Biomechanical comparison between stainless steel, titanium and carbon-fiber reinforced polyetheretherketone volar locking plates for distal radius fractures

被引:46
|
作者
Mugnai, Raffaele [1 ]
Tarallo, Luigi [1 ]
Capra, Francesco
Catani, Fabio [1 ]
机构
[1] Modena Univ Hosp, Orthopaed & Traumatol Dept, Via Pietro Giardini 1355, I-41126 Modena, MO, Italy
关键词
Distal radius; Plate; Bending stiffness; Load to failure; Biomechanical; Material; ORTHOPEDIC-SURGERY; COMPOSITE-MATERIAL; INTERNAL-FIXATION; SHAFT FRACTURES; PLASTIC PLATES; SCREW DESIGN; BONE; IMPLANTS; COMPRESSION; STABILITY;
D O I
10.1016/j.otsr.2018.05.002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: As the popularity of volar locked plate fixation for distal radius fractures has increased, so have the number and variety of implants, including variations in plate design, the size and angle of the screws, the locking screw mechanism, and the material of the plates. Hypothesis: Carbon-fiber reinforced polyetheretherketone (CFR-PEEK) plate features similar biomechanical properties to metallic plates, representing, therefore, an optimal alternative for the treatment of distal radius fractures. Materials and Methods: Three different materials-composed plates were evaluated: stainless steel volar lateral column (Zimmer); titanium DVR (Hand Innovations); CFR-PEEK DiPHOS-RM (Lima Corporate). Six plates for each type were implanted in sawbones and an extra-articular rectangular osteotomy was created. Three plates for each material were tested for load to failure and bending stiffness in axial compression. Moreover, 3 constructs for each plate were evaluated after dynamically loading for 6000 cycles of fatigue. Results: The mean bending stiffness pre-fatigue was significantly higher for the stainless steel plate. The titanium plate yielded the higher load to failure both pre and post fatigue. After cyclic loading, the bending stiffness increased by a mean of 24% for the stainless steel plate; 33% for the titanium; and 17% for the CFR-PEEK plate. The mean load to failure post-fatigue increased by a mean of 10% for the stainless steel and 14% for CFR-PEEK plates, whereas it decreased ( -16%) for the titanium plate. Statistical analysis between groups reported significant values (p < 001) for all comparisons except for Hand Innovations vs. Zimmer bending stiffness post fatigue (p = .197). Discussion: The significant higher load to failure of the titanium plate, makes it indicated for patients with higher functional requirements or at higher risk of trauma in the post-operative period. The CFR-PEEK plate showed material-specific disadvantages, represented by little tolerance to plastic deformation, and lower load to failure. (C) 2018 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:877 / 882
页数:6
相关论文
共 50 条
  • [21] CARBON-FIBER REINFORCED PLATES FOR PROBLEM FRACTURES
    PEMBERTON, DJ
    MCKIBBIN, B
    SAVAGE, R
    TAYTON, K
    STUART, D
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01): : 88 - 92
  • [22] Comparison of the Orientation Angles of Volar Locking Plate Distal Ulnar Locking Screw for Distal Radius Fractures
    Abe, Shingo
    Shimada, Toshiki
    Murase, Tsuyoshi
    Kuriyama, Kohji
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2022, 47 (04): : 389.e1 - 389.e13
  • [23] Osteosynthesis with long volar locking plates for metaphyseal-diaphyseal fractures of the distal radius
    Paritosh Gogna
    Harpal Singh Selhi
    Rohit Singla
    Mukul Mohindra
    Amit Batra
    Reetadyuti Mukhopadhyay
    Rajesh Rohilla
    Umesh Yadav
    中华创伤杂志(英文版), 2013, 16 (06) : 339 - 343
  • [24] Osteosynthesis with long volar locking plates for metaphyseal-diaphyseal fractures of the distal radius
    Gogna, Paritosh
    Selhi, Harpal Singh
    Singla, Rohit
    Mohindra, Mukul
    Batra, Amit
    Mukhopadhyay, Reetadyuti
    Rohilla, Rajesh
    Yadav, Umesh
    CHINESE JOURNAL OF TRAUMATOLOGY, 2013, 16 (06) : 339 - 343
  • [25] Ulnar Styloid Fracture in Distal Radius Fractures Managed with Volar Locking Plates: To Fix or Not?
    Gogna, Paritosh
    Selhi, Harpal Singh
    Mohindra, Mukul
    Thora, Ankit
    Yamin, Mohammad
    JOURNAL OF HAND AND MICROSURGERY, 2014, 6 (02) : 53 - 58
  • [26] Flexor tendon complications in distal radius fractures treated with volar rim locking plates
    Fardellas, A.
    Vernet, P.
    Facca, S.
    Liverneaux, P.
    HAND SURGERY & REHABILITATION, 2020, 39 (06): : 511 - 515
  • [27] Locking versus nonlocking T-plates for dorsal and volar fixation of dorsally comminuted distal radius fractures: A biomechanical study
    Trease, C
    McIff, T
    Toby, EB
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2005, 30A (04): : 756 - 763
  • [28] More is not necessarily better. A biomechanical study on distal screw numbers in volar locking distal radius plates
    Drobetz, Herwig
    Weninger, Patrick
    Grant, Caroline
    Heal, Clare
    Muller, Reinhold
    Schuetz, Michael
    Pham, Minh
    Steck, Roland
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2013, 44 (04): : 535 - 539
  • [29] Biomechanical Properties of 3-Dimensional Printed Volar Locking Distal Radius Plate: Comparison With Conventional Volar Locking Plate
    Kim, Sung-Jae
    Jo, Young-Hoon
    Choi, Wan-Sun
    Lee, Chang-Hun
    Lee, Bong-Gun
    Kim, Joo-Hak
    Lee, Kwang-Hyun
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2017, 42 (09): : 747.e1 - 747.e6
  • [30] Prospective Evaluation of Distal Radius Fractures Treated With Variable-Angle Volar Locking Plates
    Fowler, John R.
    Ilyas, Asif M.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2013, 38A (11): : 2198 - 2203