Characterization of on-site digital mammography systems: Direct versus indirect conversion detectors

被引:11
|
作者
Youn, Hanbean [1 ]
Han, Jong Chul [1 ]
Yun, Seungman [1 ]
Kam, Soohwa [1 ]
Cho, Seungryong [2 ]
Kim, Ho Kyung [1 ,3 ]
机构
[1] Pusan Natl Univ, Sch Mech Engn, Busan 609735, South Korea
[2] Korea Adv Inst Sci & Technol, Dept Nucl & Quantum Engn, Taejon 305701, South Korea
[3] Pusan Natl Univ, Ctr Adv Med Engn Res, Busan 609735, South Korea
基金
新加坡国家研究基金会;
关键词
Digital mammography; Cesium iodide; Amorphous selenium; MTF; NPS; DQE; Contrast; MODULATION TRANSFER-FUNCTION; PHYSICAL-CHARACTERISTICS; IMAGING PERFORMANCE; FILM MAMMOGRAPHY; RAY; SIGNAL; RESOLUTION; MTF;
D O I
10.3938/jkps.66.1926
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
We investigated the performances of two digital mammography systems. The systems use a cesium-iodide (CsI) scintillator and an amorphous selenium (a-Se) photoconductor for X-ray detection and are installed in the same hospital. As physical metrics, we measured the modulationtransfer function (MTF), the noise-power spectrum (NPS), and the detective quantum efficiency (DQE). In addition, we analyzed the contrast-detail performances of the two systems by using a commercial contrast-detail phantom. The CsI-based indirect conversion detector provided better MTF and DQE performances than the a-Se-based direct conversion detector whereas the former provided a poorer NPS performance than the latter. These results are explained by the fact that the CsI-based detector used an MTF restoration preprocessing algorithm. The a-Se-based detector showed better contrast-detail performance than the CsI-based detector. We believe that the highfrequency noise characteristic of a detector is more responsible for the visibility of small details than its spatial-resolution performance.
引用
下载
收藏
页码:1926 / 1935
页数:10
相关论文
共 50 条
  • [1] Characterization of on-site digital mammography systems: Direct versus indirect conversion detectors
    Hanbean Youn
    Jong Chul Han
    Seungman Yun
    Soohwa Kam
    Seungryong Cho
    Ho Kyung Kim
    Journal of the Korean Physical Society, 2015, 66 : 1926 - 1935
  • [2] Hybrid direct conversion detectors for digital mammography
    Yi, S
    Tümer, TO
    Maeding, D
    Mainprize, J
    Mawdsley, G
    Yaffe, MJ
    Hamilton, WJ
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (06) : 2093 - 2097
  • [3] Hybrid direct conversion detectors for digital mammography
    NOVA R&D Inc, Riverside, United States
    IEEE Trans Nucl Sci, 6 III (2093-2097):
  • [4] Direct conversion CdZnTe and CdTe detectors for digital mammography
    Yin, S
    Tümer, TO
    Maeding, D
    Mainprize, J
    Mawdsley, G
    Yaffe, MJ
    Gordon, EE
    Hamilton, WJ
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (01) : 176 - 181
  • [5] Direct conversion Si and CdZnTe detectors for digital mammography
    Yin, S
    Tümer, TO
    Maeding, D
    Mainprize, J
    Mawdsley, G
    Yaffe, MJ
    Gordon, EE
    Hamilton, WJ
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 448 (03): : 591 - 597
  • [6] Direct conversion detector for digital mammography
    Polischuk, B
    Rougeot, H
    Wong, K
    Debrie, A
    Poliquin, E
    Hansroul, M
    Martin, JP
    Truong, TT
    Choquette, M
    Laperrière, L
    Shukri, Z
    MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 : 417 - 425
  • [7] Analysis of the imaging performance in indirect digital mammography detectors by linear systems and signal detection models
    Liaparinos, P.
    Kalyvas, N.
    Kandarakis, I.
    Cavouras, D.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 697 : 87 - 98
  • [8] Optimization of operational conditions for direct digital mammography detectors for digital tomosynthesis
    Zhao, W
    Deych, R
    Dolazza, E
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 1272 - 1281
  • [9] Interpolation algorithms for digital mammography systems with multiple detectors
    Liu, H
    Wang, G
    Chen, JH
    Fajardo, LL
    ACADEMIC RADIOLOGY, 1999, 6 (03) : 170 - 175
  • [10] Image quality of direct conversion detectors for mammography and radiography: A theoretical comparison
    Mainprize, JG
    Yaffe, MJ
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 564 - 574