Dosimetric validation of the Acuros XB Advanced Dose Calculation algorithm: fundamental characterization in water

被引:76
|
作者
Fogliata, Antonella [1 ]
Nicolini, Giorgia [1 ]
Clivio, Alessandro [1 ]
Vanetti, Eugenio [1 ]
Mancosu, Pietro [2 ]
Cozzi, Luca [1 ]
机构
[1] Oncol Inst So Switzerland, Med Phys Unit, Bellinzona, Switzerland
[2] Ist Clin Humanitas, Radio Oncol Dept, Milan, Italy
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 06期
关键词
ANISOTROPIC ANALYTICAL ALGORITHM; TREATMENT PLANNING SYSTEMS; RADIATION-THERAPY; QUALITY-ASSURANCE; PHOTON BEAMS; RADIOTHERAPY; VERIFICATION; OPTIMIZATION; PARAMETERS; ACCURACY;
D O I
10.1088/0031-9155/56/6/022
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new algorithm, Acuros (R) XB Advanced Dose Calculation, has been introduced by Varian Medical Systems in the Eclipse planning system for photon dose calculation in external radiotherapy. Acuros XB is based on the solution of the linear Boltzmann transport equation (LBTE). The LBTE describes the macroscopic behaviour of radiation particles as they travel through and interact with matter. The implementation of Acuros XB in Eclipse has not been assessed; therefore, it is necessary to perform these preclinical validation tests to determine its accuracy. This paper summarizes the results of comparisons of Acuros XB calculations against measurements and calculations performed with a previously validated dose calculation algorithm, the Anisotropic Analytical Algorithm (AAA). The tasks addressed in this paper are limited to the fundamental characterization of Acuros XB in water for simple geometries. Validation was carried out for four different beams: 6 and 15 MV beams from a Varian Clinac 2100 iX, and 6 and 10 MV 'flattening filter free' (FFF) beams from a TrueBeam linear accelerator. The TrueBeam FFF are new beams recently introduced in clinical practice on general purpose linear accelerators and have not been previously reported on. Results indicate that Acuros XB accurately reproduces measured and calculated (with AAA) data and only small deviations were observed for all the investigated quantities. In general, the overall degree of accuracy for Acuros XB in simple geometries can be stated to be within 1% for open beams and within 2% for mechanical wedges. The basic validation of the Acuros XB algorithm was therefore considered satisfactory for both conventional photon beams as well as for FFF beams of new generation linacs such as the Varian TrueBeam.
引用
收藏
页码:1879 / 1904
页数:26
相关论文
共 50 条
  • [41] COMPARISON OF ACUROS XB AND AAA DOSE CALCULATIONS FOR ADVANCED LUNG TREATMENTS
    Vanetti, E.
    Fogliata, A.
    Clivio, A.
    Nicolini, G.
    Cozzi, L.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (08) : 516 - 516
  • [42] Critical Appraisal of Acuros XB and Anisotropic Analytic Algorithm Dose Calculation in Advanced Non-Small-Cell Lung Cancer Treatments
    Fogliata, Antonella
    Nicolini, Giorgia
    Clivio, Alessandro
    Vanetti, Eugenio
    Cozzi, Luca
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (05): : 1587 - 1595
  • [43] Implementation and Validation of 4D Acuros XB Dose Calculations
    Thomas, S.
    Huang, V.
    Milette, M.
    Yuen, C.
    Teke, T.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3441 - 3441
  • [44] Validation of a direct method to convert Acuros XB dose-to-medium to dose-to-water in lung SBRT
    Munoz-Montplet, C.
    Fuentes-Raspall, R.
    Jurado-Bruggeman, D.
    Agramunt-Chaler, S.
    Onses, A.
    Vilanova, J. C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1558 - S1559
  • [45] Clinical Impact and Utility of Acuros XB Dose Calculation Algorithm in Lung Stereotactic Body Radiation Therapy
    Clayton, R.
    Liu, H.
    Lau, H.
    Dunscombe, P.
    Nugent, Z.
    Khan, R.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S723 - S724
  • [46] Dosimetric comparison of the two dose reporting modes of Acuros XB and AAA for lung SBRT
    Mampuya, A. W.
    Nakamura, M.
    Hirose, Y.
    Ishigaki, T.
    Mizowaki, T.
    Hiraoka, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S718 - S718
  • [47] Investigation of central electrode artefacts of ionisation chamber effect on dose calculation using advanced calculation algorithms AAA and Acuros XB
    Shaiju, V. S.
    Kumar, Rajesh
    Rajasekhar, K. V.
    Zacharia, George
    Phani, Debjani
    Bhasi, Saju
    Nair, Raghuram K.
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (03) : 351 - 356
  • [48] Evaluation of Dose to Medium Vs. Dose to Water Calcualtion Mode in Acuros XB Algorithm for Lung SBRT
    Pokharel, S.
    Oyewale, S.
    Rana, S.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3422 - 3422
  • [49] Dosimetric and Radiobiological Variations in Different Dose Calculation Grid Sizes Between Acuros XB and AAA Algorithms for Prostate VMAT Plans
    Kim, K.
    Chung, J.
    Kang, S.
    Kang, S.
    Kim, J.
    Eom, K.
    Song, C.
    Suh, T.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E301 - E301
  • [50] End-to-End Verification and Dosimetric Comparative Study of Brainlab Monte Carlo and Eclipse Acuros XB Photon Dose Calculation Algorithm for VMAT SRS Planning
    Liu, H.
    Polizzi, M.
    Yuan, L.
    Kim, S.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E689 - E690