Characterization of on-site digital mammography systems: Direct versus indirect conversion detectors

被引:11
|
作者
Youn, Hanbean [1 ]
Han, Jong Chul [1 ]
Yun, Seungman [1 ]
Kam, Soohwa [1 ]
Cho, Seungryong [2 ]
Kim, Ho Kyung [1 ,3 ]
机构
[1] Pusan Natl Univ, Sch Mech Engn, Busan 609735, South Korea
[2] Korea Adv Inst Sci & Technol, Dept Nucl & Quantum Engn, Taejon 305701, South Korea
[3] Pusan Natl Univ, Ctr Adv Med Engn Res, Busan 609735, South Korea
基金
新加坡国家研究基金会;
关键词
Digital mammography; Cesium iodide; Amorphous selenium; MTF; NPS; DQE; Contrast; MODULATION TRANSFER-FUNCTION; PHYSICAL-CHARACTERISTICS; IMAGING PERFORMANCE; FILM MAMMOGRAPHY; RAY; SIGNAL; RESOLUTION; MTF;
D O I
10.3938/jkps.66.1926
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
We investigated the performances of two digital mammography systems. The systems use a cesium-iodide (CsI) scintillator and an amorphous selenium (a-Se) photoconductor for X-ray detection and are installed in the same hospital. As physical metrics, we measured the modulationtransfer function (MTF), the noise-power spectrum (NPS), and the detective quantum efficiency (DQE). In addition, we analyzed the contrast-detail performances of the two systems by using a commercial contrast-detail phantom. The CsI-based indirect conversion detector provided better MTF and DQE performances than the a-Se-based direct conversion detector whereas the former provided a poorer NPS performance than the latter. These results are explained by the fact that the CsI-based detector used an MTF restoration preprocessing algorithm. The a-Se-based detector showed better contrast-detail performance than the CsI-based detector. We believe that the highfrequency noise characteristic of a detector is more responsible for the visibility of small details than its spatial-resolution performance.
引用
下载
收藏
页码:1926 / 1935
页数:10
相关论文
共 50 条
  • [31] Characterization of Energy Consumers in Production Systems with Renewable On-Site Power Generation
    Schulz, Julia
    Rosenberg, Felix
    Scharmer, Valerie M.
    Zaeh, Michael F.
    ADVANCES IN PRODUCTION MANAGEMENT SYSTEMS: THE PATH TO DIGITAL TRANSFORMATION AND INNOVATION OF PRODUCTION MANAGEMENT SYSTEMS, PT I, 2020, 591 : 19 - 27
  • [32] New Digital Method of Direct and Indirect Control Applied to Wind Energy Conversion Systems using a Double Fed Induction Generator
    Gourma, Anass
    Berdai, Abdelmajid
    Reddak, Moussa
    Belfqih, Abdelaziz
    Elmariami, Faisal
    PROCEEDINGS OF 2016 INTERNATIONAL RENEWABLE & SUSTAINABLE ENERGY CONFERENCE (IRSEC' 16), 2016, : 548 - 553
  • [33] Comparison of a-Se direct.-conversion and CsI(Tl) indirect-conversion flat-panel digital detectors: A clinical assessment of image quality for general radiography applications
    Barski, Lori L.
    Wang, Xiaohui
    Wandtke, John
    Waldman, David
    Davis, Delphine
    Foos, David H.
    Dupin, Michael
    Huang, Weidong
    Yorkston, John
    MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
  • [34] Digital mammography: DQE versus optimized image quality in clinical environment-an on site study
    Oberhofer, Nadia
    Fracchetti, Alessandro
    Springeth, Margareth
    Moroder, Ehrenfried
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [35] Characterization of a direct, full-field flat-panel digital mammography detector
    Zhao, B
    Zhao, W
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 157 - 167
  • [36] Physical characterization of a scanning photon counting digital mammography system based on Si-strip detectors
    Aslund, Magnus
    Cederstrom, Bjorn
    Lundqvist, Mats
    Danielsson, Mats
    MEDICAL PHYSICS, 2007, 34 (06) : 1918 - 1925
  • [37] Image Quality of Digital Direct Flat-Panel Mammography Versus an Indirect Small-Field CCD Technique Using a High-Contrast Phantom
    Krug, Kathrin Barbara
    Stuetzer, Hartmut
    Frommolt, Peter
    Boecker, Julia
    Bovenschulte, Henning
    Sendler, Volker
    Lackner, Klaus
    INTERNATIONAL JOURNAL OF BREAST CANCER, 2011, 2011
  • [38] Characterization of polycrystalline lead oxide for application in direct conversion X-ray detectors
    Semeniuk, O.
    Grynko, O.
    Decrescenzo, G.
    Juska, G.
    Wang, K.
    Reznik, A.
    SCIENTIFIC REPORTS, 2017, 7
  • [39] Characterization of Thick Films of Cadmium Telluride for Direct Conversion X-ray Detectors
    Kim, Seo-Han
    Lee, Gun-Hwan
    Song, Pung-Keun
    SCIENCE OF ADVANCED MATERIALS, 2016, 8 (09) : 1799 - 1808
  • [40] Characterization of polycrystalline lead oxide for application in direct conversion X-ray detectors
    O. Semeniuk
    O. Grynko
    G. Decrescenzo
    G. Juska
    K. Wang
    A. Reznik
    Scientific Reports, 7