A comparison on three neutron-gamma discrimination methods used with NE213, UGLLT and UGAB scintillators

被引:4
|
作者
Lotfi, Y. [1 ]
Moussavi-Zarandi, S. A. [1 ]
Ghal-Eh, N. [2 ,3 ]
Bayat, E. [4 ]
机构
[1] Amirkabir Univ Technol, Dept Energy Engn & Phys, Tehran 158754413, Iran
[2] Damghan Univ, Sch Phys, Damghan 3671641167, Iran
[3] Ferdowsi Univ Mashhad, Fac Sci, Dept Phys, Mashhad 917751436, Razavi Khorasan, Iran
[4] AEOI, Nucl Sci & Technol Res Ctr, Tehran 141551339, Iran
关键词
Liquid scintillator; NE213; UGLLT; UGAB; FoM; Zero-cross method; Charge integration method; Filtered charge integration method;
D O I
10.1016/j.radphyschem.2020.108701
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Liquid organic scintillators are widely used in radiation measurements. NE213 (or its equivalent, BC-501A) in neutron-gamma discrimination (NGD) and two new scintillators, UGLLT and UGAB, in alpha-beta discrimination are two common examples. In this paper, the NGD capabilities of these three scintillators were studied and their figure-of-merits (FoMs) were measured for comparison. The three different NGD methods used in this study were zero-cross method (ZCM), charge integration method (CIM) and filtered charge integration method (FCIM), where ZCM was analog and the two other methods were digital. A 1.1 mu Ci point Na-22 gamma ray source and 100 mCi cylindrical Am-241-Be neutron source (3.5 cm diameter by 5 cm length) were used for the energy calibration and the NGD, respectively. The FoM values for NE213, UGLLT and UGAB at 100 keVee discrimination bias (i.e., the preset threshold voltage, generally expressed in keVee unit, upon which the generated pulse-heights are recorded) using ZCM, CIM and FCIM methods were (1.060 +/- 0.037, 1.430 +/- 0.040, 1.70 +/- 0.054), (0.910 +/- 0.029, 1.270 +/- 0.018, 1.460 +/- 0.027) and (0.890 +/- 0.027, 1.200 +/- 0.020, 1.460 +/- 0.026), respectively. Although UGAB scintillator exhibits the least discrimination quality, one can choose an appropriate neutrongamma discrimination method to improve its FoM. The low-temperature flammability, toxicity and high-evaporation properties of NE213 scintillator may motivate some researchers to use UGLLT and UGAB scintillators as alternatives to NE213, however, by selecting the most efficient discrimination method.
引用
收藏
页数:8
相关论文
共 43 条
  • [11] Triplet Harvesting Plastic Scintillators with Neutron-Gamma Pulse Shape Discrimination
    van Loef, Edgar V.
    Feng, Patrick
    Markosyan, Gary
    Shirwadkar, Urmila
    Doty, Patrick
    Shah, Kanai S.
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS XVI, 2014, 9213
  • [12] Fast neutron-gamma discrimination in organic scintillators via convolution neural network
    Seonkwang Yoon
    Chaehun Lee
    Byung-Hee Won
    Sang-Bum Hong
    Hee Seo
    Ho-Dong Kim
    Journal of the Korean Physical Society, 2022, 80 : 427 - 433
  • [13] Fast neutron-gamma discrimination in organic scintillators via convolution neural network
    Yoon, Seonkwang
    Lee, Chaehun
    Won, Byung-Hee
    Hong, Sang-Bum
    Seo, Hee
    Kim, Ho-Dong
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2022, 80 (05) : 427 - 433
  • [14] NEUTRON DETECTION EFFICIENCY FOR NE213 AND BC501 SCINTILLATORS AT ENERGIES BETWEEN 25 AND 200 MEV
    BYRD, RC
    SAILOR, WC
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 274 (03): : 494 - 500
  • [15] THE LIGHT YIELD RESPONSE OF NE213 ORGANIC SCINTILLATORS TO CHARGED-PARTICLES RESULTING FROM NEUTRON INTERACTIONS
    MOUATASSIM, S
    COSTA, GJ
    GUILLAUME, G
    HEUSCH, B
    HUCK, A
    MOSZYNSKI, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 359 (03): : 530 - 536
  • [16] Digital Neutron-Gamma Discrimination Methods: Charge Comparison versus Zero-Crossing
    Szczcsniak, T.
    Grodzicka, M.
    Moszynski, M.
    Wolski, D.
    Swiderski, L.
    Szawlowski, M.
    Schotanus, P.
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [17] Neutron-Gamma Pulse Shape Discrimination Using Organic Scintillators With Silicon Photomultiplier Readout
    Preston, Rhys M.
    Eberhardt, John E.
    Tickner, James R.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2014, 61 (04) : 2410 - 2418
  • [18] Neutron measurements on Joint European Torus using an NE213 scintillator with digital pulse shape discrimination
    Esposito, B
    Bertalot, L
    Marocco, D
    Riva, M
    Kaschuck, Y
    Skopintsev, S
    Zimbal, A
    Reginatto, M
    Schuhmacher, H
    Adams, JM
    Murari, A
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10): : 3550 - 3552
  • [19] NEUTRON-GAMMA PULSE SHAPE DISCRIMINATION WITH A NE-213 LIQUID SCINTILLATOR BY USING DIGITAL SIGNAL PROCESSING COMBINED WITH SIMILARITY METHOD
    Mardiyanto
    ATOM INDONESIA, 2008, 34 (02) : 103 - 115
  • [20] Optimization of digital neutron-gamma discrimination methods in time and frequency domain
    Singh, Harleen
    Mehra, and Rohit
    JOURNAL OF INSTRUMENTATION, 2024, 19 (02):