Automated antinuclear immunofluorescence antibody analysis is a reliable approach in routine clinical laboratories

被引:11
|
作者
Zheng, Bing [1 ]
Li, Enling [1 ]
Zhu, Haoming [1 ]
Lu, Jingbo [1 ]
Shi, Xinming [1 ]
Zhang, Jie [1 ]
Li, Min [1 ]
机构
[1] Shanghai Jiao Tong Univ, Dept Lab Med, Renji Hosp, Sch Med, 160 Pujian Rd, Shanghai 200127, Peoples R China
基金
中国国家自然科学基金;
关键词
antinuclear antibodies (ANAs); automation; indirect immunofluorescence; value-added ANA report; AUTOANTIBODIES; GUIDELINES; DIAGNOSIS; TESTS;
D O I
10.1515/cclm-2017-0050
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Indirect immunofluorescence (IIF) assays are recommended as the gold standard method for the detection of antinuclear antibodies (ANAs). This study aimed to investigate the reliability of an automated system. Methods: We compared 3745 serum samples using NOVA View archived images with manual analysis via microscopy. A custom cutoff value was established to distinguish ANA titers and was validated in two clinical laboratories. The automatic ANA pattern recognition system was evaluated, and all ANA-positive sera were subjected to two commercial ANA IIF kits to compare the consistency of the pattern interpretation results. For inconsistent patterns, a third ANA IIF testing kit was utilized. Results: Agreement of the interpretation of the ANA IIF test using the platform of NOVA View and manual microscopy was 96.9%. The local cutoff value to discriminate ANA titers in four main ANA patterns was calculated based on 1390 serum samples. In our laboratory, the titer prediction accuracy was superior to the preset cutoff in NOVA View (p < 0.01); the performance was similar in another laboratory (p = 0.11). The automatic pattern recognition accuracies of speckled, homogeneous, centromere, nucleolar and nuclear dot patterns were 62.7%, 57.4%, 92.6%, 30.5% and 27.3%, respectively. The consistency of the pattern interpretation results between INOVA and MBL kits was 95.3%. Conclusions: It is necessary to establish a custom value-added ANA report. However, confirmation of the digital immunofluorescence images by expert technicians was essential, and suspect results of an ANA pattern should be reconfirmed by another commercial ANA IIF kit to achieve more reliable results.
引用
收藏
页码:1922 / 1930
页数:9
相关论文
共 50 条
  • [21] Comparison of the analytical and clinical performances of two different routine testing protocols for antinuclear antibody screening
    Gonzalez Rodriguez, Concepcion
    Fuentes Cantero, Sandra
    Perez Perez, Antonio
    Vazquez Barbero, Francisco Javier
    Leon Justel, Antonio
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2021, 35 (09)
  • [22] ANALYSIS OF ANTINUCLEAR ANTIBODY TITERS AND PATTERNS USING HEP 2 INDIRECT IMMUNOFLUORESCENCE IN VARIOUS LIVER DISEASES
    Wei, Q.
    Jiang, Y.
    Yang, M.
    Xie, J.
    Xiao, M.
    Gu, J.
    ANNALS OF THE RHEUMATIC DISEASES, 2020, 79 : 1838 - 1838
  • [23] An indirect comparison of automated indirect immunofluorescence vs automated solid-phase immunoassays for antinuclear antibody detection: A meta-analysis and adjusted indirect comparison of diagnostic test accuracy
    Zhang, Minjie
    Zhang, Xiaoyan
    INTERNATIONAL JOURNAL OF RHEUMATIC DISEASES, 2022, 25 (09) : 991 - 1002
  • [24] THE PREVALENCE OF DENSE FINE SPECKLED PATTERN IN ROUTINE SCREENING FOR SYSTEMIC AUTOIMMUNE DISEASES USING INDIRECT IMMUNOFLUORESCENCE-ANTINUCLEAR ANTIBODY TEST
    Kim, S.
    Jeon, Y. L.
    Kim, M. H.
    Lee, W. I.
    Kang, S. Y.
    ANNALS OF THE RHEUMATIC DISEASES, 2017, 76 : 1425 - 1425
  • [25] Characteristics of ELISA positive and immunofluorescence negative anti-dsDNA antibody in a routine clinical laboratory
    Kim, K
    Lee, S
    Chung, W
    Kim, B
    ANNALS OF THE RHEUMATIC DISEASES, 2005, 64 : 373 - +
  • [26] A comparative study of tissue transglutaminase antibodies and endomysium antibody immunofluorescence in routine clinical laboratory practice
    Sinclair, D
    Pearce, CB
    Saas, MSL
    Poller, D
    ANNALS OF CLINICAL BIOCHEMISTRY, 2003, 40 : 411 - 416
  • [27] Sigma and Risk in the Quality Control Routine: Analysis in Chilean Clinical Laboratories
    Carboni-Huerta, Roberto
    Saenz-Flor, Klever, V
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2022, 7 (02): : 456 - 466
  • [28] Harmonization of clinical interpretation of antinuclear antibody test results by solid phase assay and by indirect immunofluorescence through likelihood ratios
    Bossuyt, Xavier
    Claessens, Jolien
    Belmondo, Thibaut
    De Langhe, Ellen
    Westhovens, Rene
    Poesen, Koen
    Hue, Sophie
    Blockmans, Daniel
    Fritzler, Marvin J.
    Mahler, Michael
    Fierz, Walter
    AUTOIMMUNITY REVIEWS, 2019, 18 (11)
  • [29] Prevalence and clinical characteristics of the dense fine speckled pattern: Indirect immunofluorescence-antinuclear antibody screening in the Chinese population
    Meng, Jingjing
    Wang, Rong
    Luo, Yueming
    Li, Siting
    Bai, Yina
    Song, Ning
    Li, Mengtao
    Zeng, Xiaofeng
    Hu, Chaojun
    SCANDINAVIAN JOURNAL OF IMMUNOLOGY, 2023, 97 (01)
  • [30] Are automated topographic analyses of visual-evoked potentials reliable and feasible in clinical routine?
    Horster, Henrik
    Hardmeier, Martin
    Hatz, Florian
    Fuhr, Peter
    NEUROLOGY, 2017, 88