Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm: fundamental characterization in water

被引:76
|
作者
Fogliata, Antonella [1 ]
Nicolini, Giorgia [1 ]
Clivio, Alessandro [1 ]
Vanetti, Eugenio [1 ]
Mancosu, Pietro [2 ]
Cozzi, Luca [1 ]
机构
[1] Oncol Inst So Switzerland, Med Phys Unit, Bellinzona, Switzerland
[2] Ist Clin Humanitas, Radio Oncol Dept, Milan, Italy
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 06期
关键词
ANISOTROPIC ANALYTICAL ALGORITHM; TREATMENT PLANNING SYSTEMS; RADIATION-THERAPY; QUALITY-ASSURANCE; PHOTON BEAMS; RADIOTHERAPY; VERIFICATION; OPTIMIZATION; PARAMETERS; ACCURACY;
D O I
10.1088/0031-9155/56/6/022
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new algorithm, Acuros (R) XB Advanced Dose Calculation, has been introduced by Varian Medical Systems in the Eclipse planning system for photon dose calculation in external radiotherapy. Acuros XB is based on the solution of the linear Boltzmann transport equation (LBTE). The LBTE describes the macroscopic behaviour of radiation particles as they travel through and interact with matter. The implementation of Acuros XB in Eclipse has not been assessed; therefore, it is necessary to perform these preclinical validation tests to determine its accuracy. This paper summarizes the results of comparisons of Acuros XB calculations against measurements and calculations performed with a previously validated dose calculation algorithm, the Anisotropic Analytical Algorithm (AAA). The tasks addressed in this paper are limited to the fundamental characterization of Acuros XB in water for simple geometries. Validation was carried out for four different beams: 6 and 15 MV beams from a Varian Clinac 2100 iX, and 6 and 10 MV 'flattening filter free' (FFF) beams from a TrueBeam linear accelerator. The TrueBeam FFF are new beams recently introduced in clinical practice on general purpose linear accelerators and have not been previously reported on. Results indicate that Acuros XB accurately reproduces measured and calculated (with AAA) data and only small deviations were observed for all the investigated quantities. In general, the overall degree of accuracy for Acuros XB in simple geometries can be stated to be within 1% for open beams and within 2% for mechanical wedges. The basic validation of the Acuros XB algorithm was therefore considered satisfactory for both conventional photon beams as well as for FFF beams of new generation linacs such as the Varian TrueBeam.
引用
收藏
页码:1879 / 1904
页数:26
相关论文
共 50 条
  • [31] Dosimetric Verification of Acuros XB Dose Calculation Algorithm On An Air Cavity for 6-MV Flattening Filter-Free Beam.
    Kang, S.
    Chung, J.
    Suh, T.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3411 - 3411
  • [32] Validation of the preconfigured Varian Ethos Acuros XB Beam Model for treatment planning dose calculations: A dosimetric study
    Hu, Yunfei
    Byrne, Mikel
    Archibald-Heeren, Ben
    Collett, Nick
    Liu, Guilin
    Aland, Trent
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (12): : 27 - 42
  • [33] Dosimetric accuracy and clinical quality of Acuros XB and AAA dose calculation algorithm for stereotactic and conventional lung volumetric modulated arc therapy plans
    Kroon, Petra S.
    Hol, Sandra
    Essers, Marion
    [J]. RADIATION ONCOLOGY, 2013, 8
  • [34] Dosimetric comparison of analytical anisotropic algorithm and the two dose reporting modes of Acuros XB dose calculation algorithm in volumetric modulated arc therapy of carcinoma lung and carcinoma prostate
    Gopalakrishnan, Zhenia
    Bhasi, Saju
    Raghukumar, P.
    V. Menon, Sharika
    Sarin, B.
    Thayil, Anna George
    Nair, Raghuram K.
    [J]. MEDICAL DOSIMETRY, 2022, 47 (03) : 280 - 287
  • [35] Dosimetric Comparison of Acuros XB and Anisotropic Analytic Algorithm with Commercial Monte Carlo Based Dose Calculation Algorithm for Stereotactic Body Radiation Therapy of Lung Cancer
    Cao, M.
    Tenn, S.
    Lee, C.
    Yang, Y.
    Lamb, J.
    Agazaryan, N.
    Lee, P.
    Low, D.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 337 - 337
  • [36] From AAA to Acuros XB-clinical implications of selecting either Acuros XB dose-to-water or dose-to-medium
    Jackson M. Zifodya
    Cameron H. C. Challens
    Wen-Long Hsieh
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2016, 39 : 431 - 439
  • [37] Dosimetric comparison of AAA and Acuros XB algorithm for breast DIBH VMAT treatment
    Deshpande, S.
    Naidu, S.
    Chavan, K. N.
    Kannan, V.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S741 - S742
  • [38] Dosimetric comparison of peripheral NSCLC SBRT using Acuros XB and AAA calculation algorithms
    Ong, Chloe C. H.
    Ang, Khong Wei
    Soh, Roger C. X.
    Tin, Kah Ming
    Yap, Jerome H. H.
    Lee, James C. L.
    Bragg, Christopher M.
    [J]. MEDICAL DOSIMETRY, 2017, 42 (03) : 216 - 222
  • [39] From AAA to Acuros XB-clinical implications of selecting either Acuros XB dose-to-water or dose-to-medium
    Zifodya, Jackson M.
    Challens, Cameron H. C.
    Hsieh, Wen-Long
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (02) : 431 - 439
  • [40] Dosimetric evaluation of Acuros XB dose calculation algorithm with measurements in predicting doses beyond different air gap thickness for smaller and larger field sizes
    Rana, Suresh
    Rogers, Kevin
    [J]. JOURNAL OF MEDICAL PHYSICS, 2013, 38 (01) : 9 - 14