Biorealistic Control of Hand Prosthesis Augments Functional Performance of Individuals With Amputation

被引:10
|
作者
Luo, Qi [1 ]
Niu, Chuanxin M. [1 ,2 ,3 ]
Chou, Chih-Hong [1 ,2 ]
Liang, Wenyuan [4 ]
Deng, Xiaoqian [5 ]
Hao, Manzhao [1 ,2 ]
Lan, Ning [1 ,2 ]
机构
[1] Shanghai Jiao Tong Univ, Sch Biomed Engn, Lab Neurorehabil Engn, Shanghai, Peoples R China
[2] Shanghai Jiao Tong Univ, Inst Med Robot, Shanghai, Peoples R China
[3] Shanghai Jiao Tong Univ, Sch Med, Ruijin Hosp, Dept Rehabil Med, Shanghai, Peoples R China
[4] Natl Res Ctr Rehabil Tech Aids, Beijing, Peoples R China
[5] Guangdong Work Injury Rehabil Hosp, Guangzhou, Peoples R China
基金
国家重点研发计划;
关键词
biorealistic control; neuromuscular reflex; neuromorphic computation; tendon-driven prosthesis; hand grasp; UPPER-LIMB LOSS; MECHANICAL MODEL; MUSCLE-STIFFNESS; TENDON; FEEDBACK; REFLEX; INFORMATION; VALIDATION; BOX;
D O I
10.3389/fnins.2021.783505
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
The human hand has compliant properties arising from muscle biomechanics and neural reflexes, which are absent in conventional prosthetic hands. We recently proved the feasibility to restore neuromuscular reflex control (NRC) to prosthetic hands using real-time computing neuromorphic chips. Here we show that restored NRC augments the ability of individuals with forearm amputation to complete grasping tasks, including standard Box and Blocks Test (BBT), Golf Balls Test (GBT), and Potato Chips Test (PCT). The latter two were more challenging, but novel to prosthesis tests. Performance of a biorealistic controller (BC) with restored NRC was compared to that of a proportional linear feedback (PLF) controller. Eleven individuals with forearm amputation were divided into two groups: one with experience of myocontrol of a prosthetic hand and another without any. Controller performances were evaluated by success rate, failure (drop/break) rate in each grasping task. In controller property tests, biorealistic control achieved a better compliant property with a 23.2% wider range of stiffness adjustment than that of PLF control. In functional grasping tests, participants could control prosthetic hands more rapidly and steadily with neuromuscular reflex. For participants with myocontrol experience, biorealistic control yielded 20.4, 39.4, and 195.2% improvements in BBT, GBT, and PCT, respectively, compared to PLF control. Interestingly, greater improvements were achieved by participants without any myocontrol experience for BBT, GBT, and PCT at 27.4, 48.9, and 344.3%, respectively. The functional gain of biorealistic control over conventional control was more dramatic in more difficult grasp tasks of GBT and PCT, demonstrating the advantage of NRC. Results support the hypothesis that restoring neuromuscular reflex in hand prosthesis can improve neural motor compatibility to human sensorimotor system, hence enabling individuals with amputation to perform delicate grasps that are not tested with conventional prosthetic hands.
引用
收藏
页数:17
相关论文
共 50 条
  • [11] Residual limb strength and functional performance measures in individuals with unilateral transtibial amputation
    Ihmels, Wyatt D.
    Miller, Ross H.
    Esposito, E. Russell
    GAIT & POSTURE, 2022, 97 : 159 - 164
  • [12] On the Design of a Novel Underactuated Robotic Finger Prosthesis for Partial Hand Amputation
    Jeong, Gu-Cheol
    Kim, Yeoeun
    Choi, Wooseong
    Gu, Gangyong
    Lee, Hyun-Joo
    Hong, Man Bok
    Kim, Keehoon
    2019 IEEE 16TH INTERNATIONAL CONFERENCE ON REHABILITATION ROBOTICS (ICORR), 2019, : 861 - 867
  • [13] Prosthesis rejection in individuals with limb amputation: a narrative review with respect to rehabilitation
    Piscitelli, Daniele
    Beghi, Massimiliano
    Bigoni, Marco
    Diotti, Silvia
    Perin, Cecilia
    Peroni, Federica
    Turati, Marco
    Zanchi, Nicolo
    Mazzucchelli, Myriam
    Cornaggia, Cesare Maria
    RIVISTA DI PSICHIATRIA, 2021, 56 (04) : 175 - 181
  • [14] Functional rehabilitation after transfemoral amputation Shaft prosthesis or endo-exo prosthesis?
    Richter, Katherina
    Krause, Katharina
    Rotter, Robert
    Fischer, Dagmar-C
    Aschoff, Horst-H
    Mittlmeier, Thomas
    UNFALLCHIRURG, 2022, 125 (04): : 266 - 274
  • [15] SENSORY CONTROL OF A MULTIFUNCTION HAND PROSTHESIS
    KYBERD, PJ
    CHAPPELL, PH
    NIGHTINGALE, JM
    BIOSENSORS, 1988, 3 (06) : 347 - 357
  • [16] PREHENSILE CONTROL OF A HAND PROSTHESIS BY A MICROCONTROLLER
    CHAPPELL, PH
    KYBERD, PJ
    JOURNAL OF BIOMEDICAL ENGINEERING, 1991, 13 (05): : 363 - 369
  • [17] ADAPTIVE-CONTROL OF A HAND PROSTHESIS
    SWAIN, ID
    INTERNATIONAL JOURNAL OF REHABILITATION RESEARCH, 1980, 3 (04) : 620 - 620
  • [18] A Noninvasive Neuroprosthesis Augments Hand Grasp Force in Individuals with Cervical Spinal Cord Injury: The Functional and Therapeutic Effects
    Thorsen, Rune
    Costa, Davide Dalla
    Chiaramonte, Sara
    Binda, Luca
    Beghi, Ettore
    Redaelli, Tiziana
    Occhi, Eugenio
    Ferrarin, Maurizio
    SCIENTIFIC WORLD JOURNAL, 2013,
  • [19] ADAPTIVE MULTI-FUNCTIONAL HAND PROSTHESIS
    CODD, RD
    NIGHTINGALE, JM
    TODD, RW
    JOURNAL OF PHYSIOLOGY-LONDON, 1973, 232 (02): : P55 - P56
  • [20] Hybrid Tongue-Myoelectric Control Improves Functional Use of a Robotic Hand Prosthesis
    Johansen, Daniel
    Popovic, Dejan B.
    Dosen, Strahinja
    Struijk, Lotte N. S. Andreasen
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2021, 68 (06) : 2011 - 2020