Intraosseous blood samples for point-of-care analysis: agreement between intraosseous and arterial analyses

被引:9
|
作者
Jousi, Milla [1 ,2 ,3 ,5 ]
Saikko, Simo [4 ]
Nurmi, Jouni [1 ,2 ,3 ]
机构
[1] Helsinki Univ Hosp, Emergency Med & Serv, Helsinki, Finland
[2] Univ Helsinki, Dept Emergency Med, Helsinki, Finland
[3] HYKS Akuutti, PL 340, Hus Helsinki 00029, Finland
[4] Saimaa Univ Appl Sci, Skinnarilankatu 36, Lappeenranta 53850, Finland
[5] FinnHEMS Res & Dev Unit, Lentajantie 3, Vantaa 01530, Finland
关键词
Point-of-care; Blood gas analysis; Intraosseous access; I-STAT; BONE-MARROW ASPIRATE; ACID-BASE STATUS; ANESTHETIZED PIG; CARDIOPULMONARY-RESUSCITATION; VENOUS SITES; RELIABILITY; GUIDELINES; VALUES; SHOCK;
D O I
10.1186/s13049-017-0435-4
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Point-of-care (POC) testing is highly useful when treating critically ill patients. In case of difficult vascular access, the intraosseous (IO) route is commonly used, and blood is aspirated to confirm the correct position of the IO-needle. Thus, IO blood samples could be easily accessed for POC analyses in emergency situations. The aim of this study was to determine whether IO values agree sufficiently with arterial values to be used for clinical decision making. Methods: Two samples of IO blood were drawn from 31 healthy volunteers and compared with arterial samples. The samples were analysed for sodium, potassium, ionized calcium, glucose, haemoglobin, haematocrit, pH, blood gases, base excess, bicarbonate, and lactate using the i-STAT (R) POC device. Agreement and reliability were estimated by using the Bland-Altman method and intraclass correlation coefficient calculations. Results: Good agreement was evident between the IO and arterial samples for pH, glucose, and lactate. Potassium levels were clearly higher in the IO samples than those from arterial blood. Base excess and bicarbonate were slightly higher, and sodium and ionised calcium values were slightly lower, in the IO samples compared with the arterial values. The blood gases in the IO samples were between arterial and venous values. Haemoglobin and haematocrit showed remarkable variation in agreement. Discussion: POC diagnostics of IO blood can be a useful tool to guide treatment in critical emergency care. Seeking out the reversible causes of cardiac arrest or assessing the severity of shock are examples of situations in which obtaining vascular access and blood samples can be difficult, though information about the electrolytes, acid-base balance, and lactate could guide clinical decision making. The analysis of IO samples should though be limited to situations in which no other option is available, and the results should be interpreted with caution, because there is not yet enough scientific evidence regarding the agreement of IO and arterial results among unstable patients. Conclusions: IO blood samples are suitable for analysis with the i-STAT (R) point-of-care device in emergency care. The aspirate used to confirm the correct placement of the IO needle can also be used for analysis. The results must be interpreted within a clinical context while taking the magnitude and direction of bias into account.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Intraosseous blood samples for point-of-care analysis: agreement between intraosseous and arterial analyses
    Milla Jousi
    Simo Saikko
    Jouni Nurmi
    Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine, 25
  • [2] Point-of-care analysis of intraosseous blood samples
    Jousi, Milla
    Nurmi, Jouni
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2019, 63 (08) : E21 - E21
  • [3] Point-of-care analyses of blood samples from intraosseous access in pre-hospital critical care
    Jousi, Milla
    Bjorkman, Johannes
    Nurmi, Jouni
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2019, 63 (10) : 1419 - 1425
  • [4] Point-of-care laboratory analyses of intraosseous, arterial and central venous samples during experimental cardiopulmonary resuscitation
    Jousi, Mille
    Skrifvars, Markus B.
    Nelskyla, Annika
    Ristagno, Giuseppe
    Schramko, Alexey
    Nurmi, Jouni
    RESUSCITATION, 2019, 137 : 124 - 132
  • [5] Analysis of intraosseous blood samples using an EPOC point of care analyzer during resuscitation
    Tallman, Crystal Ives
    Darracq, Michael
    Young, Megann
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2017, 35 (03): : 499 - 501
  • [6] Intraosseous Blood Correlates with Venous Blood in Healthy Subjects Using Point-of-Care Analyzers
    Montez, D. F.
    Puga, T. A.
    Garcia, M. R.
    Miller, L. J.
    Davlantes, C.
    Saussy, J. M.
    Philbeck, T. E., Jr.
    ANNALS OF EMERGENCY MEDICINE, 2013, 62 (04) : S40 - S40
  • [7] Analysis of bloodgas, electrolytes and glucose from intraosseous samples using an i-STAT® point-of-care analyser
    Veldhoen, Esther S.
    de Vooght, Karen M. K.
    Slieker, Martijn G.
    Versluys, Anne B.
    Turner, Nigel Mcb.
    RESUSCITATION, 2014, 85 (03) : 359 - 363
  • [8] Analysis of intraosseous samples using point of care technology-An experimental study in the anaesthetised pig
    Strandberg, Gunnar
    Eriksson, Mats
    Gustafsson, Mats G.
    Lipcsey, Miklos
    Larsson, Anders
    RESUSCITATION, 2012, 83 (11) : 1381 - 1385
  • [9] POINT-OF-CARE ULTRASOUND DIAGNOSIS OF INTRAVASCULAR AIR AFTER LOWER EXTREMITY INTRAOSSEOUS ACCESS
    Azan, Benjamin
    Teran, Felipe
    Nelson, Bret P.
    Andrus, Phillip
    JOURNAL OF EMERGENCY MEDICINE, 2016, 51 (06): : 680 - 683
  • [10] Determining Intraosseous Needle Placement Using Point-of-Care Ultrasound in a Swine (Sus scrofa) Model
    Kyle, Adrianna, I
    Auten, Jonathan D.
    Zarow, Gregory J.
    Natarajan, Ramesh
    Bianchi, William D.
    Speicher, Matthew, V
    Palma, James
    Gaspary, Micah J.
    MILITARY MEDICINE, 2023, 188 (9-10) : 2969 - 2974