Biomechanical comparison between stainless steel, titanium and carbon-fiber reinforced polyetheretherketone volar locking plates for distal radius fractures

被引:46
|
作者
Mugnai, Raffaele [1 ]
Tarallo, Luigi [1 ]
Capra, Francesco
Catani, Fabio [1 ]
机构
[1] Modena Univ Hosp, Orthopaed & Traumatol Dept, Via Pietro Giardini 1355, I-41126 Modena, MO, Italy
关键词
Distal radius; Plate; Bending stiffness; Load to failure; Biomechanical; Material; ORTHOPEDIC-SURGERY; COMPOSITE-MATERIAL; INTERNAL-FIXATION; SHAFT FRACTURES; PLASTIC PLATES; SCREW DESIGN; BONE; IMPLANTS; COMPRESSION; STABILITY;
D O I
10.1016/j.otsr.2018.05.002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: As the popularity of volar locked plate fixation for distal radius fractures has increased, so have the number and variety of implants, including variations in plate design, the size and angle of the screws, the locking screw mechanism, and the material of the plates. Hypothesis: Carbon-fiber reinforced polyetheretherketone (CFR-PEEK) plate features similar biomechanical properties to metallic plates, representing, therefore, an optimal alternative for the treatment of distal radius fractures. Materials and Methods: Three different materials-composed plates were evaluated: stainless steel volar lateral column (Zimmer); titanium DVR (Hand Innovations); CFR-PEEK DiPHOS-RM (Lima Corporate). Six plates for each type were implanted in sawbones and an extra-articular rectangular osteotomy was created. Three plates for each material were tested for load to failure and bending stiffness in axial compression. Moreover, 3 constructs for each plate were evaluated after dynamically loading for 6000 cycles of fatigue. Results: The mean bending stiffness pre-fatigue was significantly higher for the stainless steel plate. The titanium plate yielded the higher load to failure both pre and post fatigue. After cyclic loading, the bending stiffness increased by a mean of 24% for the stainless steel plate; 33% for the titanium; and 17% for the CFR-PEEK plate. The mean load to failure post-fatigue increased by a mean of 10% for the stainless steel and 14% for CFR-PEEK plates, whereas it decreased ( -16%) for the titanium plate. Statistical analysis between groups reported significant values (p < 001) for all comparisons except for Hand Innovations vs. Zimmer bending stiffness post fatigue (p = .197). Discussion: The significant higher load to failure of the titanium plate, makes it indicated for patients with higher functional requirements or at higher risk of trauma in the post-operative period. The CFR-PEEK plate showed material-specific disadvantages, represented by little tolerance to plastic deformation, and lower load to failure. (C) 2018 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:877 / 882
页数:6
相关论文
共 50 条
  • [1] Comparison between Carbon-Peek volar locking plates and titanium volar locking plates in the treatment of distal radius fractures
    Perugia, Dario
    Guzzini, Matteo
    Mazza, Daniele
    Iorio, Carlo
    Civitenga, Carolina
    Ferretti, Andrea
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 : S24 - S29
  • [2] Biomechanical Comparison of Volar Fixed-Angle Locking Plates for AO C3 Distal Radius Fractures: Titanium Versus Stainless Steel With Compression
    Marshall, Tyler
    Momaya, Amit
    Eberhardt, Alan
    Chaudhari, Nilesh
    Hunt, Thomas R., III
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2015, 40 (10): : 2032 - 2038
  • [3] A biomechanical comparison between two volar locking plate systems for distal radius fractures
    Chen, Li
    Dai, Qiang
    Wongworawat, Montri D.
    ORTHOPEDICS, 2006, 29 (10) : 927 - 929
  • [4] Letter Regarding "Biomechanical Comparison of Volar Fixed-Angle Locking Plates for AO C3 Distal Radius Fractures: Titanium Versus Stainless Steel With Compression"
    Spiegel, Adrian
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2016, 41 (07): : E225 - E225
  • [5] Stainless steel versus titanium volar multi-axial locking plates for fixation of distal radius fractures: a randomised clinical trial
    Gregory B Couzens
    Susan E Peters
    Kenneth Cutbush
    Benjamin Hope
    Fraser Taylor
    Christopher D James
    Carly R Rankin
    Mark Ross
    BMC Musculoskeletal Disorders, 15
  • [6] Stainless steel versus titanium volar multi-axial locking plates for fixation of distal radius fractures: a randomised clinical trial
    Couzens, Gregory B.
    Peters, Susan E.
    Cutbush, Kenneth
    Hope, Benjamin
    Taylor, Fraser
    James, Christopher D.
    Rankin, Carly R.
    Ross, Mark
    BMC MUSCULOSKELETAL DISORDERS, 2014, 15
  • [7] Biomechanical Comparison of Titanium Locking Fragment-Specific and Volar Locking Plates for AO B1 and B2 Fractures of the Distal Radius
    Martin, Dennis P.
    Park, Andrew G.
    Jamison, David
    Jones, Christopher M.
    Ilyas, Asif M.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2019, 44 (12): : 1093.e1 - 1093.e8
  • [8] A Comparison Between Volar Locking Plates and Percutaneous Pinning in the Treatment of Distal Radius Fractures: A Systematic Review
    Gopal, Nikhil
    Anil, Abhijith
    Gopal, Meera
    Bhat, Anil K.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2025, 50 (03): : 265 - 273
  • [9] Randomized Comparison of Volar Locking Plates and Intramedullary Nails for Unstable Distal Radius Fractures
    Plate, Johannes F.
    Gaffney, Daniel L.
    Emory, Cynthia L.
    Mannava, Sandeep
    Smith, Beth P.
    Koman, L. Andrew
    Wiesler, Ethan R.
    Li, Zhongyu
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2015, 40 (06): : 1095 - 1101
  • [10] Surgical treatment of unstable distal radius fractures with volar locking plates
    Murakami, Kenichi
    Abe, Yoshihiro
    Takahashi, Kazuhisa
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2007, 12 (02) : 134 - 140