Development of a long trace profiler in situ at National Synchrotron Radiation Research Center

被引:3
|
作者
Lin, Shang-Wei [1 ]
Wang, Duan-Jen [1 ]
Fu, Huang-Wen [1 ]
Tsai, Huang-Ming [1 ]
Hua, Chih-Yu [1 ]
Kuo, Chang-Yang [1 ,2 ]
Hsu, Ming-Ying [1 ]
Kao, Kai-Yang [1 ]
Yin, Gung-Chian [1 ]
Fung, Hok-Sum [1 ]
Perng, Shen-Yaw [1 ]
Chang, Chia-Feng [1 ]
机构
[1] Natl Synchrotron Radiat Res Ctr, Hsinchu 30076, Taiwan
[2] Max Planck Inst Chem Phys Solids, Nothnitzer Str 40, D-01187 Dresden, Germany
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2019年 / 90卷 / 02期
关键词
DISTORTION PROFILE; MIRROR;
D O I
10.1063/1.5055634
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
To achieve an ultrahigh resolution of a beamline for soft X-rays at the Taiwan Photon Source (TPS), the profile of a highly precise grating is required at various curvatures. The slope error could be decreased to 0.1 mu rad (rms) at a thermal load with a specially designed bender having 25 actuators. In the meantime, a long-trace profiler (LTP) was developed in situ to monitor the grating profile under a thermal load; it consists of a moving optical head, an air-bearing slide, an adjustable stand, and a glass viewport on the vacuum chamber. In the design of this system, a test chamber with an interior mirror was designed to simulate the chamber in the beamline. To prevent an error induced from a commercial viewport, a precision glass viewport (150CF, flatness 1/150 lambda rms at 632.8 nm) was designed. The error induced from the slope error of the glass surface and the vacuum deformation was also simulated. The performance of the optical head of the LTP in situ (ISLTP) has been tested in the metrology laboratory. The sources of error of this LTP including the linearity and the glass viewport were corrected after the measurement. For the beamline measurement, an optical head was mounted outside the vacuum chamber; the measuring beam passed through the glass viewport to measure the grating profile in vacuum. The measurement of the LTP after correction of the above errors yielded a precision about 0.2 mu rad (rms). In a preliminary test, an ISLTP was used to measure the grating profile at soft X-ray beamline TPS45A. The measured profile was for the bending mechanism to optimize the slope profile. From the measured energy spectrum, the slope error of the grating was estimated with software for optical simulation to be about 0.3 mu rad (rms), consistent with our estimate of the ISLTP. In the future, it will be used to monitor the thermal bump under a large thermal load. In addition, an ISLTP was used to monitor the properties of optical elements-the twist and radius in the beamline during the installation phase. Published under license by AIP Publishing.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] The highly efficient protein crystallography beamline TLS 13B1 at the National Synchrotron Radiation Research Center
    Chen, Yi-Hui
    Tseng, Chien-Chang
    Chou, Chung-Kuang
    Liu, Yi-Chun
    Chiang, Cheng-Hung
    Huang, Chen-Ying
    Chao, Chun-Hsiung
    Huang, Chun-Hsiang
    JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 2023, 70 (05) : 1219 - 1227
  • [32] Vacuum system developments at the national synchrotron radiation research Center - From the 1.5 GeV TLS to the 3.3 GeV TPS
    National Synchrotron Radiation Research Center, Hsinchu Science Park, Hsinchu 30076, Taiwan
    不详
    Shinku, 2006, 12 (748-753)
  • [33] Development of a portable laser heating device for synchrotron radiation in situ experiment
    Li H.
    Zhao B.
    Zhang X.
    Guo Z.
    Wang Y.
    Zhu D.
    Yao Z.
    Cui C.
    Jiao Y.
    Liu H.
    Xu Z.
    Ma L.
    Tai R.
    He Jishu/Nuclear Techniques, 2023, 46 (02):
  • [34] Development of a long trace profiler at SPring-8 using a newly developed slope sensor
    Senba, Y.
    Kishimoto, H.
    Miura, T.
    Ohashi, H.
    ADVANCES IN METROLOGY FOR X-RAY AND EUV OPTICS VI, 2016, 9962
  • [35] Design and Commission of a Superconducting Wiggler X-ray Beamline for Advanced Materials Investigation at the National Synchrotron Radiation Research Center
    Chang, Shih-Hung
    Chang, Chien-Hung
    Juang, Jwei-Ming
    Huang, Liang-Jen
    Lin, Tsai-Fu
    Liu, Chin-Yen
    Chang, Chian-Feng
    Liu, Din-Goa
    Tsang, King-Long
    Pong, Wei-Fong
    Du, Chao-Hung
    Chang, Shih-Lin
    Soo, Yun-Liang
    Tang, Mau-Tsu
    CHINESE JOURNAL OF PHYSICS, 2012, 50 (02) : 220 - 228
  • [36] CONSTRUCTION AND COMMISSIONING OF THE SYNCHROTRON-RADIATION-RESEARCH-CENTER VACUUM-SYSTEM
    HSIUNG, GY
    HUANG, JR
    SHYY, JG
    WANG, DJ
    CHEN, JR
    LIU, YC
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1994, 12 (04): : 1639 - 1643
  • [37] Vacuum chamber for the wiggler of the Taiwan Light Source at the Synchrotron Radiation Research Center
    Wang, DJ
    Chen, JR
    Hsiung, GY
    Shyy, JG
    Huang, JR
    Hsu, SN
    Hsiao, KM
    Liu, YC
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1996, 14 (04): : 2624 - 2626
  • [38] In situ analysis of trace elements in metalloproteins of human liver by synchrotron radiation X-ray fluorescence
    Chunying Chen
    Peiqun Zhang
    Zhifang Chai
    Guangcheng Li
    Yuying Huang
    Science in China Series A: Mathematics, 2000, 43 : 88 - 92
  • [39] In situ analysis of trace elements in metalloproteins of human liver by synchrotron radiation X-ray fluorescence
    Chen, CY
    Zhang, PQ
    Chai, ZF
    Li, GC
    Huang, YY
    SCIENCE IN CHINA SERIES A-MATHEMATICS PHYSICS ASTRONOMY, 2000, 43 (01): : 88 - 92
  • [40] In situ analysis of trace elements in metalloproteins of human liver by synchrotron radiation X-ray fluorescence
    陈春英
    章佩群
    柴之芳
    李光城
    黄宇营
    Science China Mathematics, 2000, (01) : 88 - 92