Prompt Gamma Imaging for In Vivo Range Verification of Pencil Beam Scanning Proton Therapy

被引:123
|
作者
Xie, Yunhe [1 ]
Bentefour, El Hassane [2 ]
Janssens, Guillaume [2 ]
Smeets, Julien [2 ]
Vander Stappen, Francois [2 ]
Hotoiu, Lucian [2 ]
Yin, Lingshu [1 ]
Dolney, Derek [1 ]
Avery, Stephen [1 ]
O'Grady, Fionnbarr [1 ]
Prieels, Damien [2 ]
McDonough, James [1 ]
Solberg, Timothy D. [3 ]
Lustig, Robert A. [1 ]
Lin, Alexander [1 ]
Teo, Boon-Keng K. [1 ]
机构
[1] Univ Penn, Dept Radiat Oncol, 3400 Civ Ctr Blvd,TRC 2 W, Philadelphia, PA 19104 USA
[2] Ion Beam Applicat SA, Adv Technol Grp, Louvain La Neuve, Belgium
[3] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA USA
关键词
CLINICAL-APPLICATION; SLIT CAMERA; EMISSION; SYSTEM; TOMOGRAPHY; DELIVERY; EDGE;
D O I
10.1016/j.ijrobp.2017.04.027
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To report the first clinical results and value assessment of prompt gamma imaging for in vivo proton range verification in pencil beam scanning mode. Methods and Materials: A stand-alone,trolley-mounted, prototype prompt gamma camera utilizing a knife-edge slit collimator design was used to record the prompt gamma signal emitted along the proton tracks during delivery of proton therapy for a brain cancer patient. The recorded prompt gamma depth detection profiles of individual pencil beam spots were compared with the expected profiles simulated from the treatment plan. Results: In 6 treatment fractions recorded over 3 weeks, the mean ( +/- standard deviation) range shifts aggregated over all spots in 9 energy layers were -0.8 +/- 1.3 mm for the lateral field, 1.7 +/- 0.7 mm for the right-superior-oblique field, and -0.4 +/- 0.9 mm for the vertex field. Conclusions: This study demonstrates the feasibility and illustrates the distinctive benefits of promptgamma imaging in pencil beam scanning treatmentmode. Accuracy in range verification was found in this first clinical case to be better than the range uncertainty margin applied in the treatment plan. These first results lay the foundation for additional work toward tighter integration of the system for in vivo proton range verification and quantification of range uncertainties. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:210 / 218
页数:9
相关论文
共 50 条
  • [41] Prompt gamma imaging of proton pencil beams at clinical dose rate
    Perali, I.
    Celani, A.
    Bombelli, L.
    Fiorini, C.
    Camera, F.
    Clementel, E.
    Henrotin, S.
    Janssens, G.
    Prieels, D.
    Roellinghoff, F.
    Smeets, J.
    Stichelbaut, F.
    Stappen, F. Vander
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (19): : 5849 - 5871
  • [42] Pencil Beam Scanning:: a dynamical approach to proton therapy
    Sepulchre, R
    Gérard, M
    Marchand, B
    Prieels, D
    Bauvir, B
    [J]. IEEE INTERNATIONAL SYMPOSIUM ON BIO-INFORMATICS AND BIOMEDICAL ENGINEERING, PROCEEDINGS, 2000, : 370 - 375
  • [43] Application of the HICAM Camera for Imaging of Prompt Gamma Rays in Measurements of Proton Beam Range
    Peloso, R.
    Busca, P.
    Fiorini, C.
    Basilavecchia, M.
    Frizzi, T.
    Smeets, J.
    Roellinghoff, F.
    Prieels, D.
    Stichelbaut, F.
    Benilov, A.
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2285 - 2289
  • [44] Effective Treatment Plan Integrity Verification for Passive Scattering and Pencil Beam Scanning Proton Therapy
    Zhang, Y.
    Ho, M.
    Liu, C.
    Liang, X.
    Johnson, P.
    Park, J.
    Artz, M.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [45] Hybrid in-beam PET- and Compton prompt-gamma imaging aimed at enhanced proton-range verification
    J. Balibrea-Correa
    J. Lerendegui-Marco
    I. Ladarescu
    C. Guerrero
    T. Rodríguez-González
    M. C. Jiménez-Ramos
    B. Fernández-Martínez
    C. Domingo-Pardo
    [J]. The European Physical Journal Plus, 137
  • [46] Hybrid in-beam PET- and Compton prompt-gamma imaging aimed at enhanced proton-range verification
    Balibrea-Correa, J.
    Lerendegui-Marco, J.
    Ladarescu, I
    Guerrero, C.
    Rodriguez-Gonzalez, T.
    Jimenez-Ramos, M. C.
    Fernandez-Martinez, B.
    Domingo-Pardo, C.
    [J]. EUROPEAN PHYSICAL JOURNAL PLUS, 2022, 137 (11):
  • [47] EXPERIMENTAL VERIFICATION OF KEY CROSS SECTIONS FOR PROMPT-GAMMA IMAGING IN PROTON THERAPY
    Wronska, A.
    Ahmed, A. Anees
    Andres, A.
    Bednarczyk, P.
    Besuglow, J.
    Gazdowicz, G.
    Herweg, K.
    Hetzel, R.
    Kasper, J.
    Kelleter, L.
    Konefal, A.
    Kulessa, P.
    Laihem, K.
    Leidner, J.
    Magiera, A.
    Rusiecka, K.
    Stachura, D.
    Stahl, A.
    Zieblinski, M.
    [J]. ACTA PHYSICA POLONICA B, 2017, 48 (10): : 1631 - 1637
  • [48] A Method to Improve the Spatial Resolution of Prompt Gamma Based Compton Imaging for Proton Range Verification
    Draeger, E.
    Mackin, D.
    Peterson, S.
    Chen, H.
    Beddar, S.
    Polf, J.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3451 - 3451
  • [49] Deep Learning-Based Prompt-Gamma Imaging for Range Verification in Proton Radiotherapy
    Jiang, Z.
    Polf, J.
    Ren, L.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E210 - E210
  • [50] Predictive Models for Cone Beam CT-Based Online Verification of Pencil Beam Scanning Proton Therapy
    Yin, L.
    Janssens, G.
    Lin, A.
    Ahn, P.
    Solberg, T.
    McDonough, J.
    Teo, B.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3454 - 3454