Design and validation of a CT-guided robotic system for lung cancer brachytherapy

被引:18
|
作者
Dou, Huaisu [1 ]
Jiang, Shan [1 ,2 ]
Yang, Zhiyong [1 ]
Sun, Luqing [1 ]
Ma, Xiaodong [1 ]
Huo, Bin [3 ]
机构
[1] Tianjin Univ, Sch Mech Engn, Tianjin 300350, Peoples R China
[2] Tianjin Univ, Ctr Adv Mech & Robot, Tianjin 300350, Peoples R China
[3] Tianjin Med Univ, Dept Oncol, Hosp 2, Tianjin 300211, Peoples R China
基金
中国国家自然科学基金;
关键词
CT-guided; lung cancer brachytherapy; surgical robot; treatment planning; PUNCTURE ROBOT; ACCURACY; BIOPSIES; SAFETY;
D O I
10.1002/mp.12435
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Currently, lung brachytherapy in clinical setting is a complex procedure. Operation accuracy depends on accurate positioning of the template; however, it is difficult to guarantee the positioning accuracy manually. Application of robotic-assisted systems can simplify the procedure and improve the manual positioning accuracy. Therefore, a novel CT-guided robotic system was developed to assist the lung cancer brachytherapy. Methods: A four degree-of-freedom (DOF) robot, controlled by a lung brachytherapy treatment planning system (TPS) software, was designed and manufactured to assist the template positioning. Target position of the template can be obtained from the treatment plan, thus the robot is driven to the target position automatically. The robotic system was validated in both the laboratory and the CT environment. In laboratory environment, a 3D laser tracker and an inertial measurement unit (IMU) were used to measure the mechanical accuracy in air, which includes positioning accuracy and position repeatability. Working reliability was also validated in this procedure by observing the response reliability and calculating the position repeatability. Imaging artifacts and accuracy of the robot registration were validated in the CT environment by using an artificial phantom with fiducial markers. CT images were obtained and used to test the image artifact and calculate the registration accuracy. Phantom experiments were conducted to test the accuracy of needle insertion by using a transparent hydrogel phantom with a high imitation artificial phantom. Also, the efficiency was validated in this procedure by comparing time costs in manual positioning with robotic positioning under the same experimental conditions. Results: The robotic system achieved the positioning accuracy of 0.28 +/- 0.25 mm and the position repeatability of 0.09 +/- 0.11 mm. Experimental results showed that the robot was CT-compatible and responded reliably to the control commands. The mean registration accuracy of the robotic system was 0.49 +/- 0.29 mm. Phantom experiments indicated that the accuracy of needle insertion was 1.5 +/- 1.7 mm at a depth ranging from 30 to 80 mm. The time used to adjust the template to the target position was 12 min on average by robotic system automatically. An average of 30 min was saved compared with the manual positioning procedure in phantom experiments. Conclusions: This paper describes the design and experimental validation of a novel CT-guided robotic system for lung cancer brachytherapy. The robotic system was validated in a number of aspects which prove that it was capable of locating the template with clinically acceptable accuracy in the CT environment. All experimental results indicated that the system is reliable and ready to be applied to further studies on animals. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4828 / 4837
页数:10
相关论文
共 50 条
  • [21] Evaluation of a CT-Guided Robotic System for Precise Percutaneous Needle Insertion
    Ben-David, Eliel
    Shochat, Moran
    Roth, Ido
    Nissenbaum, Isaac
    Sosna, Jacob
    Goldberg, S. Nahum
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2018, 29 (10) : 1440 - 1446
  • [22] Percutaneous CT-guided thermal ablation of primary lung cancer
    Konstantelou, Elissavet
    Hardavella, Georgia
    Assi, Avraam
    Dimitrakopoulou, Marianna
    Kosma, Aikaterini
    Liano, Evangelos
    Fergadi, Evangelos
    Ntokou, Anna
    Charalampakis, Nikolaos
    Kontis, Elissaios
    Karampinis, Ioannis
    Bolanos, Nikolaos
    Zias, Nikolaos
    Kosmas, Christos
    Stefanidis, Konstantinos
    EUROPEAN RESPIRATORY JOURNAL, 2024, 64
  • [23] CT-GUIDED BIOPSY OF LUNG LESIONS
    BELFIORE, G
    DIFILIPPO, S
    GUIDA, C
    MARANO, I
    PORTA, E
    NUCLEAR MEDICINE AND BIOLOGY, 1994, 21 (05): : 713 - 719
  • [24] LUNG ABSCESS - CT-GUIDED DRAINAGE
    VANSONNENBERG, E
    DAGOSTINO, HB
    CASOLA, G
    WITTICH, GR
    VARNEY, RR
    HARKER, C
    RADIOLOGY, 1991, 178 (02) : 347 - 351
  • [25] CT-GUIDED DRAINAGE FOR LUNG ABSCESS
    Maruyama, Eiichi
    Takahara, Norihiro
    Watanabe, Kenichi
    Arakawa, Toshinao
    Watanabe, Azumi
    RESPIROLOGY, 2018, 23 : 183 - 183
  • [26] CT-guided radioactive 125I seeds brachytherapy for lung oligometastases from colorectal cancer: initial results
    Song, Mengyao
    Zhou, Xueliang
    Hou, Rongna
    Sigdel, Milan
    Liu, Yiming
    Zhang, Chengzhi
    Xu, Kaihao
    Han, Xinwei
    Jiao, Dechao
    BMC CANCER, 2024, 24 (01)
  • [27] CT-guided radioactive 125I seeds brachytherapy for lung oligometastases from colorectal cancer: initial results
    Mengyao Song
    Xueliang Zhou
    Rongna Hou
    Milan Sigdel
    Yiming Liu
    Chengzhi Zhang
    Kaihao Xu
    Xinwei Han
    Dechao Jiao
    BMC Cancer, 24
  • [28] A new robotic assistance system for percutaneous CT-guided punctures: Initial experience
    Groetz, Simon
    Wilhelm, Kai
    Willinek, Winfried
    Pieper, Claus
    Schild, Hans
    Thomas, Daniel
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2016, 25 (02) : 79 - 85
  • [29] CT-Guided Interstitial Low Dose-RATE Brachytherapy for Recurrent Colorectal Cancer
    Ng, V.
    Boas, F. E.
    Cohen, G.
    Moore, A.
    Kemeny, N.
    Weiser, M.
    Paty, P.
    Crane, C. H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : E65 - E65
  • [30] DIAGNOSIS OF LUNG NODULES BY CT-GUIDED LUNG BIOPSY
    Song, Yong
    JOURNAL OF THORACIC ONCOLOGY, 2013, 8 : S129 - S129