Whole-Brain Cerebral Blood Flow Mapping Using 3D Echo Planar Imaging and Pulsed Arterial Tagging

被引:22
|
作者
Gai, Neville D. [1 ]
Talagala, S. Lalith [2 ]
Butman, John A. [1 ]
机构
[1] NIH, Ctr Clin, Bethesda, MD 20892 USA
[2] NINDS, NIH MRI Res Facil, NIH, Bethesda, MD 20892 USA
基金
美国国家卫生研究院;
关键词
arterial spin labeling; 3D PULSAR; blurring reduction; whole brain perfusion imaging; LABELING PERFUSION MEASUREMENTS; MRI; SIGNAL; INVERSION;
D O I
10.1002/jmri.22437
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To quantitate cerebral blood flow (CBF) in the entire brain using the 3D echo planar imaging (EPI) PULSAR (pulsed star labeling) technique. Materials and Methods: The PULSAR technique was modified to 1) incorporate a nonselective inversion pulse to suppress background signal; 2) to use 3D EPI acquisition; and 3) to modulate flip angle in such a manner as to minimize the blurring resulting from T1 modulation along the slice encoding direction. Computation of CBF was performed using the general kinetic model (GKM). In a series of healthy volunteers (n = 12), we first investigated the effects of introducing an inversion pulse on the measured value of CBF and on the temporal stability of the perfusion signal. Next we investigated the effect of flip angle modulation on the spatial blurring of the perfusion signal. Finally, we evaluated the repeatability of the CBF measurements, including the influence of the measurement of arterial blood magnetization (a calibration factor for the GKM). Results: The sequence provides sufficient perfusion signal to achieve whole brain coverage in approximate to 5 minutes. Introduction of the inversion pulse for background suppression did not significantly affect computed CBF values, but did reduce the fluctuation in the perfusion signal. Flip angle modulation reduced blurring, resulting in higher estimates of gray matter (GM) CBF and lower estimates of white matter (WM) CBF. The repeatability study showed that measurement of arterial blood signal did not result in significantly higher error in the perfusion measurement. Conclusion: Improvements in acquisition and sequence preparation presented here allow for better quantification and localization of perfusion signal, allowing for accurate whole-brain CBF measurements in 5 minutes.
引用
收藏
页码:287 / 295
页数:9
相关论文
共 50 条
  • [31] Segmented spin-echo echo-planar imaging improves whole-brain BOLD functional MRI in awake pigeon brains
    Khodadadi, Mina
    Helluy, Xavier
    Guentuerkuen, Onur
    Behroozi, Mehdi
    NMR IN BIOMEDICINE, 2024, 37 (01)
  • [32] Whole-brain snapshot CEST imaging at 7 T using 3D-EPI
    Akbey, Suzan
    Ehses, Philipp
    Stirnberg, Ruediger
    Zaiss, Moritz
    Stoecker, Tony
    MAGNETIC RESONANCE IN MEDICINE, 2019, 82 (05) : 1741 - 1752
  • [33] Three-dimensional echo-planar cine imaging of cerebral blood supply using arterial spin labeling
    Shrestha, Manoj
    Mildner, Toralf
    Schlumm, Torsten
    Robertson, Scott Haile
    Moeller, Harald
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2016, 29 (06) : 799 - 810
  • [34] Three-dimensional echo-planar cine imaging of cerebral blood supply using arterial spin labeling
    Manoj Shrestha
    Toralf Mildner
    Torsten Schlumm
    Scott Haile Robertson
    Harald Möller
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2016, 29 : 799 - 810
  • [35] Improved velocity-selective-inversion arterial spin labeling for cerebral blood flow mapping with 3D acquisition
    Liu, Dapeng
    Xu, Feng
    Li, Wenbo
    van Zijl, Peter C.
    Lin, Doris D.
    Qin, Qin
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (05) : 2512 - 2522
  • [36] Fast 3D magnetic resonance fingerprinting for a whole-brain coverage
    Ma, Dan
    Jiang, Yun
    Chen, Yong
    McGivney, Debra
    Mehta, Bhairav
    Gulani, Vikas
    Griswold, Mark
    MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (04) : 2190 - 2197
  • [37] 3D Nucleus Instance Segmentation for Whole-Brain Microscopy Images
    Ma, Junbo
    Krupa, Oleh
    Glass, Madison Rose
    McCormick, Carolyn M.
    Borland, David
    Kim, Minjeong
    Stein, Jason L.
    Wu, Guorong
    INFORMATION PROCESSING IN MEDICAL IMAGING, IPMI 2021, 2021, 12729 : 504 - 516
  • [38] Fast high-resolution 3D segmented echo planar imaging for dose mapping using a superheated emulsion chamber
    Lamba, M
    Holland, SK
    Schmithorst, V
    Dardzinski, B
    d'Errico, F
    Nath, R
    MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (04) : 675 - 681
  • [39] 3D whole-brain metabolite imaging to improve characterization of low-to-intermediate grade gliomas
    Zhong, Jim
    Huang, Vicki
    Gurbani, Saumya S.
    Ramesh, Karthik
    Scott Cordova, J.
    Schreibmann, Eduard
    Shu, Hui-Kuo G.
    Olson, Jeffrey
    Han, Hui
    Giuffrida, Alexander
    Shim, Hyunsuk
    Weinberg, Brent D.
    JOURNAL OF NEURO-ONCOLOGY, 2021, 153 (02) : 303 - 311
  • [40] Optimized interleaved whole-brain 3D double inversion recovery (DIR) sequence for imaging the neocortex
    Boulby, PA
    Symms, MR
    Barker, GJ
    MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (06) : 1181 - 1186