Cyclic Fatigue Resistance of Nickel-titanium Rotary Instruments according to the Angle of File Access and Radius of Root Canal

被引:39
|
作者
Pedulla, Eugenio [1 ]
La Rosa, Giusy Rita Maria [1 ]
Virgillito, Chiara [1 ]
Rapisarda, Ernesto [1 ]
Kim, Hyeon-Cheol [2 ]
Generali, Luigi [3 ]
机构
[1] Univ Catania, Dept Gen Surg & Surg Med Specialties, Catania, Italy
[2] Pusan Natl Univ, Sch Dent, Dent & Life Sci Inst, Dent Res Inst,Dept Conservat Dent, Yangsan, South Korea
[3] Univ Modena & Reggio Emilia, Dept Surg Med Dent & Morphol Sci Transplant Surg, Modena, Italy
关键词
2Shape; angle of file access; canal curvature; canal radius; cyclic fatigue resistance; HyFlex CM; ENDODONTIC INSTRUMENTS; PROFILE VORTEX; RECIPROC BLUE; FRACTURE; DEFECTS; 2SHAPE; SINGLE;
D O I
10.1016/j.joen.2019.11.015
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: The aim of this study was to compare the influences from different access angles and curvature radii on cyclic fatigue resistance of nickel-titanium rotary files. Methods: Two file systems (2Shape [TS; MicroMega, Besancon, France] and HyFlex CM [HCM; Coltene/Whaledent, Allstatten, Switzerland]) were used. A total of 192 instruments of TS #25/.04 (TS1), TS #25/.06 (TS2), HCM #25/.04, and HCM #25/.06 were evaluated at 3 insertion angles (0 degrees, 10 degrees, and 20 degrees) and 2 radii (5 mm and 3 mm) in 16-mm stainless steel artificial canals with a 60 curvature. Cyclic fatigue resistance was determined by the number of cycles to failure (NCF) using a customized testing device. Data were analyzed statistically with the significance level established at 95%. Results: In the 3-mm radius canal, the instruments showed lower cyclic fatigue resistance than in the 5-mm radius canal (P < .05). HCM #25/.06 and all .04 taper instruments had a significant NCF reduction at 20 and 10 in the 3-mm radius canal (P < .05), whereas TS2 showed no significant differences. In the 5-mm radius of curvature, although .06 taper instruments had no significant NCF reduction for each angle tested, .04 taper files exhibited significant NCF reduction when tested at 20 (P < .05). Comparing the same size instruments, HCM had higher NCF than TS (P < .05). Instruments with a .04 taper exhibited higher NCF than the .06 ones with the same heat treatment (P < .05). Conclusions: An inclined insertion into the canals decreased cyclic fatigue resistance of thermal-treated instruments with a .04 taper at all radii of curvature tested. The synergistic effect of a small radius of curvature and access angulation of heat-treated instruments decreases their fatigue resistance.
引用
收藏
页码:431 / 436
页数:6
相关论文
共 50 条
  • [11] Angle of Insertion and Torsional Resistance of Nickel-Titanium Rotary Instruments
    Di Nardo, Dario
    Zanza, Alessio
    Seracchiani, Marco
    Donfrancesco, Orlando
    Gambarini, Gianluca
    Testarelli, Luca
    MATERIALS, 2021, 14 (13)
  • [12] Cyclic Fatigue Resistance of Nickel-Titanium Rotary Instruments after Simulated Clinical Use
    Ubaed, Hana R.
    Bakr, Diyar Kh.
    APPLIED BIONICS AND BIOMECHANICS, 2022, 2022
  • [13] Effect of clinical use on the cyclic fatigue resistance of ProTaper nickel-titanium rotary instruments
    Ounsi, Hani F.
    Salameh, Ziad
    Al-Shalan, Thakib
    Ferrari, Marco
    Grandini, Simone
    Pashley, David H.
    Tay, Franklin R.
    JOURNAL OF ENDODONTICS, 2007, 33 (06) : 737 - 741
  • [14] Buckling resistance of pathfinding nickel-titanium rotary instruments in narrow root canal: an in vitro study
    Kill, Kleber Borgo
    Lopes, Helio Pereira
    Schultz Neves, Monica Aparecida
    Leal Vieira, Victor Talarico
    Elias, Carlos Nelson
    de Souza, Leticia Chaves
    ENDO-ENDODONTIC PRACTICE TODAY, 2018, 12 (02): : 133 - 138
  • [15] Cyclic Fatigue of Different Nickel-Titanium Rotary Instruments: A Comparative Study
    Testarelli, L.
    Grande, N. M.
    Plotino, G.
    Lendini, M.
    Pongione, G.
    De Paolis, G.
    Rizzo, F.
    Milana, V.
    Gambarini, G.
    OPEN DENTISTRY JOURNAL, 2009, 3 : 55 - 58
  • [16] The Effect of Temperature on Cyclic Fatigue of Nickel-titanium Rotary Endodontic Instruments
    Dosanjh, Amandeep
    Paurazas, Susan
    Askar, Mazin
    JOURNAL OF ENDODONTICS, 2017, 43 (05) : 823 - 826
  • [17] Effect of sterilization on cyclic fatigue of rotary nickel-titanium endodontic instruments
    Mize, SB
    Clement, DJ
    Pruett, JP
    Carnes, DL
    JOURNAL OF ENDODONTICS, 1998, 24 (12) : 843 - 847
  • [18] Cyclic Fatigue Testing and Metallographic Analysis of Nickel-Titanium Rotary Instruments
    Pirani, Chiara
    Cirulli, Pier Paolo
    Chersoni, Stefano
    Micele, Lorenzo
    Ruggeri, Oddone
    Prati, Carlo
    JOURNAL OF ENDODONTICS, 2011, 37 (07) : 1013 - 1016
  • [19] Evaluation of cyclic flexural fatigue resistance of 25/0.04 and 25/0.06 twisted file rotary nickel-titanium endodontic instruments
    Al-Hadlaq, Solaiman M.
    AUSTRALIAN ENDODONTIC JOURNAL, 2013, 39 (02) : 62 - 65
  • [20] Conditioning of root canal anatomy on static and dynamics of nickel-titanium rotary instruments
    Di Giuseppe, Italo
    Di Giuseppe, Davide
    Malagnino, Vito Antonio
    Silla, Enrico Paolo
    Somma, Francesco
    GIORNALE ITALIANO DI ENDODONZIA, 2015, 29 (02): : 58 - 64