Failure analysis of a modular revision total HIP arthroplasty femoral stem fractured in vivo

被引:9
|
作者
Ortega, P. C. [1 ,2 ]
Medeiros, W. B., Jr. [1 ]
More, A. D. O. [1 ]
Vasconcelos, R. F. [1 ]
da Rosa, E. [1 ,2 ]
Roesler, C. R. M. [1 ,2 ]
机构
[1] Univ Fed Santa Catarina, Univ Hosp, Biomech Engn Lab LEBm, BR-88040900 Florianopolis, SC, Brazil
[2] Univ Fed Santa Catarina, Mech Engn Dept, BR-88040900 Florianopolis, SC, Brazil
关键词
Total hip arthroplasty revision; Modular femoral stem; Titanium alloy; Failure analysis; MECHANICAL-PROPERTIES; TITANIUM-ALLOYS; UNITED-STATES; MICROSTRUCTURE;
D O I
10.1016/j.engfailanal.2020.104591
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
An implant fracture is an unusual reason for total hip arthroplasty (THA) revision; however, it may cause serious clinical complications. The stress in femoral stems may increase considerably with the loss of bone support around the implant, leading to implant failure. This paper investigates the main reasons for the in vivo fracture of a Ti6Al4V revision femoral ZMR stem. Various methods were used: examination of the patient's by radiographies; optical emission spectrometry; and Scanning Electron Microscopy (SEM). The titanium (Ti) alloy was also characterized through optical microscopy and a Vickers hardness test was carried out. A finite element analysis (FEA) was conducted to evaluate the stress distribution occurring in the femoral stem revision under loading. The chemical composition and microstructure of the prosthesis did not reveal any non-conformity. The SEM analysis revealed fretting damage near the crack initiation site and the fractured surfaces exhibited characteristic ratchet marks, poorly formed striations and a small final fracture area. The FEA showed a stress concentration located at the proximal-distal taper junction, with a smaller diameter, coinciding with the crack initiation site. The conclusion was that fatigue failure was due to a lack of bone sustainability that increased the stress, associated with design characteristics that caused a localized concentration of stress, and these were potentialized by a fretting mechanism, leading to the failure.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Revision Total Hip Arthroplasty Using an Extensively Porous Coated Femoral Stem
    Moon, Kyoung Ho
    Kang, Joon Soon
    Lee, Sang Hyup
    Jung, Sae Rom
    CLINICS IN ORTHOPEDIC SURGERY, 2009, 1 (02) : 105 - 109
  • [32] Unique Failure Mechanism of a Femoral Component After Revision Total Hip Arthroplasty
    Martin, J. Ryan
    Trousdale, Robert T.
    ORTHOPEDICS, 2013, 36 (10) : E1327 - E1329
  • [33] Failure Following Revision Total Hip Arthroplasty After Cobalt-Chrome Femoral Heads are Placed on a Retained Femoral Stem
    Pagani, Nicholas R.
    Coden, Gloria S.
    Ramsden, David M.
    Zink, Thomas M.
    Ward, Daniel M.
    V. Bono, James
    Talmo, Carl T.
    JOURNAL OF ARTHROPLASTY, 2024, 39 (10): : 2569 - 2574
  • [35] Modular versus monoblock stem in revision total hip arthroplasty: a systematic review and meta-analysis
    Zampogna, Biagio
    Papalia, Giuseppe Francesco
    Parisi, Francesco Rosario
    Luciano, Claudia
    Zampoli, Andrea
    Vorini, Ferruccio
    Marongiu, Giuseppe
    Marinozzi, Andrea
    Farsetti, Pasquale
    Papalia, Rocco
    ANNALS OF JOINT, 2023, 8
  • [36] FEMORAL WINDOWS IN REVISION TOTAL HIP-ARTHROPLASTY
    KLEIN, AH
    RUBASH, HE
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1993, (291) : 164 - 170
  • [37] Revision total hip arthroplasty with cemented femoral component
    Nouri, Habib
    Kallel, Sofiene
    Slimane, Mohamed Hadj
    Meherzi, Mohamed Hedi
    Ouertatani, Moez
    Karray, Salahedine
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2008, 18 (05): : 327 - 332
  • [38] Revision total hip arthroplasty with cemented femoral component
    Habib Nouri
    Sofiene Kallel
    Mohamed Hadj Slimane
    Mohamed Hédi Meherzi
    Moez Ouertatani
    Salahedine Karray
    European Journal of Orthopaedic Surgery & Traumatology, 2008, 18 : 327 - 332
  • [39] Cortical Strut Allograft Support of Modular Femoral Junctions During Revision Total Hip Arthroplasty
    Lim, Chin Tat
    Amanatullah, Derek F.
    Huddleston, James I., III
    Hwang, Katherine L.
    Maloney, William J.
    Goodman, Stuart B.
    JOURNAL OF ARTHROPLASTY, 2017, 32 (05): : 1586 - 1592
  • [40] Mid-Term Results of Modular Tapered Femoral Stems in Revision Total Hip Arthroplasty
    Marfo, Kojo A.
    Berend, Keith R.
    Adams, Joanne B.
    Morris, Michael J.
    Lombardi, Adolph V., Jr.
    SURGICAL TECHNOLOGY INTERNATIONAL-INTERNATIONAL DEVELOPMENTS IN SURGERY AND SURGICAL RESEARCH, 2019, 35