High-resolution monolithic LYSO detector with 6-layer depth-of-interaction for clinical PET

被引:21
|
作者
Stockhoff, Mariele [1 ]
Decuyper, Milan [1 ]
Van Holen, Roel [1 ]
Vandenberghe, Stefaan [1 ]
机构
[1] Univ Ghent, Dept Elect & Informat Syst, Med Image & Signal Proc MEDISIP, Ghent, Belgium
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2021年 / 66卷 / 15期
关键词
PET; TB-PET; monolithic detector; high-resolution; nearest neighbour; neural networks; CALIBRATION PROCEDURE; POSITION ESTIMATION; SPATIAL-RESOLUTION; NEURAL-NETWORK; TOF-PET; CRYSTALS; DOI;
D O I
10.1088/1361-6560/ac1459
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The system spatial resolution of whole-body positron emission tomography (PET) is limited to around 2 mm due to positron physics and the large diameter of the bore. To stay below this 'physics'-limit a scintillation detector with an intrinsic spatial resolution of around 1.3 mm is needed. Currently used detector technology consists of arrays of 2.6-5 mm segmented scintillator pixels which are the dominant factor contributing to the system resolution. Pixelated detectors using smaller pixels exist but face major drawbacks in sensitivity, timing, energy resolution and cost. Monolithic continuous detectors, where the spatial resolution is determined by the shape of the light distribution on the photodetector array, are a promising alternative. Without having the drawbacks of pixelated detectors, monolithic ones can also provide depth-of-interaction (DOI) information. In this work we present a monolithic detector design aiming to serve high-resolution clinical PET systems while maintaining high sensitivity. A 50 x 50 x 16 mm(3) Lutetium-Yttrium oxyorthosilicate scintillation crystal with silicon photomultiplier (SiPM) backside readout is calibrated in singles mode by a collimated beam obtaining a reference dataset for the event positioning. A mean nearest neighbour (MNN) algorithm and an artificial neural network for positioning are compared. The targeted intrinsic detector resolution of 1.3 mm needed to reach a 2 mm resolution on system level was accomplished with both algorithms. The neural network achieved a mean spatial resolution of 1.14 mm FWHM for the whole detector and 1.02 mm in the centre (30 x 30 mm(2)). The MNN algorithm performed slightly worse with 1.17 mm for the whole detector and 1.13 mm in the centre. The intrinsic DOI information will also result in uniform system spatial resolution over the full field of view.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Development of a High-Resolution and Depth-of-Interaction Capable Detector for Time-of-Flight PET
    Krishnamoorthy, Srilalan
    Wiener, Rony I.
    Kaul, Madhuri
    Panetta, Joseph
    Karp, Joel S.
    Surti, Suleman
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2457 - 2461
  • [2] Scalable, Time-of-Flight and Depth-of-Interaction Detector Units for High-Resolution PET Systems
    Nadig, Vanessa
    Gundacker, Stefan
    Schug, David
    Herweg, Katrin
    Weindel, Konstantin
    Radermacher, Harald
    Mueller, Florian
    Weissler, Bjoern
    Schulz, Volkmar
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2024, 8 (01) : 1 - 14
  • [3] A PET detector module with monolithic crystal, single end readout, SiPM array and high depth-of-interaction resolution
    Zhang, H.
    Zhou, R.
    Yang, C.
    JOURNAL OF INSTRUMENTATION, 2016, 11
  • [4] A PET DETECTOR WITH DEPTH-OF-INTERACTION DETERMINATION
    BARTZAKOS, P
    THOMPSON, CJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (06): : 735 - 748
  • [5] Spatial resolution of a small cubic LYSO scintillator crystal detector with depth-of-interaction capabilities in a small animal PET scanner
    Perera, Lakshal
    Lerch, Michael
    Rosenfeld, Anatoly
    Meikle, Steve R.
    Chan, Jan
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3754 - +
  • [6] Evaluation of a prototype PET detector with depth-of-interaction encoding
    Burr, KC
    Ivan, A
    Castleberry, DE
    LeBlanc, JW
    Shah, KS
    Farrell, R
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 877 - 881
  • [7] DMice - a depth-of-interaction detector design for PET scanners
    Lewellen, TK
    Janes, M
    Miyaoka, RS
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2388 - 2392
  • [8] High spatial resolution measurement of depth-of-interaction of a PET LSO crystal
    Simon, A
    Balkay, L
    Kalinka, G
    Kerek, A
    Novák, D
    Sipos, A
    Végh, J
    Trón, L
    Molnár, J
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 546 (1-2): : 33 - 36
  • [9] Preliminary performance evaluation of a high resolution small animal PET scanner with monolithic crystals and depth-of-interaction encoding
    Balcerzyk, Marcin
    Kontaxakis, George
    Delgado, Mercedes
    Garcia, Luis
    Benlloch, Jose M.
    Pozo, Miguel A.
    8TH IEEE INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOENGINEERING, VOLS 1 AND 2, 2008, : 944 - +
  • [10] Four-layer depth-of-interaction PET detector for high resolution PET using a multi-pixel S8550 avalanche photodiode
    Nishikido, Fumihiko
    Inadama, Naoko
    Oda, Ichiro
    Shibuya, Kengo
    Yoshida, Eiji
    Yamaya, Taiga
    Kitamura, Keishi
    Murayama, Hideo
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3): : 570 - 575