The effect of disinfecting procedure on the glucose concentration using a personal glucose meter

被引:3
|
作者
Jonca, Mateusz [1 ]
Krotki, Filipina [1 ]
Tomasik, Przemyslaw [2 ]
机构
[1] Jagiellonian Univ Med Coll, Student Res Grp Clin Biochem, Krakow, Poland
[2] Jagiellonian Univ Med Coll, Pediat Inst, Dept Clin Biochem, Ul Wielicka 265, PL-30663 Krakow, Poland
关键词
Personal glucose meters; Disinfectant; Glucose concentration; POCT; VOLUME;
D O I
10.1016/j.pcd.2021.05.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Personal glucose meters, primarily designed for self-control of glucose concentration in patients with diabetes, are frequently used in clinical practice as point-of-care equipment. The procedure of washing hands with water and soap before testing, as recommended by the manufacturer of personal glucose meters, in hospital wards is often difficult to fulfil and is replaced by disinfecting the place of blood sampling with isopropyl alcohol. Aim: The purpose of the research was to evaluate the effect of different disinfecting procedures on glycemia measurement in capillary blood using personal glucose meters. Materials and methods: Four measurements of glycemia were taken in each of 50-volunteer group using Accu-Check Performa (Roche, Swiss) glucose meter using different procedures: washing hands with soap and drying them (1), disinfection with isopropyl alcohol and waiting for the disinfectant to evaporate (2) and the immediate puncture after disinfection. In the latter case two kind of single-use lancets were use - typical for adult (3) and smaller preferred by diabetics or children (4). Results: Paired T-student test showed no statistically significant difference between measurements 1st and 2nd 95% CI (-4.282 to 0.322), p = 0.09. However, the comparison between 1st/3rd and 1st/4th showed statistically significant difference, 95% CI (-4.964 to -0.796), p = 0,0077 (p < 0,01) and 95% CI (-7.842 to - 2.917), p = 0.00006 (p < 0.001). Conclusions: The procedure of disinfection when fluid completely evaporates has no effect on glucose measurements. Sampling the wet finger influences results of glucose measurement but observed changes were clinically irrelevant. (C) 2021 Primary Care Diabetes Europe. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:848 / 852
页数:5
相关论文
共 50 条
  • [31] EFFECT ON BLOOD GLUCOSE CONCENTRATION OF CHANGES IN AVAILABILITY OF GLUCOSE TO BRAIN
    SAKATA, K
    HAYANO, S
    SLOVITER, HA
    AMERICAN JOURNAL OF PHYSIOLOGY, 1963, 204 (06): : 1127 - &
  • [32] Portable and sensitive detection of non-glucose target by enzyme-encapsulated metal-organic-framework using personal glucose meter
    Cao, Yunzhe
    Mo, Fengye
    Liu, Yahua
    Liu, Yu
    Li, Gaiping
    Yu, Wenqian
    Liu, Xiaoqing
    BIOSENSORS & BIOELECTRONICS, 2022, 198
  • [33] Talking glucose meter for the visually impaired diabetic patient: the effect of haematocrit on glucose measurement
    Peake, R. W. A.
    Devgun, M. S.
    O'Brien, I. A. D.
    BRITISH JOURNAL OF BIOMEDICAL SCIENCE, 2012, 69 (01) : 41 - 42
  • [34] Highly sensitive detection of M.SssI DNA methyltransferase activity using a personal glucose meter
    Huimin Deng
    Si Ying Peng
    Zhiqiang Gao
    Analytical and Bioanalytical Chemistry, 2016, 408 : 5867 - 5872
  • [35] Portable detection of Staphylococcus aureus using personal glucose meter based on hybridization chain reaction strategy
    Yang, Yuemeng
    Wu, Tingting
    Xu, Li-Ping
    Zhang, Xueji
    TALANTA, 2021, 226
  • [36] Label-free and washing-free alkaline phosphatase assay using a personal glucose meter
    Jun Ki Ahn
    Hyo Yong Kim
    Chang Yeol Lee
    Ki Soo Park
    Hyun Gyu Park
    Journal of Biological Engineering, 13
  • [37] Point-of-care testing of methotrexate using a controlled release sensor based on a personal glucose meter
    Yang, Liyu
    Yang, Siyi
    Zhang, Ting
    Zhao, Langyi
    Fa, Huanbao
    Wang, Yongzhong
    Huo, Danqun
    Hou, Changjun
    Yang, Mei
    ANALYTICAL METHODS, 2025, 17 (06) : 1345 - 1353
  • [38] DNAzyme assisted single amplification for FEN1 activity detection using a personal glucose meter
    Li, Mingxi
    Zang, Weijie
    Wang, Shuiliang
    Lin, Zhannuo
    Liu, Shu
    Chen, Yiling
    Ruan, Xiaohui
    Luo, Yuanfang
    Xing, Chao
    Lu, Chunhua
    ANALYTICA CHIMICA ACTA, 2025, 1336
  • [39] Label-free and washing-free alkaline phosphatase assay using a personal glucose meter
    Ahn, Jun Ki
    Kim, Hyo Yong
    Lee, Chang Yeol
    Park, Ki Soo
    Park, Hyun Gyu
    JOURNAL OF BIOLOGICAL ENGINEERING, 2019, 13 (1)
  • [40] CRISPR-Cas12a-driven aptasensor for sensitive detection of alternariol by using a personal glucose meter
    Yang, Ruiqi
    Zhao, Liping
    Fang, Mengqi
    Kong, Weijun
    Luan, Yunxia
    TALANTA, 2025, 286