Continuous glucose monitoring using a microneedle array sensor coupled with a wireless signal transmitter

被引:77
|
作者
Kim, Kwang Bok [1 ]
Lee, Won-Chul [2 ]
Cho, Chul-Ho [1 ]
Park, Deog-Su [3 ]
Cho, Seong Je [1 ]
Shim, Yoon-Bo [2 ]
机构
[1] Samsung Elect, Samsung Res, Seoul R&D Campus,34,Seongchon Gil, Seoul 06765, South Korea
[2] Pusan Natl Univ, Dept Chem, Busan 46241, South Korea
[3] Pusan Natl Univ, IBST, Busan 46241, South Korea
关键词
Glucose sensor; Wireless biosensor; Microneedle array; Carboxylated-poly(terthiophene); Point-of care; ELECTRON-TRANSFER; SYSTEMS;
D O I
10.1016/j.snb.2018.10.081
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A minimally invasive wireless biosensor was developed for glucose monitoring using human blood. The sensor probe was shaped in a microneedle array with a cyclic olefin copolymer. FAD-glucose oxidase (GOx) was covalently bound to a terthiophene carboxylic acid (TCA) monomer followed by electropolymerization on a gold-coated microneedle (AuMN) array using the potential cycling method (AuMN/pTCA-GOx). The formation of the sensing probe layer on the needle was confirmed by employing electrochemical and surface characterization experiments. To prevent the interfering species and the detaching of the probe material, the AuMN/pTCA-GOx surface was protected with Nafion. To enhance the repeatability and stability of the sensor, it was pretreated by applying the sequential negative and positive pulses immediately prior to each measurement. The sequential pulse allowed the sensor to have a high repeatability (Coefficient of variation = 1.1%, n = 10), and continuously long-term stability (Coefficient of variation = 5.56% for 72 h). The sensor has an excellent linear response at glucose concentrations between 0.05 and 20.0 mM (R-2 = 0.99) with a sensitivity of 0.22 mu A/mM(-1) cm(-2) (Coefficient of variation = 1.7%) and the detection limit of 19.4 ( +/- 0.62) mu A. The reliability of the proposed method was performed on finger through coupling of the sensor part to a reusable wireless transmitter that sends the measured values to a mobile phone via Bluetooth.
引用
收藏
页码:14 / 21
页数:8
相关论文
共 50 条
  • [21] Microelectronic sensor for continuous glucose monitoring
    Sanches Nascimento, Raphael Aparecido
    Mulato, Marcelo
    APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 2019, 125 (03):
  • [22] Impaction of a continuous glucose monitoring sensor
    Park, Kyong Chan
    Choi, Hwan Jun
    ARCHIVES OF PLASTIC SURGERY-APS, 2021, 48 (04): : 392 - 394
  • [23] Microelectronic sensor for continuous glucose monitoring
    Raphael Aparecido Sanches Nascimento
    Marcelo Mulato
    Applied Physics A, 2019, 125
  • [24] Minimally invasive microneedle array electrodes employing direct electron transfer type glucose dehydrogenase for the development of continuous glucose monitoring sensors
    Sharma, Sanjiv
    Takagi, Eri
    Cass, Tony
    Tsugawa, Wakako
    Sode, Koji
    BIOSENSORS 2016, 2017, 27 : 208 - 209
  • [25] An Electrochemical Microneedle Biosensor with Wide Linear Range for Continuous Glucose Monitoring
    Reza, Md Selim
    Song, HyeSu
    Lee, YeYoung
    Asaduzzaman, Md
    Cha, Cheolung
    Park, Jae Yeong
    2023 IEEE SENSORS, 2023,
  • [27] A Minimally Invasive Implantable Sensor for Continuous Wireless Glucose Monitoring Based on a Passive Resonator
    Hassan, Rehab S.
    Lee, Jongheon
    Kim, Sanghoek
    IEEE ANTENNAS AND WIRELESS PROPAGATION LETTERS, 2020, 19 (01): : 124 - 128
  • [28] Fully Passive Sensor Coated With Phenylboronic Acid Hydrogel for Continuous Wireless Glucose Monitoring
    Hassan, Rehab S.
    Lee, Yongshik
    IEEE SENSORS JOURNAL, 2024, 24 (08) : 12025 - 12033
  • [29] A wireless, fully implantable continuous glucose sensor
    DeHennis, A. D.
    Modzelewska, A.
    Pearsall, B.
    Lorenz, C.
    Mdingi, C.
    Jiang, H.
    Lawrence, J.
    Mortellaro, M.
    Tymchyshyn, O.
    Huffstetler, P.
    Tankiewicz, S.
    Maiorana, T.
    Wang, X.
    Walters, S.
    Colvin, A.
    DIABETOLOGIA, 2011, 54 : S117 - S117
  • [30] A Novel Wireless Continuous Glucose Monitoring System Using Semiconductor Integration
    Mujeeb-U-Rahman, Muhammad
    Nazari, Meisam Honarvar
    Sencan, Mehmet
    DIABETES, 2019, 68