conditions. Results: The robotic system achieved the positioning accuracy of 0.28 +/- 0.25 mm and the position repeatability of 0.09 +/- 0.11 mm. Experimental results showed that the robot was CT-compatible and responded reliably to the control commands. The mean registration accuracy of the robotic system was 0.49 +/- 0.29 mm. Phantom experiments indicated that the accuracy of needle insertion was 1.5 +/- 1.7 mm at a depth ranging from 30 to 80 mm. The time used to adjust the template to the target position was 12 min on average by robotic system automatically. An average of 30 min was saved compared with the manual positioning procedure in phantom experiments. Conclusions: This paper describes the design and experimental validation of a novel CT-guided robotic system for lung cancer brachytherapy. The robotic system was validated in a number of aspects which prove that it was capable of locating the template with clinically acceptable accuracy in the CT environment. All experimental results indicated that the system is reliable and ready to be applied to further studies on animals. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4828 / 4837
页数:10
相关论文
共 50 条
  • [11] CT-guided 125I brachytherapy for recurrent ovarian cancer
    Liu, Ping
    Tong, Lina
    Huo, Bin
    Dai, Dong
    Liu, Wenxin
    Wang, Ke
    Wang, Ying
    Guo, Zhi
    Ni, Hong
    ONCOTARGET, 2017, 8 (35) : 59766 - 59776
  • [12] Estimation of the optimal exposure to urethra in CT-guided brachytherapy of prostate cancer
    Panarina, N.
    Sviridov, P.
    Volodina, T.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S256 - S256
  • [13] CT-GUIDED INTRACAVITARY BRACHYTHERAPY FOR THE TREATMENT OF CERVICAL CANCER: A PROSPECTIVE STUDY
    Du, X.
    Sheng, X.
    Wang, C.
    Lu, C.
    INTERNATIONAL JOURNAL OF GYNECOLOGICAL CANCER, 2013, 23 (08)
  • [14] CT-Guided Pelvic Lymph Nodal Brachytherapy
    Kunogi, Hiroaki
    Hsu, I-Chow
    Yamaguchi, Nanae
    Kusunoki, Soshi
    Nakagawa, Keiko
    Sugimori, Yayoi
    Fujino, Kazunari
    Terao, Yasuhisa
    Ogishima, Daiki
    Yoshimura, Ryoichi
    Sasai, Keisuke
    FRONTIERS IN ONCOLOGY, 2021, 10
  • [15] A Path Planning System for CT-guided Lung Biopsy
    Feng R.
    Tang L.
    Emu Y.
    Zhang X.
    Zhang J.
    Zhong J.
    He L.
    Jiqiren/Robot, 2022, 44 (06): : 694 - 707
  • [16] CT-guided robotic needle biopsy of lung nodules with respiratory motion - experimental system and preliminary test
    Zhou, Yu
    Thiruvalluvan, Kaarvannan
    Krzeminski, Lukasz
    Moore, William H.
    Xu, Zhigang
    Liang, Zhengrong
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2013, 9 (03): : 317 - 330
  • [17] A CT-guided robotic needle puncture method for lung tumours with respiratory motion
    Wei, Lin
    Jiang, Shan
    Yang, Zhiyong
    Zhang, Guobin
    Ma, Liang
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 73 : 48 - 56
  • [18] Clinical evaluation of a robotic system for precise CT-guided percutaneous procedures
    Levy, Shiran
    Goldberg, S. Nahum
    Roth, Ido
    Shochat, Moran
    Sosna, Jacob
    Leichter, Isaac
    Flacke, Sebastian
    ABDOMINAL RADIOLOGY, 2021, 46 (10) : 5007 - 5016
  • [19] Clinical evaluation of a robotic system for precise CT-guided percutaneous procedures
    Shiran Levy
    S. Nahum Goldberg
    Ido Roth
    Moran Shochat
    Jacob Sosna
    Isaac Leichter
    Sebastian Flacke
    Abdominal Radiology, 2021, 46 : 5007 - 5016
  • [20] A new robotic system for CT-guided percutaneous procedures with haptic feedback
    Maurin, B
    Piccin, O
    Bayle, B
    Gangloff, J
    de Mathelin, M
    Soler, L
    Gangi, A
    CARS 2004: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2004, 1268 : 515 - 520