摘要
Objective. The objective of this study was to separately quantify the stability of the megavoltage imager (MVI) and radiation head of an Elekta Unity MRL, throughout full gantry rotation. Approach. A ball-bearing (BB) phantom was attached to the radiation head of the Unity, while a single BB was placed at isocentre. Images were acquired during rotation, using the MVI. These images were processed using an in-house developed MATLAB program to reduce the errors resulted by noise, and the positions of the BBs in the images were analysed to extract MVI and radiation head sag data. Main results. The results returned by this method showed reproducibility, with a mean standard deviation of 7 mu m for the position of BBs across all gantry angles. The radiation head was found to sag throughout rotation, with a maximum course of movement of 0.59 mm. The sag pattern was stable over a period greater than a year but showed some dependence on gantry rotation direction. Significance. As MRL is a relatively new system, it is promising to have data supporting the high level of precision on one Elekta Unity machine. Isolating and quantifying the sources of uncertainty in radiation delivery may allow more sophisticated analysis of how the system performance may be improved.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] MRI B0 homogeneity and geometric distortion with continuous linac gantry rotation on an Elekta Unity MR-linac
    Jackson, Steven
    Glitzner, Markus
    Tijssen, Rob H. N.
    Raaymakers, Bas W.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (12):
  • [2] ADC measurements on the Unity MR-linac - A recommendation on behalf of the Elekta Unity MR-linac consortium
    Kooreman, Ernst S.
    van Houdt, Petra J.
    Keesman, Rick
    Pos, Floris J.
    van Pelt, Vivian W. J.
    Nowee, Marlies E.
    Wetscherek, Andreas
    Tijssen, Rob H. N.
    Philippens, Marielle E. P.
    Thorwarth, Daniela
    Wang, Jihong
    Shukla-Dave, Amita
    Hall, William A.
    Paulson, Eric S.
    van der Heide, Uulke A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 153 : 106 - 113
  • [3] Commissioning measurements on an Elekta Unity MR-Linac
    Powers, Marcus
    Baines, John
    Crane, Robert
    Fisher, Chantelle
    Gibson, Stephen
    Marsh, Linda
    Oar, Bronwyn
    Shoobridge, Ariadne
    Simpson-Page, Emily
    Van der Walt, Marchant
    de Vine, Glenn
    [J]. PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (02) : 457 - 473
  • [4] Commissioning measurements on an Elekta Unity MR-Linac
    Marcus Powers
    John Baines
    Robert Crane
    Chantelle Fisher
    Stephen Gibson
    Linda Marsh
    Bronwyn Oar
    Ariadne Shoobridge
    Emily Simpson-Page
    Marchant Van der Walt
    Glenn de Vine
    [J]. Physical and Engineering Sciences in Medicine, 2022, 45 : 457 - 473
  • [5] Adaptive radiotherapy: The Elekta Unity MR-linac concept
    Winkel, Dennis
    Bol, Gijsbert H.
    Kroon, Petra S.
    van Asselen, Bram
    Hackett, Sara S.
    Werensteijn-Honingh, Anita M.
    Intven, Martijn P. W.
    Eppinga, Wietse S. C.
    Tijssen, Rob H. N.
    Kerkmeijer, Linda G. W.
    de Boer, Hans C. J.
    Mook, Stella
    Meijer, Gert J.
    Hes, Jochem
    Willemsen-Bosman, Mirjam
    de Groot-van Breugel, Eline N.
    Jurgenliemk-Schulz, Ina M.
    Raaymakers, Bas W.
    [J]. CLINICAL AND TRANSLATIONAL RADIATION ONCOLOGY, 2019, 18 : 54 - 59
  • [6] Developing a QA programme for the Elekta Unity MR-linac
    Chick, J.
    Hanson, I.
    Nill, S.
    Oelfke, U.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S1167 - S1168
  • [7] Elekta Unity MR-linac commissioning: mechanical and dosimetry tests
    Tsuneda, Masato
    Abe, Kota
    Fujita, Yukio
    Ikeda, Yohei
    Furuyama, Yoshinobu
    Uno, Takashi
    [J]. JOURNAL OF RADIATION RESEARCH, 2023, 64 (01) : 73 - 84
  • [8] Machine QA for the Elekta Unity system: A Report from the Elekta MR-linac consortium
    Roberts, David A.
    Sandin, Carlos
    Vesanen, Panu T.
    Lee, Hannah
    Hanson, Ian M.
    Nill, Simeon
    Perik, Thijs
    Lim, Seng Boh
    Vedam, Sastry
    Yang, Jinzhong
    Woodings, Simon W.
    Wolthaus, Jochem W. H.
    Keller, Brian
    Budgell, Geoff
    Chen, Xinfeng
    Li, X. Allen
    [J]. MEDICAL PHYSICS, 2021, 48 (05) : E67 - E85
  • [9] Feasibility of spinal stereotactic body radiotherapy in Elekta Unity® MR-Linac
    Han, Eun Young
    Aima, Manik
    Hughes, Neil
    Briere, Tina M.
    Yeboa, Debra N.
    Castillo, Pam
    Wang, Jihong
    Yang, Jinzhong
    Vedam, Sastry
    [J]. JOURNAL OF RADIOSURGERY AND SBRT, 2020, 7 (02): : 127 - 134
  • [10] Secondary MU Check Using RadCalc for Elekta Unity MR-Linac
    Yang, J.
    Vedam, S.
    Brink, C.
    Nielsen, M.
    Ji, M.
    Bist, L.
    Ohrt, J.
    Wang, J.
    Balter, P.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E267 - E267