Tube Current Modulation Between Single- and Dual-Energy CT With a Second-Generation Dual-Source Scanner: Radiation Dose and Image Quality

被引:6
|
作者
Matsubara, Kosuke [1 ]
Takata, Tadanori [2 ]
Kobayashi, Masanao [3 ]
Kobayashi, Satoshi [1 ]
Koshida, Kichiro [1 ]
Gabata, Toshifumi [4 ]
机构
[1] Kanazawa Univ, Fac Hlth Sci, Inst Med Pharmaceut & Hlth Sci, Dept Quantum Med Technol, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 9200942, Japan
[2] Kanazawa Univ Hosp, Dept Radiol Technol, Kanazawa, Ishikawa, Japan
[3] Fujita Hlth Univ, Sch Hlth Sci, Fac Radiol Technol, Toyoake, Aichi, Japan
[4] Kanazawa Univ, Inst Med Pharmaceut & Hlth Sci, Dept Radiol, Fac Med, Kanazawa, Ishikawa, Japan
关键词
dosimetry; dual-energy CT; image noise; radiation dose; tube current modulation; AUTOMATIC EXPOSURE CONTROL; COMPUTED-TOMOGRAPHY; MULTIDETECTOR CT; PATIENT SIZE; ABDOMINAL CT; ITERATIVE RECONSTRUCTION; STONE COMPOSITION; BLOOD-VOLUME; ANGIOGRAPHY; REDUCTION;
D O I
10.2214/AJR.15.15404
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study was to compare the effects of tube current modulation between single- and dual-energy CT with a second-generation dual-source scanner. MATERIALS AND METHODS. Custom-made elliptic polymethylmethacrylate phantoms for slim and large patients were used. Absorbed radiation dose at the central point of the phantoms was measured with a solid-state detector while the phantoms were scanned in single-energy (120 kV) and dual-energy (100/Sn140, 80/Sn140, and 140/80 kV) modes with a second-generation dual-source CT scanner. Tube current modulation was activated in both modes, and quality reference tube current-time settings of 150, 300, 450, and 600 mAs were selected. Scanning was performed three times under the same conditions, and image noise was evaluated by measuring the SD of CT numbers in four separate regions of three adjacent images of the phantoms. RESULTS. Absorbed dose increased and image noise decreased with an increase in quality reference tube current-time setting when the slim phantom was scanned. For the large phantom, the radiation dose and noise level reached a plateau above quality reference tube current-time settings of 300 mAs for 100/Sn140 kV and 450 mAs for 120 kV. The radiation dose was small and the noise level was large with 80/Sn140 kV compared with that obtained with 120 and 100/Sn140 kV at all quality reference tube current-time settings. CONCLUSION. When a large phantom is scanned with 100/Sn140 kV, exposure demand for tube current modulation exceeds system limits at a lower quality reference tube current-time setting than for scanning 120 kV.
引用
下载
收藏
页码:354 / 361
页数:8
相关论文
共 50 条
  • [1] Single- and dual-energy CT pulmonary angiography using second- and third-generation dual-source CT systems: comparison of radiation dose and image quality
    Lenga, Lukas
    Trapp, Franziska
    Albrecht, Moritz H.
    Wichmann, Julian L.
    Johnson, Addison A.
    Yel, Ibrahim
    D'Angelo, Tommaso
    Booz, Christian
    Vogl, Thomas J.
    Martin, Simon S.
    EUROPEAN RADIOLOGY, 2019, 29 (09) : 4603 - 4612
  • [2] Single- and dual-energy CT pulmonary angiography using second- and third-generation dual-source CT systems: comparison of radiation dose and image quality
    Lukas Lenga
    Franziska Trapp
    Moritz H. Albrecht
    Julian L. Wichmann
    Addison A. Johnson
    Ibrahim Yel
    Tommaso D’Angelo
    Christian Booz
    Thomas J. Vogl
    Simon S. Martin
    European Radiology, 2019, 29 : 4603 - 4612
  • [3] Systematic radiation dose optimization of abdominal dual-energy CT on a second-generation dual-source CT scanner: assessment of the accuracy of iodine uptake measurement and image quality in an in vitro and in vivo investigations
    Schindera, Sebastian T.
    Zaehringer, Caroline
    D'Errico, Luigia
    Schwartz, Fides
    Kekelidze, Maka
    Szucs-Farkas, Zsolt
    Benz, Matthias R.
    ABDOMINAL RADIOLOGY, 2017, 42 (10) : 2562 - 2570
  • [4] Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT
    Wichmann, Julian L.
    Hardie, Andrew D.
    Schoepf, U. Joseph
    Felmly, Lloyd M.
    Perry, Jonathan D.
    Varga-Szemes, Akos
    Mangold, Stefanie
    Caruso, Damiano
    Canstein, Christian
    Vogl, Thomas J.
    De Cecco, Carlo N.
    EUROPEAN RADIOLOGY, 2017, 27 (02) : 642 - 650
  • [5] Systematic radiation dose optimization of abdominal dual-energy CT on a second-generation dual-source CT scanner: assessment of the accuracy of iodine uptake measurement and image quality in an in vitro and in vivo investigations
    Sebastian T. Schindera
    Caroline Zaehringer
    Luigia D’Errico
    Fides Schwartz
    Maka Kekelidze
    Zsolt Szucs-Farkas
    Matthias R. Benz
    Abdominal Radiology, 2017, 42 : 2562 - 2570
  • [6] Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CT
    Julian L. Wichmann
    Andrew D. Hardie
    U. Joseph Schoepf
    Lloyd M. Felmly
    Jonathan D. Perry
    Akos Varga-Szemes
    Stefanie Mangold
    Damiano Caruso
    Christian Canstein
    Thomas J. Vogl
    Carlo N. De Cecco
    European Radiology, 2017, 27 : 642 - 650
  • [7] Virtual monochromatic imaging in dual-source dual-energy CT: Radiation dose and image quality
    Yu, Lifeng
    Christner, Jodie A.
    Leng, Shuai
    Wang, Jia
    Fletcher, Joel G.
    McCollough, Cynthia H.
    MEDICAL PHYSICS, 2011, 38 (12) : 6371 - 6379
  • [8] Head and neck single- and dual-energy CT: differences in radiation dose and image quality of 2nd and 3rd generation dual-source CT
    Lenga, Lukas
    Lange, Marvin
    Martin, Simon S.
    Albrecht, Moritz H.
    Booz, Christian
    Yel, Ibrahim
    Arendt, Christophe T.
    Vogl, Thomas J.
    Leithner, Doris
    BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1122):
  • [9] Comparison of Radiation Dose and Image Quality of Contrast-Enhanced Dual-Source CT of the Chest: Single- Versus Dual-Energy and Second-Versus Third-Generation Technology
    Lenga, Lukas
    Leithnert, Doris
    Peterke, Julia L.
    Albrecht, Moritz H.
    Gudauskas, Tomas
    D'Angelo, Tommaso
    Booz, Christian
    Hammerstingl, Renate
    Vogl, Thomas J.
    Martin, Simon S.
    Wichmann, Julian L.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2019, 212 (04) : 741 - 747
  • [10] Trends in radiation dose and image quality for pediatric patients with a multidetector CT and a third-generation dual-source dual-energy CT
    Andrea Agostini
    Alberto Mari
    Cecilia Lanza
    Nicolo’ Schicchi
    Alessandra Borgheresi
    Stefania Maggi
    Andrea Giovagnoni
    La radiologia medica, 2019, 124 : 745 - 752