Development of a long trace profiler in situ at National Synchrotron Radiation Research Center

被引:3
|
作者
Lin, Shang-Wei [1 ]
Wang, Duan-Jen [1 ]
Fu, Huang-Wen [1 ]
Tsai, Huang-Ming [1 ]
Hua, Chih-Yu [1 ]
Kuo, Chang-Yang [1 ,2 ]
Hsu, Ming-Ying [1 ]
Kao, Kai-Yang [1 ]
Yin, Gung-Chian [1 ]
Fung, Hok-Sum [1 ]
Perng, Shen-Yaw [1 ]
Chang, Chia-Feng [1 ]
机构
[1] Natl Synchrotron Radiat Res Ctr, Hsinchu 30076, Taiwan
[2] Max Planck Inst Chem Phys Solids, Nothnitzer Str 40, D-01187 Dresden, Germany
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2019年 / 90卷 / 02期
关键词
DISTORTION PROFILE; MIRROR;
D O I
10.1063/1.5055634
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
To achieve an ultrahigh resolution of a beamline for soft X-rays at the Taiwan Photon Source (TPS), the profile of a highly precise grating is required at various curvatures. The slope error could be decreased to 0.1 mu rad (rms) at a thermal load with a specially designed bender having 25 actuators. In the meantime, a long-trace profiler (LTP) was developed in situ to monitor the grating profile under a thermal load; it consists of a moving optical head, an air-bearing slide, an adjustable stand, and a glass viewport on the vacuum chamber. In the design of this system, a test chamber with an interior mirror was designed to simulate the chamber in the beamline. To prevent an error induced from a commercial viewport, a precision glass viewport (150CF, flatness 1/150 lambda rms at 632.8 nm) was designed. The error induced from the slope error of the glass surface and the vacuum deformation was also simulated. The performance of the optical head of the LTP in situ (ISLTP) has been tested in the metrology laboratory. The sources of error of this LTP including the linearity and the glass viewport were corrected after the measurement. For the beamline measurement, an optical head was mounted outside the vacuum chamber; the measuring beam passed through the glass viewport to measure the grating profile in vacuum. The measurement of the LTP after correction of the above errors yielded a precision about 0.2 mu rad (rms). In a preliminary test, an ISLTP was used to measure the grating profile at soft X-ray beamline TPS45A. The measured profile was for the bending mechanism to optimize the slope profile. From the measured energy spectrum, the slope error of the grating was estimated with software for optical simulation to be about 0.3 mu rad (rms), consistent with our estimate of the ISLTP. In the future, it will be used to monitor the thermal bump under a large thermal load. In addition, an ISLTP was used to monitor the properties of optical elements-the twist and radius in the beamline during the installation phase. Published under license by AIP Publishing.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] A novel long trace profiler for synchrotron radiation optics
    Li, Z
    Zhao, Y
    Li, DC
    Xiao, TQ
    Xia, SJ
    OPTICS AND LASER TECHNOLOGY, 2005, 37 (05): : 391 - 396
  • [2] Principle of π-phase plate long trace profiler for synchrotron radiation optics
    Zhao, Y
    Li, Z
    Li, DC
    Xiao, TQ
    Xia, SJ
    OPTICS COMMUNICATIONS, 2001, 200 (1-6) : 23 - 26
  • [3] STATUS OF THE RESEARCH-AND-DEVELOPMENT SYNCHROTRON RADIATION CENTER ZELENOGRAD
    KOVALCHUK, MV
    KULIPANOV, GN
    LITVINOV, YM
    MAZURENKO, SN
    SAMSONOV, S
    ZHILTSOV, VI
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 308 (1-2): : 1 - 5
  • [4] Precise measuring method for detecting the in situ distortion profile of a high-heat-load mirror for synchrotron radiation by use of a pentaprism long trace profiler
    Qian, SN
    Jark, W
    Sostero, G
    Gambitta, A
    Mazzolini, F
    Savoia, A
    APPLIED OPTICS, 1997, 36 (16): : 3769 - 3775
  • [5] Cryogenic oxygen deficiency hazard assessment at the National Synchrotron Radiation Research Center
    Kao, S. P.
    Wen, P. J.
    Lin, Y. C.
    IPAC23 PROCEEDINGS, 2024, 2687
  • [6] Development of a multi-beam long trace profiler
    Kilaru, Kiranmayee
    Merthe, Daniel J.
    Ali, Zulfiqar
    Gubarev, Mikhail V.
    Kester, Thomas
    Benson, Carl M.
    McKinney, Wayne R.
    Takacs, Peter Z.
    Yashchuk, Valeriy V.
    OPTICS FOR EUV, X-RAY, AND GAMMA-RAY ASTRONOMY V, 2011, 8147
  • [7] Improvement of synchrotron radiation mirrors below the 0.1 arcsec rms slope error limit with the help of a long trace profiler
    Lammert, H
    Senf, F
    Berger, M
    MATERIALS, MANUFACTURING, AND MEASUREMENT FOR SYNCHROTRON RADIATION MIRRORS, 1997, 3152 : 168 - 179
  • [8] In Situ Synchrotron Radiation Research in Materials Science
    Pedro A. Montano
    Hiroyuki Oyanagi
    MRS Bulletin, 1999, 24 : 13 - 20
  • [9] In situ synchrotron radiation research in materials science
    Montano, PA
    Oyanagi, H
    MRS BULLETIN, 1999, 24 (01) : 13 - 16
  • [10] Respiratory protective equipment fit tests for researchers at the national synchrotron radiation research center
    Wen, P. J.
    Kao, S. P.
    Lin, Y. C.
    Lin, S. Y.
    Chang, M. H.
    IPAC23 PROCEEDINGS, 2024, 2687