Safety verification of carbon-ion radiotherapy for patients with cardiac implantable electronic devices (CIEDs)

被引:2
|
作者
Okano, Naoko [1 ,2 ]
Sakai, Makoto [2 ]
Shibuya, Kei [2 ]
Tsuda, Kazuhisa [3 ]
Kanzaki, Takao [3 ]
Sano, Masato [4 ]
Kaneko, Yoshiaki [5 ]
Ohno, Tatsuya [1 ,2 ]
机构
[1] Gunma Univ, Dept Radiat Oncol, Grad Sch Med, 3-39-15 Showa Machi, Maebashi, Gunma 3718511, Japan
[2] Gunma Univ, Heavy Ion Med Ctr, 3-39-15 Showa Machi, Maebashi, Gunma 3718511, Japan
[3] Gunma Univ Hosp, Dept Radiol, 3-9-15 Showa Machi, Maebashi, Gunma 3718511, Japan
[4] Gunma Univ Hosp, Me Supply Ctr, 3-9-15 Showa Machi, Maebashi, Gunma 3718511, Japan
[5] Gunma Univ, Dept Cardiovasc Med, Grad Sch Med, 3-39-15 Showa Machi, Maebashi, Gunma 3718511, Japan
关键词
pacemaker; cardiac implantable electronic devices (CIEDs); implantable cardioverter-defibrillator (ICD); carbon-ion; radiotherapy; Monte Carlo simulation; DEGRO/DGK-GUIDELINE; LATEST GENERATION; RADIATION; MANAGEMENT; IRRADIATION; EXPOSURE; SYSTEM; ENERGY;
D O I
10.1093/jrr/rrab105
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
According to guidelines, carbon-ion beam therapy is considered to carry a high safety risk for patients with cardiac implantable electronic devices (CIEDs), although the actual impacts remain unclear. In this study, we investigated the safety of carbon-ion beam therapy in patients with CIEDs. Patients with CIEDs who underwent carbon-ion therapy at Gunma University Heavy Ion Medical Center between June 2010 and December 2019 were identified and investigated for abnormalities in the operation of their CIEDs, such as oversensing and resetting during irradiation, and abnormalities in operation after treatment. In addition, the risk of irradiation from carbon-ion beam therapy was evaluated by model simulations. Twenty patients (22 sites) with CIEDs were identified, 19 with pacemakers and one with an implantable cardioverter-defibrillator (ICD). Treatments were completed without any problems, except for one case in which the treatment was discontinued because of worsening of the primary disease. Monte Carlo simulation indicated that the carbon beam irradiation produced neutrons at a constant and high level in the irradiation field. Nevertheless, with the distances between the CIEDs and the irradiation fields in the analyzed cases, the quantity of neutrons at the CIEDs was lower than that within the irradiation. Although carbon-ion beam therapy can be safely administered to patients with CIEDs, it is advisable to perform the therapy with sufficient preparation and backup devices because of the risks involved.
引用
收藏
页码:122 / 127
页数:6
相关论文
共 50 条
  • [31] Magnetic resonance imaging safety in patients with cardiac implantable electronic devices
    Yang, Eunice
    Suzuki, Masahito
    Nazarian, Saman
    Halperin, Henry R.
    [J]. TRENDS IN CARDIOVASCULAR MEDICINE, 2022, 32 (07) : 440 - 447
  • [32] Safety and outcomes of noncardiac surgery in patients with cardiac implantable electronic devices
    Akhondi, Bahman
    Ansari-Ramandi, Mohammadmostafa
    Heidarali, Mona
    Kamali, Farzad
    Haghjoo, Majid
    [J]. RESEARCH IN CARDIOVASCULAR MEDICINE, 2019, 8 (03) : 76 - 78
  • [33] Safety of repetitive nerve stimulation in patients with cardiac implantable electronic devices
    Cronin, Edmond M.
    Gray, Jennifer
    Abi-Saleh, Bernard
    Wilkoff, Bruce L.
    Levin, Kerry H.
    [J]. MUSCLE & NERVE, 2013, 47 (06) : 840 - 844
  • [34] Evaluation of Safety for Scanning Carbon-Ion Radiotherapy in Hemodialysis Patients With Prostate Cancer
    Takakusagi, Yosuke
    Suga, Makito
    Kusano, Yohsuke
    Kano, Kio
    Shima, Satoshi
    Tsuchida, Keisuke
    Mizoguchi, Nobutaka
    Serizawa, Itsuko
    Yoshida, Daisaku
    Kamada, Tadashi
    Minohara, Shinichi
    Katoh, Hiroyuki
    [J]. CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (02)
  • [35] Cardiac Implantable Electronic Devices and Radiotherapy with 6 MV Photons: Are the Patients Safe?
    Blamek, S.
    Konefal, A.
    Wronska, A.
    Gabrys, D.
    Orlef, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S1007 - S1007
  • [36] Radiotherapy-induced malfunctions of cardiac implantable electronic devices in cancer patients
    Vincenzo L. Malavasi
    Giuseppina De Marco
    Jacopo F. Imberti
    Filippo Placentino
    Marco Vitolo
    Ercole Mazzeo
    Gianfranco Cicoria
    Edoardo Casali
    Vincenzo Turco
    Frank Lohr
    Giuseppe Boriani
    [J]. Internal and Emergency Medicine, 2020, 15 : 967 - 973
  • [37] Radiotherapy-induced malfunction in cancer patients with cardiac implantable electronic devices
    Imberti, J. F.
    Placentino, F.
    Malavasi, V. L.
    Demarco, G.
    Lohr, F.
    Boriani, G.
    [J]. EUROPEAN HEART JOURNAL, 2019, 40 : 290 - 290
  • [38] Radiotherapy-induced malfunctions of cardiac implantable electronic devices in cancer patients
    Malavasi, Vincenzo L.
    De Marco, Giuseppina
    Imberti, Jacopo F.
    Placentino, Filippo
    Vitolo, Marco
    Mazzeo, Ercole
    Cicoria, Gianfranco
    Casali, Edoardo
    Turco, Vincenzo
    Lohr, Frank
    Boriani, Giuseppe
    [J]. INTERNAL AND EMERGENCY MEDICINE, 2020, 15 (06) : 967 - 973
  • [39] Provision of MR imaging for patients with cardiac implantable electronic devices (CIEDs): a single-center experience and national survey
    Murray, Aoife S.
    Gilligan, Paddy J.
    Bisset, James M.
    Nolan, Chris
    Galvin, Joseph M.
    Murray, John G.
    [J]. IRISH JOURNAL OF MEDICAL SCIENCE, 2019, 188 (03) : 999 - 1004
  • [40] Provision of MR imaging for patients with cardiac implantable electronic devices (CIEDs): a single-center experience and national survey
    Aoife S. Murray
    Paddy J. Gilligan
    James M. Bisset
    Chris Nolan
    Joseph M. Galvin
    John G. Murray
    [J]. Irish Journal of Medical Science (1971 -), 2019, 188 : 999 - 1004