Retention prediction of highly polar ionizable solutes under gradient conditions on a mixed-mode reversed-phase and weak anion-exchange stationary phase

被引:10
|
作者
Balkatzopoulou, P. [1 ]
Fasoula, S. [1 ]
Gika, H. [2 ]
Nikitas, P. [1 ]
Pappa-Louisi, A. [1 ]
机构
[1] Aristotle Univ Thessaloniki, Dept Chem, Thessaloniki 54124, Greece
[2] Aristotle Univ Thessaloniki, Dept Chem Engn, Thessaloniki 54124, Greece
关键词
Mixed-mode stationary phase; RP and HILIC gradients; Modeling dual gradient elution; TANDEM MASS-SPECTROMETRY; PERFORMANCE LIQUID-CHROMATOGRAPHY; HUMAN URINE; SEPARATION; ELUTION; METABOLITES; OPTIMIZATION; COLUMNS; PH;
D O I
10.1016/j.chroma.2015.03.082
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In the present work the retention of three highly polar and ionizable solutes - uric acid, nicotinic acid and ascorbic acid - was investigated on a mixed-mode reversed-phase and weak anion-exchange (RP/WAX) stationary phase in buffered aqueous acetonitrile (ACN) mobile phases. A U-shaped retention behavior was observed for all solutes with respect to the eluent organic modifier content studied in a range of 5-95% (v/v). This retention behavior clearly demonstrates the presence of a HILIC-type retention mechanism at ACN-rich hydro-organic eluents and an RP-like retention at aqueous-rich hydro-organic eluents. Hence, this column should be promising for application under both RP and HILIC gradient elution modes. For this reason, a series of programmed elution runs were carried out with increasing (RP) and decreasing (HILIC) organic solvent concentration in the mobile phase. This dual gradient process was successfully modeled by two retention models exhibiting a quadratic or a cubic dependence of the logarithm of the solute retention factor (In k) upon the organic modifier volume fraction (phi). It was found that both models produced by gradient retention data allow the prediction of solute retention times for both types of programmed elution on the mixed-mode column. Four, in the case of the quadratic model, or five, in the case of the cubic model, initial HILIC- and RP-type gradient runs gave satisfactory retention predictions of any similar kind elution program, even with different flow rate, with an overall error of only 2.5 or 1.7%, respectively. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:72 / 76
页数:5
相关论文
共 50 条
  • [41] The secret of reversed-phase/weak cation exchange retention mechanisms in mixed-mode liquid chromatography applied for small drug molecule analysis
    Svrkota, Bojana
    Krmar, Jovana
    Protic, Ana
    Otasevic, Biljana
    JOURNAL OF CHROMATOGRAPHY A, 2023, 1690
  • [42] A reversed phase/hydrophilic interaction/ion exchange mixed-mode stationary phase for liquid chromatography
    Sainan Zhang
    Feifang Zhang
    Bingcheng Yang
    Xinmiao Liang
    Chinese Chemical Letters, 2019, 30 (02) : 470 - 472
  • [43] A reversed phase/hydrophilic interaction/ion exchange mixed-mode stationary phase for liquid chromatography
    Zhang, Sainan
    Zhang, Feifang
    Yang, Bingcheng
    Liang, Xinmiao
    CHINESE CHEMICAL LETTERS, 2019, 30 (02) : 470 - 472
  • [44] Mixed-phase weak anion-exchange/reversed-phase LC-MS/MS for analysis of nucleotide sugars in human fibroblasts
    Rahm, Moritz
    Kwast, Hanneke
    Wessels, Hans J. C. T.
    Noga, Marek J.
    Lefeber, Dirk J.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (15) : 3595 - 3604
  • [45] Study of the mechanism of mixed-mode retention of a fluorescent dye by reversed-phase liquid chromatography
    Anthony, SR
    Wirth, MJ
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2003, 225 : U113 - U113
  • [46] Liquid chromatography and characterization of ether-functionalized imidazolium ionic liquids on mixed-mode reversed-phase/cation exchange stationary phase
    Leicunaite, Jelena
    Klimenkovs, Igors
    Kviesis, Jorens
    Zacs, Dzintars
    Kreismanis, Juris Pauls
    COMPTES RENDUS CHIMIE, 2010, 13 (10) : 1335 - 1340
  • [47] Flow microcalorimetric measurements for bovine serum albumin on reversed-phase and anion-exchange supports under overloaded conditions
    Thrash, ME
    Pinto, NG
    JOURNAL OF CHROMATOGRAPHY A, 2001, 908 (1-2) : 293 - 299
  • [48] Preparation of a monolithic column with a mixed-mode stationary phase of reversed-phase/hydrophilic interaction for capillary liquid chromatography
    Zhang, Meng
    Gong, Li
    Liu, Guizhen
    Kang, Jingwu
    JOURNAL OF SEPARATION SCIENCE, 2019, 42 (03) : 662 - 669
  • [49] Synthesis, chromatographic evaluation and hydrophilic interaction/reversed-phase mixed-mode behavior of a "Click β-cyclodextrin" stationary phase
    Guo, Zhimou
    Jin, Yu
    Liang, Tu
    Liu, Yanfang
    Xu, Qing
    Liang, Xinmiao
    Lei, Aiwen
    JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (02) : 257 - 263
  • [50] Fast separation of pyrimidine derivatives by capillary electrochromatography on ion-exchange/reversed-phase mixed-mode stationary phases
    Klampfl, CW
    Buchberger, W
    Haddad, PR
    JOURNAL OF CHROMATOGRAPHY A, 2001, 911 (02) : 277 - 283