Variation in isocentre location of an Elekta Unity MR-linac through full gantry rotation

被引:2
|
作者
Hunt, James R. [1 ]
Ebert, Martin A. [1 ,2 ,3 ]
Rowshanfarzad, Pejman [1 ]
Riis, Hans L. [4 ,5 ]
机构
[1] Univ Western Australia, Sch Phys Math & Comp, Crawley, WA 6009, Australia
[2] Sir Charles Gairdner Hosp, Dept Radiat Oncol, Nedlands, WA 6009, Australia
[3] 5D Clin, Claremont, WA 6010, Australia
[4] Odense Univ Hosp, Dept Oncol, Radiofys Lab, Odense, Denmark
[5] Univ Southern Denmark, Dept Clin Res, Odense, Denmark
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2022年 / 67卷 / 01期
关键词
MR linac; sag; accuracy; radiotherapy; LINEAR ACCELERATORS; MLC; PERFORMANCE; SAG;
D O I
10.1088/1361-6560/ac4564
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. The objective of this study was to separately quantify the stability of the megavoltage imager (MVI) and radiation head of an Elekta Unity MRL, throughout full gantry rotation. Approach. A ball-bearing (BB) phantom was attached to the radiation head of the Unity, while a single BB was placed at isocentre. Images were acquired during rotation, using the MVI. These images were processed using an in-house developed MATLAB program to reduce the errors resulted by noise, and the positions of the BBs in the images were analysed to extract MVI and radiation head sag data. Main results. The results returned by this method showed reproducibility, with a mean standard deviation of 7 mu m for the position of BBs across all gantry angles. The radiation head was found to sag throughout rotation, with a maximum course of movement of 0.59 mm. The sag pattern was stable over a period greater than a year but showed some dependence on gantry rotation direction. Significance. As MRL is a relatively new system, it is promising to have data supporting the high level of precision on one Elekta Unity machine. Isolating and quantifying the sources of uncertainty in radiation delivery may allow more sophisticated analysis of how the system performance may be improved.
引用
收藏
页数:14
相关论文
共 50 条
  • [21] Beam and Cryostat Scatter Characteristics of the Elekta MR-Linac
    Duglio, M.
    Towe, S.
    Roberts, D.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3442 - 3442
  • [22] Developing a Source Model for the Elekta Unity MR-Linac Using a Grid-Based Boltzmann Solver
    Nowak, S.
    Hyer, D.
    Yang, R.
    St-Aubin, J.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E159 - E159
  • [23] Feasibility of Skull Base Stereotactic Radiotherapy on Elekta MR-Linac
    Wang, H.
    Yang, J.
    Phan, J.
    Han, E.
    Lim, T.
    Castillo, P.
    Wang, X.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [24] Can a dual isocentre technique enable cervix treatments on the MR-Linac?
    Chuter, R.
    Cree, A.
    Whitehurst, P.
    Hales, R.
    McWilliam, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S195 - S196
  • [25] Clinical Experience of Patient-Specific QA for Online Adaptive Radiotherapy Using Elekta Unity MR-Linac
    Yang, J.
    Vedam, S.
    Wang, J.
    Fuller, C.
    Choi, S.
    Chung, C.
    McAleer, M.
    Lee, B.
    Hughes, N.
    Ungchusri, G.
    Gillin, M.
    Martel, M.
    Balter, P.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E204 - E204
  • [26] Superparamagnetic Iron Oxide Enhanced MRI for Elekta Unity MR-Linac: A Pre-Clinical Phantom Study
    Lee, D.
    Klepchick, P.
    Lee, H.
    Pavord, D.
    Oh, S.
    Sohn, J.
    Kirichenko, A.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [27] Proof-of-concept delivery of intensity modulated arc therapy on the Elekta Unity 1.5 T MR-linac
    Kontaxis, C.
    Woodhead, P. L.
    Bol, G. H.
    Lagendijk, J. J. W.
    Raaymakers, B. W.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (04):
  • [28] Real-Time Adaptive Dose Scaling During MLC-Tracking on the Elekta Unity MR-Linac
    Borman, P. T. S.
    Woodhead, P. L.
    Uijtewaal, P.
    Hackett, S. L.
    Lagendijk, J. J. W.
    Raaymakers, B. W.
    Fast, M. F.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E437 - E437
  • [29] Feasibility of MRI-guided VMAT: investigating image quality during gantry rotation on an MR-linac
    Jackson, S.
    Glitzner, M.
    Tijssen, R. H. N.
    Raaymakers, B. W.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S270 - S271
  • [30] Benchmarking daily plan adaptation on the Unity MR-Linac
    Jagt, T.
    Janssen, T.
    Betgen, A.
    Wiersema, L.
    Verhage, R.
    Garritsen, S.
    Vijlbrief-Bosman, T.
    de Ruiter, P.
    Peters, F.
    Remeijer, P.
    Marijnen, C.
    Sonke, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S60 - S62