Biomechanical comparison between stainless steel, titanium and carbon-fiber reinforced polyetheretherketone volar locking plates for distal radius fractures

被引:46
|
作者
Mugnai, Raffaele [1 ]
Tarallo, Luigi [1 ]
Capra, Francesco
Catani, Fabio [1 ]
机构
[1] Modena Univ Hosp, Orthopaed & Traumatol Dept, Via Pietro Giardini 1355, I-41126 Modena, MO, Italy
关键词
Distal radius; Plate; Bending stiffness; Load to failure; Biomechanical; Material; ORTHOPEDIC-SURGERY; COMPOSITE-MATERIAL; INTERNAL-FIXATION; SHAFT FRACTURES; PLASTIC PLATES; SCREW DESIGN; BONE; IMPLANTS; COMPRESSION; STABILITY;
D O I
10.1016/j.otsr.2018.05.002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: As the popularity of volar locked plate fixation for distal radius fractures has increased, so have the number and variety of implants, including variations in plate design, the size and angle of the screws, the locking screw mechanism, and the material of the plates. Hypothesis: Carbon-fiber reinforced polyetheretherketone (CFR-PEEK) plate features similar biomechanical properties to metallic plates, representing, therefore, an optimal alternative for the treatment of distal radius fractures. Materials and Methods: Three different materials-composed plates were evaluated: stainless steel volar lateral column (Zimmer); titanium DVR (Hand Innovations); CFR-PEEK DiPHOS-RM (Lima Corporate). Six plates for each type were implanted in sawbones and an extra-articular rectangular osteotomy was created. Three plates for each material were tested for load to failure and bending stiffness in axial compression. Moreover, 3 constructs for each plate were evaluated after dynamically loading for 6000 cycles of fatigue. Results: The mean bending stiffness pre-fatigue was significantly higher for the stainless steel plate. The titanium plate yielded the higher load to failure both pre and post fatigue. After cyclic loading, the bending stiffness increased by a mean of 24% for the stainless steel plate; 33% for the titanium; and 17% for the CFR-PEEK plate. The mean load to failure post-fatigue increased by a mean of 10% for the stainless steel and 14% for CFR-PEEK plates, whereas it decreased ( -16%) for the titanium plate. Statistical analysis between groups reported significant values (p < 001) for all comparisons except for Hand Innovations vs. Zimmer bending stiffness post fatigue (p = .197). Discussion: The significant higher load to failure of the titanium plate, makes it indicated for patients with higher functional requirements or at higher risk of trauma in the post-operative period. The CFR-PEEK plate showed material-specific disadvantages, represented by little tolerance to plastic deformation, and lower load to failure. (C) 2018 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:877 / 882
页数:6
相关论文
共 50 条
  • [41] Can the Use of Variable-Angle Volar Locking Plates Compensate for Suboptimal Plate Positioning in Unstable Distal Radius Fractures? A Biomechanical Study
    Hart, Adam
    Collins, Melissa
    Chhatwal, Dane
    Steffen, Thomas
    Harvey, Edward J.
    Martineau, Paul A.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2015, 29 (01) : E1 - E6
  • [42] A biomechanical comparison of two volar locked plates in a dorsally unstable distal radius fracture model
    Liporace, Frank Anthony
    Kubiak, Erik N.
    Jeong, Gerard K.
    Iesaka, Kazuho
    Egol, Kenneth A.
    Koval, Kenneth J.
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2006, 61 (03): : 668 - 672
  • [43] Volar locking plate fixation for distal radius fractures: did variable-angle plates make difference?
    Mohamed Abdel-Wahed
    Ahmed Abdel-Zaher Khater
    Mahmoud Ahmed El-Desouky
    International Orthopaedics, 2022, 46 : 2165 - 2176
  • [44] Volar fixation of dorsally displaced distal radius fractures using the 2.4-mm locking compression plates
    Musgrave, DS
    Idler, RS
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2005, 30A (04): : 743 - 749
  • [45] Functional and Radiological Outcomes in Intra-Articular Fractures of Distal Radius with Volar Variable Angle Locking Plates
    Arora, Sahil
    Govil, Vasudha
    Neogi, Avik Kr
    Bishnoi, Sanju
    Paul, Shagnik
    Gupta, R. K.
    TRAUMA MONTHLY, 2022, 27 (01) : 380 - 385
  • [46] Radiographic Alignment of Unstable Distal Radius Fractures Fixed With 1 or 2 Rows of Screws in Volar Locking Plates
    Neuhaus, Valentin
    Badri, Omar
    Ferree, Steven
    Bot, Arjan G. J.
    Ring, David C.
    Mudgal, Chaitanya S.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2013, 38A (02): : 297 - 301
  • [47] Volar locking plate fixation for distal radius fractures: did variable-angle plates make difference?
    Abdel-Wahed, Mohamed
    Khater, Ahmed Abdel-Zaher
    El-Desouky, Mahmoud Ahmed
    INTERNATIONAL ORTHOPAEDICS, 2022, 46 (09) : 2165 - 2176
  • [48] What happens to the posterior comminution in extra-articular fractures of the distal radius treated with volar locking plates?
    Guillou, J.
    Darees, M.
    Pouges, C.
    Christiaens, N.
    Guerre, E.
    Chantelot, C.
    HAND SURGERY & REHABILITATION, 2019, 38 (02): : 91 - 96
  • [49] Are Volar Locking Plates Superior to Percutaneous K-wires for Distal Radius Fractures? A Meta-analysis
    Chaudhry, Harman
    Kleinlugtenbelt, Ydo V.
    Mundi, Raman
    Ristevski, Bill
    Goslings, J. C.
    Bhandari, Mohit
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2015, 473 (09) : 3017 - 3027
  • [50] REOPERATION FOR METALWORK COMPLICATIONS FOLLOWING THE USE OF VOLAR LOCKING PLATES FOR DISTAL RADIUS FRACTURES: A UNITED KINGDOM EXPERIENCE
    Sahu, A.
    Charalambous, C. P.
    Mills, S. P.
    Batra, S.
    Ravenscroft, M. J.
    JOURNAL OF HAND SURGERY-ASIAN-PACIFIC VOLUME, 2011, 16 (02): : 113 - 118