摘要
Background/purpose: This study evaluated the torsional resistances of nickel-titanium rotary instruments by two methods: i) dynamic resistance: repetitive torsional loading (RTL) and ii) static resistance: single torsional loading (STL) based on the International Organization for Standardization. Materials and methods: In RTL method, a pre-set rotational loading (0.5 N cm) was applied after fixing the 3-mm tip of the file, and this clockwise loading to the pre-set torque and counterclockwise unloading to original position was repeated at 50 rpm until the file fractured. The number of repetition cycles until fracture was counted. In STL method, the torsional strength was determined by continuous clockwise rotation (2 rpm) until fracture. Results from both methods were compared after testing the torsional resistances of four instrument systems (n = 15): Hyflex CM, HyFlex EDM, V-Taper2, and V-Taper2H. A scanning electron microscope (SEM) was used to examine the topographic features of the fractured surfaces and longitudinal aspects (n = 5) from both methods. Results: The RTL and STL methods had similar results: V-Taper2 had the highest resistance and the Hyflex CMhad the lowest (P < 0.05). Spearman correlation test showed the results from two methods were strongly correlated (coefficient = 1). Under the SEM, specimens from the RTLshowed ruptured aspects on cross-sections with multiple areas of crack propagation, while the STLshowed the typical features of torsional failure such as circular abrasion marks and fatigue dimples. Conclusions: This study suggests the clinically relevant torsional test (RTL) method yield similar results with the STL method, but they have different topographic findings. (C) 2018 Association for Dental Sciences of the Republic of China. Publishing services by Elsevier B.V.
引用
收藏
页码:207 / 212
页数:6
相关论文
共 50 条
  • [41] Mechanical and Metallurgical Properties of Various Nickel-Titanium Rotary Instruments
    Shim, Kyu-Sang
    Oh, Soram
    Kum, KeeYeon
    Kim, Yu-Chan
    Jee, Kwang-Koo
    Chang, Seok Woo
    BIOMED RESEARCH INTERNATIONAL, 2017, 2017
  • [42] A Review of Cyclic Fatigue Testing of Nickel-Titanium Rotary Instruments
    Plotino, Gianluca
    Grande, Nicola M.
    Cordaro, Massimo
    Testarelli, Luca
    Gambarini, Gianluca
    JOURNAL OF ENDODONTICS, 2009, 35 (11) : 1469 - 1476
  • [43] Cyclic fatigue of endodontic nickel titanium rotary instruments: Static and dynamic tests
    Li, UM
    Lee, BS
    Shih, CT
    Lan, WH
    Lin, CP
    JOURNAL OF ENDODONTICS, 2002, 28 (06) : 448 - 451
  • [44] Evaluation of surface characteristics of rotary nickel-titanium instruments produced by different manufacturing methods
    Inan, U.
    Gurel, M.
    NIGERIAN JOURNAL OF CLINICAL PRACTICE, 2017, 20 (02) : 143 - 146
  • [45] Evaluation of Cyclic Flexural Fatigue of M-Wire Nickel-Titanium Rotary Instruments
    Al-Hadlaq, Solaiman M. S.
    AlJarbou, Fahad A.
    AlThumairy, Riyadh I.
    JOURNAL OF ENDODONTICS, 2010, 36 (02) : 305 - 307
  • [46] Static and Dynamic Cyclic Fatigue Resistance of Nickel-Titanium Rotary Instruments in a Double-Curved Stainless Steel Artificial Canal
    Le, Hoang-Lan-Anh
    Tran, Thuan-Loc
    Nguyen, Thu-Thuy
    Pham, Tran-Lan-Khue
    Pham, Van-Khoa
    APPLIED SCIENCES-BASEL, 2023, 13 (04):
  • [47] Effect of sterilization on torsional fracture resistance of nickel-titanium rotary files
    Kiss, EP
    Murchison, DF
    Davis, BD
    JOURNAL OF DENTAL RESEARCH, 1997, 76 : 547 - 547
  • [48] Effect of electropolishing ProFile nickel-titanium rotary instruments on cyclic fatigue resistance, torsional resistance, and cutting efficiency
    Bui, Tung B.
    Mitchell, John C.
    Baumgartner, J. Craig
    JOURNAL OF ENDODONTICS, 2008, 34 (02) : 190 - 193
  • [49] Torsional resistance of XP-endo Shaper at body temperature compared with several nickel-titanium rotary instruments
    Elnaghy, A. M.
    Elsaka, S. E.
    INTERNATIONAL ENDODONTIC JOURNAL, 2018, 51 (05) : 572 - 576
  • [50] Evaluation of stress distribution in nickel-titanium rotary instruments with different geometrical designs subjected to bending and torsional load: a finite element study
    Manar Galal
    Tamer M. Hamdy
    Bulletin of the National Research Centre, 44 (1)