Mitigating voltage lead errors of an AC Josephson voltage standard by impedance matching

被引:16
|
作者
Zhao, Dongsheng [1 ]
van den Brom, Helko E. [1 ]
Houtzager, Ernest [1 ]
机构
[1] VSL BV, POB 654, NL-2600 AR Delft, Netherlands
关键词
AC Josephson voltage standard; impedance matching; simulation; error sources; uncertainty; CALIBRATIONS;
D O I
10.1088/1361-6501/aa7aba
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A pulse-driven AC Josephson voltage standard (ACJVS) generates calculable AC voltage signals at low temperatures, whereas measurements are performed with a device under test (DUT) at room temperature. The voltage leads cause the output voltage to show deviations that scale with the frequency squared. Error correction mechanisms investigated so far allow the ACJVS to be operational for frequencies up to 100 kHz. In this paper, calculations are presented to deal with these errors in terms of reflected waves. Impedance matching at the source side of the system, which is loaded with a high-impedance DUT, is proposed as an accurate method to mitigate these errors for frequencies up to 1 MHz. Simulations show that the influence of non-ideal component characteristics, such as the tolerance of the matching resistor, the capacitance of the load input impedance, losses in the voltage leads, non-homogeneity in the voltage leads, a non-ideal on-chip connection and inductors between the Josephson junction array and the voltage leads, can be corrected for using the proposed procedures. The results show that an expanded uncertainty of 12 parts in 10(6) (k = 2) at 1 MHz and 0.5 part in 10(6) (k = 2) at 100 kHz is within reach.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] INFLUENCE OF VOLTAGE DIVIDER ON JOSEPHSON VOLTAGE STANDARD SYSTEM
    SHIDA, K
    MURAYAMA, Y
    ENDO, T
    NAKAMURA, A
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS, 1978, 17 (07) : 1299 - 1300
  • [32] Extending voltage range to 10 V rms in AC-DC difference measurements with AC programmable Josephson voltage standard
    Amagai, Yasutaka
    Maruyama, Michitaka
    Yamamori, Hirotake
    Shimazaki, Takeshi
    Okawa, Kenjiro
    Fujiki, Hiroyuki
    Kaneko, Nobu-Hisa
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2020, 31 (06)
  • [33] Josephson standards for AC voltage metrology
    Benz, SP
    Hamilton, CA
    Burroughs, CJ
    Christian, LA
    [J]. SUCCESS IN THE 21ST CENTURY DEPENDS ON MODERN METROLOGY, VOLS 1-2, 1997, : 355 - 363
  • [34] Systematic-error signals in the AC Josephson voltage standard: Measurement and reduction
    Landim, Regis Pinheiro
    Benz, Samuel P.
    Dresselhaus, Paul D.
    Burroughs, Charles J., Jr.
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2008, 57 (06) : 1215 - 1220
  • [35] Low Crest Factor Multitone Waveform Synthesis with the AC Josephson Voltage Standard
    Zhou, Kunli
    Qu, Jifeng
    Dong, Xianying
    [J]. 2015 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2015, : 556 - 559
  • [36] An Improved Sinc Function Compensation Algorithm in AC Programmable Josephson Voltage Standard
    Jia Zhengsen
    Wang Lei
    Wang Zengmin
    Peng Xiaojuan
    Huang Hongtao
    Liu Lijuan
    Lu Zuliang
    [J]. 2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,
  • [37] An AC Josephson Voltage Standard up to the Kilohertz Range Tested in a Calibration Laboratory
    Schubert, Marco
    Starkloff, Michael
    Lee, Jinni
    Behr, Ralf
    Palafox, Luis
    Wintermeier, Alexander
    Boeck, Andreas C.
    Fleischmann, Philip M.
    May, Torsten
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2015, 64 (06) : 1620 - 1626
  • [38] Modeling the Nonlinear Amplifier with the Pulse-Driven AC Josephson Voltage Standard
    Zhou, Kunli
    Qu, Jifeng
    Zhou, Zhenyu
    Xu, Xun
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [39] Uncertainties in the Measurement of AC Voltage Using a Programmable Josephson Voltage Standard and a Phase-Sensitive Null Detector
    Georgakopoulos, Dimitrios
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (07) : 2178 - 2183
  • [40] Thermal Transfer Standard Calibrations Using a Pulse-Driven AC Josephson Voltage Standard
    Zhou, Kunli
    Qu, Jifeng
    Xu, Xun
    Zhou, Zhenyu
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,