Machine QA for the Elekta Unity system: A Report from the Elekta MR-linac consortium

被引:57
|
作者
Roberts, David A. [1 ]
Sandin, Carlos [1 ]
Vesanen, Panu T. [2 ]
Lee, Hannah [3 ]
Hanson, Ian M. [4 ,5 ]
Nill, Simeon [4 ,5 ]
Perik, Thijs [6 ]
Lim, Seng Boh [7 ]
Vedam, Sastry [8 ]
Yang, Jinzhong [8 ]
Woodings, Simon W. [9 ]
Wolthaus, Jochem W. H. [9 ]
Keller, Brian [10 ]
Budgell, Geoff [11 ]
Chen, Xinfeng [12 ,13 ]
Li, X. Allen [12 ,13 ]
机构
[1] Elekta Ltd, London Rd, Crawley RH10 9BL, England
[2] Philips Healthcare, Vantaa, Finland
[3] Allegheny Hlth Network, Canc Inst, Pittsburgh, PA USA
[4] Inst Canc Res, Joint Dept Phys, Sutton, Surrey, England
[5] Royal Marsden NHS Fdn Trust, Sutton, Surrey, England
[6] Antoni van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiat Oncol, Amsterdam, Netherlands
[7] Mem Sloan Kettering Canc Ctr, 1275 York Ave, New York, NY 10021 USA
[8] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[9] Univ Med Ctr Utrecht, Dept Radiotherapy, Utrecht, Netherlands
[10] Sunnybrook Hlth Sci Ctr, Odette Canc Ctr, Toronto, ON, Canada
[11] Christie NHS Fdn Trust, Christie Med Phys & Engn, Wilmslow Rd, Manchester, Lancs, England
[12] Froedtert Hosp, Dept Radiat Oncol, Milwaukee, WI USA
[13] Med Coll Wisconsin, Milwaukee, WI 53226 USA
关键词
radiotherapy; quality assurance; MRI‐ linac; dosimetry;
D O I
10.1002/mp.14764
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Over the last few years, magnetic resonance image-guided radiotherapy systems have been introduced into the clinic, allowing for daily online plan adaption. While quality assurance (QA) is similar to conventional radiotherapy systems, there is a need to introduce or modify measurement techniques. As yet, there is no consensus guidance on the QA equipment and test requirements for such systems. Therefore, this report provides an overview of QA equipment and techniques for mechanical, dosimetric, and imaging performance of such systems and recommendation of the QA procedures, particularly for a 1.5T MR-linac device. An overview of the system design and considerations for QA measurements, particularly the effect of the machine geometry and magnetic field on the radiation beam measurements is given. The effect of the magnetic field on measurement equipment and methods is reviewed to provide a foundation for interpreting measurement results and devising appropriate methods. And lastly, a consensus overview of recommended QA, appropriate methods, and tolerances is provided based on conventional QA protocols. The aim of this consensus work was to provide a foundation for QA protocols, comparative studies of system performance, and for future development of QA protocols and measurement methods.
引用
收藏
页码:E67 / E85
页数:19
相关论文
共 50 条
  • [1] Developing a QA programme for the Elekta Unity MR-linac
    Chick, J.
    Hanson, I.
    Nill, S.
    Oelfke, U.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S1167 - S1168
  • [2] ADC measurements on the Unity MR-linac - A recommendation on behalf of the Elekta Unity MR-linac consortium
    Kooreman, Ernst S.
    van Houdt, Petra J.
    Keesman, Rick
    Pos, Floris J.
    van Pelt, Vivian W. J.
    Nowee, Marlies E.
    Wetscherek, Andreas
    Tijssen, Rob H. N.
    Philippens, Marielle E. P.
    Thorwarth, Daniela
    Wang, Jihong
    Shukla-Dave, Amita
    Hall, William A.
    Paulson, Eric S.
    van der Heide, Uulke A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 153 : 106 - 113
  • [3] Commissioning measurements on an Elekta Unity MR-Linac
    Powers, Marcus
    Baines, John
    Crane, Robert
    Fisher, Chantelle
    Gibson, Stephen
    Marsh, Linda
    Oar, Bronwyn
    Shoobridge, Ariadne
    Simpson-Page, Emily
    Van der Walt, Marchant
    de Vine, Glenn
    [J]. PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (02) : 457 - 473
  • [4] Commissioning measurements on an Elekta Unity MR-Linac
    Marcus Powers
    John Baines
    Robert Crane
    Chantelle Fisher
    Stephen Gibson
    Linda Marsh
    Bronwyn Oar
    Ariadne Shoobridge
    Emily Simpson-Page
    Marchant Van der Walt
    Glenn de Vine
    [J]. Physical and Engineering Sciences in Medicine, 2022, 45 : 457 - 473
  • [5] Adaptive radiotherapy: The Elekta Unity MR-linac concept
    Winkel, Dennis
    Bol, Gijsbert H.
    Kroon, Petra S.
    van Asselen, Bram
    Hackett, Sara S.
    Werensteijn-Honingh, Anita M.
    Intven, Martijn P. W.
    Eppinga, Wietse S. C.
    Tijssen, Rob H. N.
    Kerkmeijer, Linda G. W.
    de Boer, Hans C. J.
    Mook, Stella
    Meijer, Gert J.
    Hes, Jochem
    Willemsen-Bosman, Mirjam
    de Groot-van Breugel, Eline N.
    Jurgenliemk-Schulz, Ina M.
    Raaymakers, Bas W.
    [J]. CLINICAL AND TRANSLATIONAL RADIATION ONCOLOGY, 2019, 18 : 54 - 59
  • [6] Elekta Unity MR-linac commissioning: mechanical and dosimetry tests
    Tsuneda, Masato
    Abe, Kota
    Fujita, Yukio
    Ikeda, Yohei
    Furuyama, Yoshinobu
    Uno, Takashi
    [J]. JOURNAL OF RADIATION RESEARCH, 2023, 64 (01) : 73 - 84
  • [7] Feasibility of spinal stereotactic body radiotherapy in Elekta Unity® MR-Linac
    Han, Eun Young
    Aima, Manik
    Hughes, Neil
    Briere, Tina M.
    Yeboa, Debra N.
    Castillo, Pam
    Wang, Jihong
    Yang, Jinzhong
    Vedam, Sastry
    [J]. JOURNAL OF RADIOSURGERY AND SBRT, 2020, 7 (02): : 127 - 134
  • [8] Secondary MU Check Using RadCalc for Elekta Unity MR-Linac
    Yang, J.
    Vedam, S.
    Brink, C.
    Nielsen, M.
    Ji, M.
    Bist, L.
    Ohrt, J.
    Wang, J.
    Balter, P.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E267 - E267
  • [9] A new multi-purpose QA phantom for use on the Elekta MR-Linac
    Hanson, I.
    Sullivan, J.
    Nill, S.
    Oelfke, U.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S38 - S39
  • [10] Clinical Experience of Patient-Specific QA for Online Adaptive Radiotherapy Using Elekta Unity MR-Linac
    Yang, J.
    Vedam, S.
    Wang, J.
    Fuller, C.
    Choi, S.
    Chung, C.
    McAleer, M.
    Lee, B.
    Hughes, N.
    Ungchusri, G.
    Gillin, M.
    Martel, M.
    Balter, P.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E204 - E204