Failure analysis of a modular revision total HIP arthroplasty femoral stem fractured in vivo

被引:9
|
作者
Ortega, P. C. [1 ,2 ]
Medeiros, W. B., Jr. [1 ]
More, A. D. O. [1 ]
Vasconcelos, R. F. [1 ]
da Rosa, E. [1 ,2 ]
Roesler, C. R. M. [1 ,2 ]
机构
[1] Univ Fed Santa Catarina, Univ Hosp, Biomech Engn Lab LEBm, BR-88040900 Florianopolis, SC, Brazil
[2] Univ Fed Santa Catarina, Mech Engn Dept, BR-88040900 Florianopolis, SC, Brazil
关键词
Total hip arthroplasty revision; Modular femoral stem; Titanium alloy; Failure analysis; MECHANICAL-PROPERTIES; TITANIUM-ALLOYS; UNITED-STATES; MICROSTRUCTURE;
D O I
10.1016/j.engfailanal.2020.104591
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
An implant fracture is an unusual reason for total hip arthroplasty (THA) revision; however, it may cause serious clinical complications. The stress in femoral stems may increase considerably with the loss of bone support around the implant, leading to implant failure. This paper investigates the main reasons for the in vivo fracture of a Ti6Al4V revision femoral ZMR stem. Various methods were used: examination of the patient's by radiographies; optical emission spectrometry; and Scanning Electron Microscopy (SEM). The titanium (Ti) alloy was also characterized through optical microscopy and a Vickers hardness test was carried out. A finite element analysis (FEA) was conducted to evaluate the stress distribution occurring in the femoral stem revision under loading. The chemical composition and microstructure of the prosthesis did not reveal any non-conformity. The SEM analysis revealed fretting damage near the crack initiation site and the fractured surfaces exhibited characteristic ratchet marks, poorly formed striations and a small final fracture area. The FEA showed a stress concentration located at the proximal-distal taper junction, with a smaller diameter, coinciding with the crack initiation site. The conclusion was that fatigue failure was due to a lack of bone sustainability that increased the stress, associated with design characteristics that caused a localized concentration of stress, and these were potentialized by a fretting mechanism, leading to the failure.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Outcome of a modular tapered uncemented titanium femoral stem in revision hip arthroplasty
    Maik Hoberg
    Christian Konrads
    Jana Engelien
    Dorothee Oschmann
    Michael Holder
    Matthias Walcher
    Maximilian Rudert
    International Orthopaedics, 2015, 39 : 1709 - 1713
  • [22] Outcome of a modular tapered uncemented titanium femoral stem in revision hip arthroplasty
    Hoberg, Maik
    Konrads, Christian
    Engelien, Jana
    Oschmann, Dorothee
    Holder, Michael
    Walcher, Matthias
    Rudert, Maximilian
    INTERNATIONAL ORTHOPAEDICS, 2015, 39 (09) : 1709 - 1713
  • [23] Long-term Outcomes of Revision Total Hip Arthroplasty Using a Modular Fluted Conical Femoral Stem
    Wallace, David T.
    Ohly, Nicholas E.
    Allen, David J.
    Baines, Joe
    JOURNAL OF ARTHROPLASTY, 2024, 39 (04): : 1048 - 1053
  • [24] Femoral Fixation in Revision Total Hip Arthroplasty
    Hartman, Curtis W.
    Garvin, Kevin L.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2011, 93A (24): : 2311 - 2322
  • [25] Modular revision stems in total hip arthroplasty
    Jones, RE
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (420) : 142 - 147
  • [26] Femoral Component Revision of Total Hip Arthroplasty
    Brown, Jonathon M.
    Mistry, Jaydev B.
    Cherian, Jeffrey J.
    Elmallah, Randa K.
    Chughtai, Morad
    Harwin, Steven F.
    Mont, Michael A.
    ORTHOPEDICS, 2016, 39 (06) : E1129 - E1139
  • [27] Failure of the stem-condyle junction of a modular femoral stem in revision total knee arthroplasty - A report of five cases
    Lim, LA
    Trousdale, RT
    Berry, DJ
    Hanssen, AD
    JOURNAL OF ARTHROPLASTY, 2001, 16 (01): : 128 - 132
  • [28] Femoral revision total hip arthroplasty using Wagner stem with allograft bone
    Takahira, N
    Uchiyama, K
    Komiya, K
    Takasaki, S
    Ohkawa, T
    Itoman, M
    OSTEOPOROSIS INTERNATIONAL, 2003, 14 : S56 - S56
  • [29] New Approach and Stem Increased Femoral Revision Rate in Total Hip Arthroplasty
    Panichkul, Phonthakorn
    Parks, Nancy L.
    Ho, Henry
    Hopper, Robert H., Jr.
    Hamilton, William G.
    ORTHOPEDICS, 2016, 39 (01) : E86 - E92
  • [30] REVISION TOTAL HIP ARTHROPLASTY IN PATIENTS WITH ASEPTIC LOOSENING OF FEMORAL STEM (REVIEW)
    Sementkovsky, A. V.