An in vivo comparison of two frequency-based electronic apex locators

被引:69
|
作者
Welk, AR [1 ]
Baumgartner, JC [1 ]
Marshall, JG [1 ]
机构
[1] Oregon Hlth Sci Univ, Sch Dent, Dept Endodontol, Portland, OR 97201 USA
关键词
D O I
10.1097/00004770-200308000-00002
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to compare the accuracy of a two-frequency (Root ZX) and a five-frequency (Endo Analyzer Model 8005) electronic apex locator under clinical conditions. Thirty-two teeth planned for extraction were used. The coronal portion of each canal was flared using Gates Glidden drills and Orifice Shapers. The canals were irrigated with 2.6% sodium hypochlorite. A K-type file was used to determine a separate working length in each canal using the electronic apex locators. The teeth were extracted and the apical 4 mm of each root canal was exposed along the long axis of the tooth. Photographic slides of each canal were projected and the file position in relation to the minor diameter was determined by two investigators. The mean distance between the electronic apex locator working length and minor diameter was 1.03 mm for the Endo Analyzer and 0.19 mm for the Root ZX. A paired sample t test showed that the Endo Analyzer had significantly longer readings beyond the minor diameter than the Root ZX (p < 0.0001). The ability to locate the minor diameter (+/- 0.5 mm) was 90.7% for the Root ZX and 34.4% for the Endo Analyzer Model 8005.
引用
收藏
页码:497 / 500
页数:4
相关论文
共 50 条
  • [21] Electronic apex locators
    Gordon, MPJ
    Chandler, NP
    INTERNATIONAL ENDODONTIC JOURNAL, 2004, 37 (07) : 425 - 437
  • [22] An in vitro comparison of two modern apex locators
    Weiger, R
    John, C
    Geigle, H
    Löst, C
    JOURNAL OF ENDODONTICS, 1999, 25 (11) : 765 - 768
  • [23] Accuracy of two electronic apex locators in primary teeth
    Bolan, Michele
    de Carvalho Rocha, Maria Jose
    ENDO-ENDODONTIC PRACTICE TODAY, 2010, 4 (01): : 49 - 54
  • [24] Ex vivo comparison of three electronic apex locators to determine working length in endodontic treatment
    Miguita, Kenner Bruno
    de Araujo, Roberta Aranha
    De Martin, Alexandre Sigrist
    da Silveira Bueno, Carlos Eduardo
    Cunha, Rodrigo Sanches
    ENDO-ENDODONTIC PRACTICE TODAY, 2011, 5 (04): : 281 - 284
  • [25] The effect of coronal preflaring on the accuracy of two electronic apex locators
    Suryantoro, R.
    Meidyawati, R.
    Suprastiwi, E.
    1ST PHYSICS AND TECHNOLOGIES IN MEDICINE AND DENTISTRY SYMPOSIUM, 2017, 884
  • [26] An in vivo comparison of accuracy of two electronic apex locators in determining working length using stainless steel and nickel titanium files
    Chaudhary, Sanjeeb
    Gharti, Archana
    Adhikari, Bhawana
    CLINICAL COSMETIC AND INVESTIGATIONAL DENTISTRY, 2018, 10 : 75 - 82
  • [27] An in-vivo investigation of the performance of five electronic apex locators.
    Hooper, SM
    Meredith, N
    JOURNAL OF DENTAL RESEARCH, 1999, 78 (05) : 1061 - 1061
  • [28] Use of electronic apex locators may improve determination of working lengthQuestion: Are electronic apex locators (EALs) accurate?
    Isabel Mello
    Evidence-Based Dentistry, 2014, 15 (4) : 120 - 120
  • [29] The influence of two forms of chlorhexidine on the accuracy of contemporary electronic apex locators
    Marek, Ewa
    Lagocka, Ryta
    Kot, Katarzyna
    Wozniak, Krzysztof
    Lipski, Mariusz
    BMC ORAL HEALTH, 2019, 20 (01)
  • [30] The influence of two forms of chlorhexidine on the accuracy of contemporary electronic apex locators
    Ewa Marek
    Ryta Łagocka
    Katarzyna Kot
    Krzysztof Woźniak
    Mariusz Lipski
    BMC Oral Health, 20