Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group

被引:53
|
作者
Chen, Min [1 ,2 ]
Zhang, Lina [1 ,3 ]
Xu, Jia [1 ,2 ]
Chen, Xinyi [1 ,2 ]
Gu, Yuxiang [1 ,2 ]
Ren, Yuping [4 ]
Wang, Kaijun [1 ,2 ]
机构
[1] Zhejiang Univ, Coll Med, Affiliated Hosp 2, Eye Ctr, 88 Jiefang Rd, Hangzhou 310009, Zhejiang, Peoples R China
[2] Zhejiang Prov Key Lab Ophthalmol, Hangzhou, Zhejiang, Peoples R China
[3] Lishui Peoples Hosp, Lishui, Zhejiang, Peoples R China
[4] Shaoxing Tradit Chinese Med Hosp, Shaoxing, Zhejiang, Peoples R China
基金
中国国家自然科学基金;
关键词
Intraocular pressure; iCare rebound tonometer; Non-contact tonometer; Goldmann applanation tonometer; Central corneal thickness; OCULAR RESPONSE ANALYZER; DYNAMIC CONTOUR; HEALTHY-SUBJECTS;
D O I
10.1186/s12886-019-1236-5
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background Measurement of intraocular pressure (IOP) is essential for glaucoma patients. Many factors such as central corneal thickness (CCT) can affect the accuracy of IOP measurement. The purpose of this study was to evaluate the agreement of IOP measured by non-contact tonometer (NCT), iCare pro rebound tonometer (iCare), and Goldmann applanation tonometer (GAT) in different IOP group. Methods This was a Hospital-based cross-sectional study. Two hundred subjects were enrolled in this study. All subjects underwent IOP measurement using an NCT-iCare-GAT sequence. Bland-Altman, Pearson correlation and intraclass correlation analysis were performed using SPSS 17.0 software. The influence of CCT on each IOP measurement methods was evaluated by linear regression analysis. Results The mean difference (Delta) of NCT-GAT did not differ from (Delta) iCare-GAT in IOP < 10 and 10-21 mmHg group. However, (Delta) NCT-GAT was significantly higher than (Delta) iCare-GAT in IOP 22-30 and > 30 mmHg group (P < 0.05). Bland-Altman analysis showed significant agreement between the three devices (P < 0.01). IOP measurements of the three methods were significantly correlated with CCT (P < 0.01). Conclusions ICare pro shows a higher agreement with GAT over a wide range of IOP compared with NCT. The consistency between the three tonometers was similar in a low and normal IOP range. However, NCT shows a greater overestimate of IOP in moderate and higher IOP group. The variability of IOP measurement affected by CCT is NCT > iCare pro > GAT.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] Inter-optometrist and intra-optometrist agreement for Intraocular Pressure (IOP) measurement by Goldmann Applanation Tonometry (GAT)
    Shah, Rakhee
    Edgar, David
    Campbell, Peter
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [42] COMPARATIVE MEASUREMENTS OF INTRAOCULAR-PRESSURE WITH THE NON-CONTACT TONOMETER AND THE GOLDMANN APPLANATION TONOMETER
    LANGMANN, G
    SCHUHMANN, G
    SCHWAIGER, W
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1985, 187 (05) : 345 - 347
  • [43] Intraocular pressure measurement in mice: a comparison between Goldmann and rebound tonometry
    C Y Kim
    M H Kuehn
    M G Anderson
    Y H Kwon
    Eye, 2007, 21 : 1202 - 1209
  • [44] Intraocular pressure measurement in mice: a comparison between Goldmann and rebound tonometry
    Kim, C. Y.
    Kuehn, M. H.
    Anderson, M. G.
    Kwon, Y. H.
    EYE, 2007, 21 (09) : 1202 - 1209
  • [45] Comparison of Non-contact Tonometry and Goldmann Applanation Tonometry Measurements in Non-pathologic High Myopia
    Wang, Peiyuan
    Song, Yunhe
    Lin, Fengbin
    Wang, Zhenyu
    Gao, Xinbo
    Cheng, Weijing
    Chen, Meiling
    Peng, Yuying
    Liu, Yuhong
    Zhang, Xiulan
    Chen, Shida
    FRONTIERS IN MEDICINE, 2022, 9
  • [46] Measuring Intraocular Pressure After Intrastromal Corneal Ring Segment Implantation With Rebound Tonometry and Goldmann Applanation Tonometry
    Arribas-Pardo, Paula
    Mendez-Hernandez, Carmen
    Cuina-Sardina, Ricardo
    Fernandez-Perez, Cristina
    Diaz-Valle, David
    Garcia-Feijoo, Julian
    CORNEA, 2015, 34 (05) : 516 - 520
  • [47] Comparison of Intraocular Pressure Measurements Obtained by Rebound, Noncontact, and Goldmann Applanation Tonometry in Children REPLY
    Feng, Chi Shian
    Jin, Ki Won
    Yi, Kayoung
    Choi, Dong Gyu
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2016, 163 : 192 - 193
  • [48] Comparison of Three Methods to Measure Intraocular Pressure: ic100 Icare, TAO1i Icare and Goldmann Applanation Tonometry.
    Sara Aurora, Garcia Y. Otero Sanchez
    Paczka, Jose A.
    Romo Sainz, Montserrat
    Dorantes Diez, Yesenia Yolanda
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [49] Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients
    Tan, Shaoying
    Yu, Marco
    Baig, Nafees
    Hansapinyo, Linda
    Tham, Clement C.
    SCIENTIFIC REPORTS, 2017, 7
  • [50] Agreement of patient-measured intraocular pressure using rebound tonometry with Goldmann applanation tonometry (GAT) in glaucoma patients
    Shaoying Tan
    Marco Yu
    Nafees Baig
    Linda Hansapinyo
    Clement C. Tham
    Scientific Reports, 7