The effect of disinfecting procedure on the glucose concentration using a personal glucose meter

被引:3
|
作者
Jonca, Mateusz [1 ]
Krotki, Filipina [1 ]
Tomasik, Przemyslaw [2 ]
机构
[1] Jagiellonian Univ Med Coll, Student Res Grp Clin Biochem, Krakow, Poland
[2] Jagiellonian Univ Med Coll, Pediat Inst, Dept Clin Biochem, Ul Wielicka 265, PL-30663 Krakow, Poland
关键词
Personal glucose meters; Disinfectant; Glucose concentration; POCT; VOLUME;
D O I
10.1016/j.pcd.2021.05.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Personal glucose meters, primarily designed for self-control of glucose concentration in patients with diabetes, are frequently used in clinical practice as point-of-care equipment. The procedure of washing hands with water and soap before testing, as recommended by the manufacturer of personal glucose meters, in hospital wards is often difficult to fulfil and is replaced by disinfecting the place of blood sampling with isopropyl alcohol. Aim: The purpose of the research was to evaluate the effect of different disinfecting procedures on glycemia measurement in capillary blood using personal glucose meters. Materials and methods: Four measurements of glycemia were taken in each of 50-volunteer group using Accu-Check Performa (Roche, Swiss) glucose meter using different procedures: washing hands with soap and drying them (1), disinfection with isopropyl alcohol and waiting for the disinfectant to evaporate (2) and the immediate puncture after disinfection. In the latter case two kind of single-use lancets were use - typical for adult (3) and smaller preferred by diabetics or children (4). Results: Paired T-student test showed no statistically significant difference between measurements 1st and 2nd 95% CI (-4.282 to 0.322), p = 0.09. However, the comparison between 1st/3rd and 1st/4th showed statistically significant difference, 95% CI (-4.964 to -0.796), p = 0,0077 (p < 0,01) and 95% CI (-7.842 to - 2.917), p = 0.00006 (p < 0.001). Conclusions: The procedure of disinfection when fluid completely evaporates has no effect on glucose measurements. Sampling the wet finger influences results of glucose measurement but observed changes were clinically irrelevant. (C) 2021 Primary Care Diabetes Europe. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:848 / 852
页数:5
相关论文
共 50 条
  • [41] Portable detection of Ochratoxin A based on a structure-switching aptamer using a personal glucose meter (PGM)
    Gu, Chunmei
    Shi, Hanchang
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2015, 250
  • [42] Highly sensitive detection of M.SssI DNA methyltransferase activity using a personal glucose meter
    Deng, Huimin
    Peng, Si Ying
    Gao, Zhiqiang
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2016, 408 (21) : 5867 - 5872
  • [43] A Catechol-Meter Based on Conventional Personal Glucose Meter for Portable Detection of Tyrosinase and Sodium Benzoate
    Tian, Tao
    Zhang, Wei-Yi
    Zhou, Hang-Yu
    Peng, Li-Jing
    Zhou, Xi
    Zhang, Hao
    Yang, Feng-Qing
    BIOSENSORS-BASEL, 2022, 12 (12):
  • [44] THE EFFECT OF HEMOLYSIS ON BLOOD-GLUCOSE METER MEASUREMENT
    KILPATRICK, ES
    RUMLEY, AG
    RUMLEY, CN
    DIABETIC MEDICINE, 1995, 12 (04) : 341 - 343
  • [45] Interference in a Glucose Dehydrogenase-Based Glucose Meter
    Kelly, Brian N.
    Haverstick, Doris M.
    Bruns, David E.
    CLINICAL CHEMISTRY, 2010, 56 (06) : 1038 - 1040
  • [46] Diabetic glucose meter for the determination of glucose in microbial cultures
    Flavigny, Raphael
    JOURNAL OF MICROBIOLOGICAL METHODS, 2014, 100 : 91 - 92
  • [47] GLUCOSE-CONCENTRATION MEASUREMENT USING IMMOBILIZED GLUCOSE-OXIDASE
    SPIRKO, K
    LINEK, V
    CERKASOV, J
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1989, 259 (1-2) : 155 - 171
  • [48] ONLINE CONTROL OF GLUCOSE-CONCENTRATION USING AN AUTOMATIC GLUCOSE ANALYZER
    MIZUTANI, S
    IIJIMA, S
    MORIKAWA, M
    SHIMIZU, K
    MATSUBARA, M
    OGAWA, Y
    IZUMI, R
    MATSUMOTO, K
    KOBAYASHI, T
    JOURNAL OF FERMENTATION TECHNOLOGY, 1987, 65 (03): : 325 - 331
  • [49] Correction to: Highly sensitive detection of M.SssI DNA methyltransferase activity using a personal glucose meter
    Huimin Deng
    Si Ying Png
    Zhiqiang Gao
    Analytical and Bioanalytical Chemistry, 2018, 410 : 6033 - 6033
  • [50] Assessing glucose meter accuracy
    Arabadjief, Darius
    Nichols, James H.
    CURRENT MEDICAL RESEARCH AND OPINION, 2006, 22 (11) : 2167 - 2174