Safety verification of carbon-ion radiotherapy for patients with cardiac implantable electronic devices (CIEDs)

被引:2
|
作者
Okano, Naoko [1 ,2 ]
Sakai, Makoto [2 ]
Shibuya, Kei [2 ]
Tsuda, Kazuhisa [3 ]
Kanzaki, Takao [3 ]
Sano, Masato [4 ]
Kaneko, Yoshiaki [5 ]
Ohno, Tatsuya [1 ,2 ]
机构
[1] Gunma Univ, Dept Radiat Oncol, Grad Sch Med, 3-39-15 Showa Machi, Maebashi, Gunma 3718511, Japan
[2] Gunma Univ, Heavy Ion Med Ctr, 3-39-15 Showa Machi, Maebashi, Gunma 3718511, Japan
[3] Gunma Univ Hosp, Dept Radiol, 3-9-15 Showa Machi, Maebashi, Gunma 3718511, Japan
[4] Gunma Univ Hosp, Me Supply Ctr, 3-9-15 Showa Machi, Maebashi, Gunma 3718511, Japan
[5] Gunma Univ, Dept Cardiovasc Med, Grad Sch Med, 3-39-15 Showa Machi, Maebashi, Gunma 3718511, Japan
关键词
pacemaker; cardiac implantable electronic devices (CIEDs); implantable cardioverter-defibrillator (ICD); carbon-ion; radiotherapy; Monte Carlo simulation; DEGRO/DGK-GUIDELINE; LATEST GENERATION; RADIATION; MANAGEMENT; IRRADIATION; EXPOSURE; SYSTEM; ENERGY;
D O I
10.1093/jrr/rrab105
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
According to guidelines, carbon-ion beam therapy is considered to carry a high safety risk for patients with cardiac implantable electronic devices (CIEDs), although the actual impacts remain unclear. In this study, we investigated the safety of carbon-ion beam therapy in patients with CIEDs. Patients with CIEDs who underwent carbon-ion therapy at Gunma University Heavy Ion Medical Center between June 2010 and December 2019 were identified and investigated for abnormalities in the operation of their CIEDs, such as oversensing and resetting during irradiation, and abnormalities in operation after treatment. In addition, the risk of irradiation from carbon-ion beam therapy was evaluated by model simulations. Twenty patients (22 sites) with CIEDs were identified, 19 with pacemakers and one with an implantable cardioverter-defibrillator (ICD). Treatments were completed without any problems, except for one case in which the treatment was discontinued because of worsening of the primary disease. Monte Carlo simulation indicated that the carbon beam irradiation produced neutrons at a constant and high level in the irradiation field. Nevertheless, with the distances between the CIEDs and the irradiation fields in the analyzed cases, the quantity of neutrons at the CIEDs was lower than that within the irradiation. Although carbon-ion beam therapy can be safely administered to patients with CIEDs, it is advisable to perform the therapy with sufficient preparation and backup devices because of the risks involved.
引用
收藏
页码:122 / 127
页数:6
相关论文
共 50 条
  • [1] Recommendations for Imaging Patients With Cardiac Implantable Electronic Devices (CIEDs)
    Vigen, Karl K.
    Reeder, Scott B.
    Hood, Maureen N.
    Steckner, Michael
    Leiner, Tim
    Dombroski, David A.
    Gulani, Vikas
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2021, 53 (05) : 1311 - 1317
  • [2] Cardiac implantable electronic devices (CIEDs) and allergy
    De Ruvo, Ermenegildo
    Petrungaro, Mattia
    De Luca, Lucia
    Bressi, Edoardo
    Bruni, Giuseppe
    Fagagnini, Alessandro
    Grieco, Domenico
    Martino, Anna Maria
    Panuccio, Marco
    Politano, Alessandro
    Rebecchi, Marco
    Calo, Leonardo
    [J]. JOURNAL OF ARRHYTHMIA, 2023, 39 (03) : 310 - 314
  • [3] Radiotherapy (RT) and cardiac implantable electronic devices (CIEDs): a prospective study.
    Villa, E.
    Michelotti, F.
    Motta, M.
    Ravasio, A.
    Gherardi, D.
    Manazzale, V.
    Locatelli, F.
    Trombetta, L.
    Salmoiraghi, P.
    Mascioli, G.
    Vavassori, V.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S612 - S612
  • [4] Monitoring Cardiovascular Implantable Electronic Devices (CIEDs) for Patients Receiving Radiotherapy (RT)
    Bowman, K.
    Popple, R.
    Doppalapudi, H.
    Fiveash, J. B.
    Meredith, R.
    Brezovich, I. A.
    Bonner, J. A.
    Spencer, S. A.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S599 - S599
  • [5] Percutaneous Extraction of Cardiac Implantable Electronic Devices (CIEDs) in Octogenarians
    Williams, Steven E.
    Arujuna, Aruna
    Whitaker, John
    Sohal, Manav
    Shetty, Anoop K.
    Roy, Debashis
    Bostock, Julian
    Cooklin, Michael
    Gill, Jaswinder
    O'Neill, Mark
    Wright, Matthew
    Patel, Nikhil
    Bucknall, Cliff
    Hamid, Shoaib
    Rinaldi, C. Aldo
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (07): : 841 - 849
  • [6] Safety of particle radiotherapy in patients with cardiac implantable electronic devices: Review of literature
    Gupta, Amulya
    Glein, Rachel
    Hossain, Nabil
    Sundhu, Murtaza
    Rotondo, Ronny
    Sheldon, Seth H.
    Noheria, Amit
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2024,
  • [7] Effects of Ionizing Radiation on Cardiac Implantable Electronic Devices (CIEDs) in Patients with Esophageal Cancer Undergoing Radiotherapy: A Pilot Study
    Uzun, Davut D.
    Salatzki, Janek
    Xynogalos, Panagiotis
    Frey, Norbert
    Debus, Juergen
    Lang, Kristin
    [J]. CANCERS, 2024, 16 (03)
  • [8] Management of Radiotherapy Patients with Cardiac Implantable Electronic Devices
    Miften, Moyed
    Mihailidis, D.
    Kry, S.
    Miften, M.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3240 - 3240
  • [9] Safety of electronic massagers in patients with cardiac implantable electronic devices
    Chia, Pow-Li
    Mok, Kwang-How
    Wong, Shiun Woei
    Foo, David
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2021, 44 (01): : 167 - 170
  • [10] MR Imaging of Patients with Cardiac Implantable Electronic Devices (CIEDs): Implementing a Program and Optimizing CMR
    Gulhane A.
    Litt H.
    [J]. Current Radiology Reports, 6 (11)