Performance of a real time selenium based x-ray detector for fluoroscopy

被引:24
|
作者
Choquette, M [1 ]
Demers, Y [1 ]
Shukri, Z [1 ]
Tousignant, O [1 ]
Aoki, K [1 ]
Honda, M [1 ]
Takahashi, A [1 ]
Tsukamoto, A [1 ]
机构
[1] ANRAD Corp, St Laurent, PQ H4R 2P1, Canada
关键词
selenium; flat-panel detector; TFT; fluoroscopy; real-time; dynamic; direct conversion; x-ray detector;
D O I
10.1117/12.430873
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
As amorphous selenium based flat panel detectors gain more interest for direct, real-time x-ray imaging, we report in this paper the performance of such a detector by ANRAD Corporation. This new detector is based on a 1536 x 1536 array of amorphous silicon TFT pixels coupled with a 1000 mum selenium converter biased at 10 V/mum. Each 150 mum x 150 mum pixel is made of a thin film transistor (TFT), a storage capacitor and a collecting electrode having a geometrical fill factor of 77% and an effective fill factor of nearby 100%. New TFT architecture and high speed electronics allow operation of this new detector under a large range of conditions from low dose fluoroscopy (few muR/frame) to DR and it supports frame rates from single shot to 60 fps. Image lag following x-ray excitation was measured to be very low. Key imaging parameters for this real time x-ray detection system such as: detective quantum efficiency (DQE), signal to noise ratio (SNR) and modulation transfer function (MTF) were measured. These results together with frames extracted from a low dose fluoroscopy sequence and an anthropomorphic phantom image will be presented. The measurements show that the dynamic performance of this new detector is very well suited for real time imaging in fluoroscopic and/or cardiac applications.
引用
收藏
页码:501 / 508
页数:8
相关论文
共 50 条
  • [41] THE X-RAY FOVEA, A DEVICE FOR REDUCING X-RAY DOSE IN FLUOROSCOPY
    LABBE, MS
    CHIU, MY
    RZESZOTARSKI, MS
    BANIHASHEMI, AR
    WILSON, DL
    MEDICAL PHYSICS, 1994, 21 (03) : 471 - 481
  • [42] Development of a large-area CMOS-based detector for real-time x-ray imaging
    Heo, Sung Kyn
    Park, Sung Kyu
    Hwang, Sung Ha
    Im, Dong Ak
    Kosonen, Jari
    Kim, Tae Woo
    Yun, Seungman
    Kim, Ho Kyung
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [43] Characterization of a full field digital mammography detector based on direct x-ray conversion in selenium
    Yorker, JG
    Jeromin, LS
    Lee, DLY
    Palecki, EF
    Golden, KP
    Jing, ZX
    MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 : 21 - 29
  • [44] X-RAY PROTECTION IN INDUSTRIAL FLUOROSCOPY
    HENNY, GC
    RADIOLOGY, 1946, 46 (01) : 66 - 69
  • [45] X-RAY FLUOROSCOPY FOR AIRCRAFT INSPECTION
    不详
    METAL CONSTRUCTION AND BRITISH WELDING JOURNAL, 1972, 4 (09): : 350 - &
  • [46] A carbon nanotube based x-ray detector
    Boucher, Richard A.
    Bauch, Juergen
    Wuensche, Dietmar
    Lackner, Gerhard
    Majumder, Anindya
    NANOTECHNOLOGY, 2016, 27 (47)
  • [47] Development and evaluation of a large-area selenium-based flat panel detector for real-time radiography and fluoroscopy
    Tsukamoto, A
    Yamada, S
    Tomisaki, T
    Asahina, H
    Kinno, A
    Suzuki, K
    RADIOLOGY, 1998, 209P : 358 - 358
  • [48] Performance of an energy resolving X-ray pixel detector
    Bates, R
    Derbyshire, G
    Gannon, WJF
    Iles, G
    Lowe, B
    Mathieson, K
    Passmore, MS
    Prydderch, M
    Seller, P
    Smith, K
    Thomas, SL
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 477 (1-3): : 161 - 165
  • [49] X-Ray Performance of SiC NPN Radiation Detector
    Wang, Jing
    Zhou, Leidang
    Chen, Liang
    Zhang, Silong
    Wang, Fangbao
    Fan, Tingting
    Chen, Zhuo
    Bai, Song
    Ouyang, Xiaoping
    MICROMACHINES, 2025, 16 (01)
  • [50] The design and performance of CUBES - a CubeSat X-ray detector
    Kushwah, R.
    Stana, T. A.
    Pearce, M.