Mitigating voltage lead errors of an AC Josephson voltage standard by impedance matching

被引:16
|
作者
Zhao, Dongsheng [1 ]
van den Brom, Helko E. [1 ]
Houtzager, Ernest [1 ]
机构
[1] VSL BV, POB 654, NL-2600 AR Delft, Netherlands
关键词
AC Josephson voltage standard; impedance matching; simulation; error sources; uncertainty; CALIBRATIONS;
D O I
10.1088/1361-6501/aa7aba
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A pulse-driven AC Josephson voltage standard (ACJVS) generates calculable AC voltage signals at low temperatures, whereas measurements are performed with a device under test (DUT) at room temperature. The voltage leads cause the output voltage to show deviations that scale with the frequency squared. Error correction mechanisms investigated so far allow the ACJVS to be operational for frequencies up to 100 kHz. In this paper, calculations are presented to deal with these errors in terms of reflected waves. Impedance matching at the source side of the system, which is loaded with a high-impedance DUT, is proposed as an accurate method to mitigate these errors for frequencies up to 1 MHz. Simulations show that the influence of non-ideal component characteristics, such as the tolerance of the matching resistor, the capacitance of the load input impedance, losses in the voltage leads, non-homogeneity in the voltage leads, a non-ideal on-chip connection and inductors between the Josephson junction array and the voltage leads, can be corrected for using the proposed procedures. The results show that an expanded uncertainty of 12 parts in 10(6) (k = 2) at 1 MHz and 0.5 part in 10(6) (k = 2) at 100 kHz is within reach.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Thermal Transfer Standard Calibrations Using a Pulse-Driven AC Josephson Voltage Standard
    Zhou, Kunli
    Qu, Jifeng
    Xu, Xun
    Zhou, Zhenyu
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [42] The BIPM Compact Josephson Voltage Standard
    Solve, Stephane
    Chayramy, Regis
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (07) : 2366 - 2371
  • [43] Progress in Josephson voltage standard development
    Hassel, J
    Seppä, H
    Grönberg, L
    Suni, I
    Helistö, P
    [J]. 2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 399 - 400
  • [44] Harmonic Voltage Measurement based on Program Josephson Voltage Standard
    Jia, Zhengsen
    Xu, Xitong
    Wang, Lei
    Zhang, Jiangtao
    Pan, Xianlin
    Shi, Zhaomin
    [J]. 2020 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM), 2020,
  • [45] THE BIPM COMPACT JOSEPHSON VOLTAGE STANDARD
    Solve, S.
    Chayramy, R.
    [J]. 2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, 2010, : 155 - 156
  • [46] New Josephson voltage standard proposed
    不详
    [J]. JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1996, 101 (03): : 395 - 396
  • [47] Compact DC Josephson Voltage Standard
    Rufenacht, Alain
    Fox, Anna E.
    Butler, Grace E.
    Burroughs, Charles J.
    Dresselhaus, Paul D.
    Schwall, Robert E.
    Cular, Stefan
    Benz, Samuel P.
    [J]. 2020 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM), 2020,
  • [48] Portable Josephson Voltage Reference Standard
    Katkov, A.
    Lovtsus, V.
    Behr, R.
    [J]. 2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,
  • [49] STANDARD MEASURE OF AC VOLTAGE
    ROITMAN, MS
    SVINOLUPOV, YG
    SUSHENTSEV, AN
    PODKOPAEV, NN
    POLSTVIN, VI
    LAUD, TA
    [J]. MEASUREMENT TECHNIQUES USSR, 1984, 27 (05): : 443 - 446
  • [50] A compact transportable Josephson voltage standard
    Hamilton, CA
    Burroughs, CJ
    Kupferman, SL
    Naujoks, GA
    Vickery, A
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (02) : 237 - 241