Development of a long trace profiler in situ at National Synchrotron Radiation Research Center

被引:3
|
作者
Lin, Shang-Wei [1 ]
Wang, Duan-Jen [1 ]
Fu, Huang-Wen [1 ]
Tsai, Huang-Ming [1 ]
Hua, Chih-Yu [1 ]
Kuo, Chang-Yang [1 ,2 ]
Hsu, Ming-Ying [1 ]
Kao, Kai-Yang [1 ]
Yin, Gung-Chian [1 ]
Fung, Hok-Sum [1 ]
Perng, Shen-Yaw [1 ]
Chang, Chia-Feng [1 ]
机构
[1] Natl Synchrotron Radiat Res Ctr, Hsinchu 30076, Taiwan
[2] Max Planck Inst Chem Phys Solids, Nothnitzer Str 40, D-01187 Dresden, Germany
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2019年 / 90卷 / 02期
关键词
DISTORTION PROFILE; MIRROR;
D O I
10.1063/1.5055634
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
To achieve an ultrahigh resolution of a beamline for soft X-rays at the Taiwan Photon Source (TPS), the profile of a highly precise grating is required at various curvatures. The slope error could be decreased to 0.1 mu rad (rms) at a thermal load with a specially designed bender having 25 actuators. In the meantime, a long-trace profiler (LTP) was developed in situ to monitor the grating profile under a thermal load; it consists of a moving optical head, an air-bearing slide, an adjustable stand, and a glass viewport on the vacuum chamber. In the design of this system, a test chamber with an interior mirror was designed to simulate the chamber in the beamline. To prevent an error induced from a commercial viewport, a precision glass viewport (150CF, flatness 1/150 lambda rms at 632.8 nm) was designed. The error induced from the slope error of the glass surface and the vacuum deformation was also simulated. The performance of the optical head of the LTP in situ (ISLTP) has been tested in the metrology laboratory. The sources of error of this LTP including the linearity and the glass viewport were corrected after the measurement. For the beamline measurement, an optical head was mounted outside the vacuum chamber; the measuring beam passed through the glass viewport to measure the grating profile in vacuum. The measurement of the LTP after correction of the above errors yielded a precision about 0.2 mu rad (rms). In a preliminary test, an ISLTP was used to measure the grating profile at soft X-ray beamline TPS45A. The measured profile was for the bending mechanism to optimize the slope profile. From the measured energy spectrum, the slope error of the grating was estimated with software for optical simulation to be about 0.3 mu rad (rms), consistent with our estimate of the ISLTP. In the future, it will be used to monitor the thermal bump under a large thermal load. In addition, an ISLTP was used to monitor the properties of optical elements-the twist and radius in the beamline during the installation phase. Published under license by AIP Publishing.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Biopharmaceutical beamline TLS 15A1 for macromolecular crystallography at the National Synchrotron Radiation Research Center
    Chiang, Cheng-Hung
    Chou, Chung-Kuang
    Tseng, Chien-Chang
    Chen, Yi-Hui
    Liu, Yi-Chun
    Huang, Chen-Ying
    Chao, Chun-Hsiung
    Huang, Chun-Hsiang
    JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 2024, 71 (07) : 721 - 731
  • [22] Design of cryogenic system for several super-conducting modules at national synchrotron radiation research center
    King, TC
    Hsiao, FZ
    Chang, SH
    Li, HC
    Chang, JC
    Chen, JR
    PROCEEDINGS OF THE 2003 PARTICLE ACCELERATOR CONFERENCE, VOLS 1-5, 2003, : 2405 - 2407
  • [23] In-Situ Synchrotron and Neutron Radiation Advanced Welding Research
    Elmer, John W.
    Wu, Amanda S.
    WELDING JOURNAL, 2016, 95 (07) : 53 - 54
  • [24] Development of a high-precision long trace profiler utilizing shear measurements
    Cui, Xiaowen
    Wu, Ming
    Dong, Han
    Qin, Xiaobo
    Yang, Fugui
    Zhang, Xiaowei
    Li, Ming
    Sheng, Weifan
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2024, 95 (10):
  • [25] Erratum to: Synchrotron Radiation Based Research at the Novosibirsk Scientific Center
    K. V. Zolotarev
    A. I. Ancharov
    Z. S. Vinokurov
    B. G. Goldenberg
    F. A. Darin
    V. V. Kriventsov
    G. N. Kulipanov
    K. E. Cooper
    A. A. Legkodymov
    G. A. Lyubas
    A. D. Nikolenko
    K. A. Ten
    B. P. Tolochko
    M. R. Sharafutdinov
    A. N. Shmakov
    E. B. Levichev
    P. A. Piminov
    A. N. Zhuravlev
    Bulletin of the Russian Academy of Sciences: Physics, 2023, 87 (10) : 1557 - 1557
  • [26] Vacuum performance of the Synchrotron Radiation Research Center 1.3 GeV synchrotron light source
    Hsiung, GY
    Huang, JR
    Shyy, JG
    Wang, DJ
    Chen, JR
    Liu, YC
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A-VACUUM SURFACES AND FILMS, 1996, 14 (06): : 3275 - 3277
  • [27] The in-situ long trace profiler window glass thermal deformation effect of measurement analysis
    Hsu, Ming-Ying
    Yin, Gung-Chian
    Lee, Chien-yu
    Chen, Bo-Yi
    Fung, Hok-Sum
    Lin, Shang-Wei
    Wang, Duan-Jen
    Huang, Yu-Shan
    OPTICAL MODELING AND SYSTEM ALIGNMENT, 2019, 11103
  • [28] In Situ Testing Using Synchrotron Radiation Computed Tomography in Materials Research
    Xinchen Ni
    Nathan K. Fritz
    Brian L. Wardle
    MRS Advances, 2019, 4 : 2831 - 2841
  • [29] In Situ Testing Using Synchrotron Radiation Computed Tomography in Materials Research
    Ni, Xinchen
    Fritz, Nathan K.
    Wardle, Brian L.
    MRS ADVANCES, 2019, 4 (51-52) : 2831 - 2841
  • [30] Wiggler x-ray beamlines at synchrotron radiation research center
    Tsang, K.-L.
    Lee, C.-H.
    Jean, Y.-C.
    Dann, T.-E.
    Chen, J.-R.
    D'Amico, K.L.
    Oversluizen, T.
    Review of Scientific Instruments, 1995, 66 (2 pt 2):