Compact DC Josephson Voltage Standard

被引:0
|
作者
Rufenacht, Alain [1 ]
Fox, Anna E. [1 ]
Butler, Grace E. [1 ]
Burroughs, Charles J. [1 ]
Dresselhaus, Paul D. [1 ]
Schwall, Robert E. [1 ]
Cular, Stefan [2 ]
Benz, Samuel P. [1 ]
机构
[1] NIST, Boulder, CO 80305 USA
[2] NIST, Gaithersburg, MD 20899 USA
关键词
Digital-analog conversion; Josephson arrays; standards; superconducting integrated circuits; voltage measurement;
D O I
10.1109/cpem49742.2020.9191805
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents precision measurements with a prototype cryogen-liquid-free DC Josephson voltage standard that produces a 1 V maximum output. Its cryostat is sufficiently compact that it can be operated on a bench close to devices under test. Compared to the National Institute of Standards and Technology (NIST) programmable Josephson voltage standard (PJVS), this bench-top Josephson voltage standard (BJVS) has a smaller cryostat (only 25 % of the volume), the Josephson junction array circuit is less complex and has only 10 % as many junctions, simpler bias electronics, and reduced mechanical cooling requirements. Direct voltage comparison between the BJVS and the PJVS achieved a relative agreement of 5 x 10(-10) at 1 V. With the recent redefinition of the International System of Units (SI), the BJVS becomes a primary realization the unit volt, requiring only access to a standard single-phase power outlet and a GPS signal (SI second).
引用
收藏
页数:2
相关论文
共 50 条
  • [1] The BIPM Compact Josephson Voltage Standard
    Solve, Stephane
    Chayramy, Regis
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (07) : 2366 - 2371
  • [2] THE BIPM COMPACT JOSEPHSON VOLTAGE STANDARD
    Solve, S.
    Chayramy, R.
    [J]. 2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, 2010, : 155 - 156
  • [3] A compact transportable Josephson voltage standard
    Hamilton, CA
    Burroughs, CJ
    Kupferman, SL
    Naujoks, GA
    Vickery, A
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (02) : 237 - 241
  • [4] Direct Comparison of DC Josephson Voltage Standard and Programmable Josephson Voltage Standard of TUBITAK UME
    Turhan, Saliha
    Kanatoglu, Naylan
    Orhan, Recep
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [5] Joint program develops compact Josephson voltage standard
    不详
    [J]. JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1996, 101 (06): : 834 - 834
  • [6] 1 volt DC programmable Josephson voltage standard
    Burroughs, CJ
    Benz, SP
    Harvey, TE
    Hamilton, CA
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1999, 9 (02) : 4145 - 4149
  • [7] Traceability of DC high voltage measurements using the Josephson voltage standard
    Mageed, Hala M. Abdel
    El-Rifaie, Ali M.
    Aladdin, Omar M.
    [J]. MEASUREMENT, 2014, 58 : 269 - 273
  • [8] DC Comparison of a Programmable Josephson Voltage Standard and a Josephson Arbitrary Waveform Synthesizer
    Rufenacht, Alain
    Flowers-Jacobs, Nathan E.
    Fox, Anna E.
    Waltman, Steven B.
    Schwall, Robert E.
    Burroughs, Charles J.
    Dresselhaus, Paul D.
    Benz, Samuel P.
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [9] Direct comparison between the NIST 10 V Compact Josephson Voltage Standard and the 2.5 V Programmable Josephson Voltage Standard
    Jaiswal, Shiv Kumar
    Tang, Y.
    [J]. MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2009, 24 (04): : 207 - 214
  • [10] Deployment of a compact, transportable, fully automated Josephson voltage standard
    Kupferman, SL
    Hamilton, CA
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1999, 48 (02) : 249 - 252