Dosimetric evaluation of Acuros XB dose calculation algorithm with measurements in predicting doses beyond different air gap thickness for smaller and larger field sizes

被引:46
|
作者
Rana, Suresh [1 ]
Rogers, Kevin [1 ]
机构
[1] Arizona Ctr Canc Care, Dept Radiat Oncol, Peoria, AZ USA
关键词
Anisotropic analytical algorithm; Acuros; air gap; dose prediction; heterogeneity correction;
D O I
10.4103/0971-6203.106600
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this study, dose prediction accuracy of Acuros XB (AXB) dose calculation algorithm beyond air gap thickness (range 2, 4, and 6 cm) in simple inhomogeneous phantoms was investigated. The evaluation of AXB was performed by comparing the doses calculated by AXB with the doses calculated by Anisotropic Analytical Algorithm (AAA) and the measured data for different field sizes (3 x 3, 5 x 5, and 10 x 10 cm(2)) of a 6 MV photon beam. The dose computation was performed within Eclipse treatment planning system, and measurements were acquired with a cylindrical ionization chamber. Central axis depth dose comparisons were done in solid-water material region up to 5 cm distance from air/solid-water interface. The results of AXB had better agreement with measurements at all measured points than that of AAA. The discrepancies between AXB and measured data were seen from -3.81% to +0.9%, whereas the AAA differences with measurement from -3.1% to -10.9%. The combination of the smallest test field size and the largest air gap produced the highest range (1-5 cm distance from air/solid-water interface) in dose difference (AAA: -4.0% to -10.6% and AXB: -3.8% to +0.6%). The AAA computational time was about 8 times faster than that of AXB. In conclusion, AXB is more appropriate to use for dose predictions, especially when low-density heterogeneities are involved.
引用
收藏
页码:9 / 14
页数:6
相关论文
共 8 条
  • [1] Dosimetric evaluation of Acuros XB Advanced Dose Calculation algorithm in heterogeneous media
    Antonella Fogliata
    Giorgia Nicolini
    Alessandro Clivio
    Eugenio Vanetti
    Luca Cozzi
    [J]. Radiation Oncology, 6
  • [2] Dosimetric evaluation of Acuros XB Advanced Dose Calculation algorithm in heterogeneous media
    Fogliata, Antonella
    Nicolini, Giorgia
    Clivio, Alessandro
    Vanetti, Eugenio
    Cozzi, Luca
    [J]. RADIATION ONCOLOGY, 2011, 6
  • [3] Dosimetric Evaluation of Acuros XB Advanced Dose Calculation Algorithm in Head and Neck Patients
    Faught, A.
    Wu, Q.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3406 - 3406
  • [4] Dosimetric and radiobiological comparison in different dose calculation grid sizes between Acuros XB and anisotropic analytical algorithm for prostate VMAT
    Kim, Kyeong-Hyeon
    Chung, Jin-Beom
    Suh, Tae Suk
    Kang, Sang-Won
    Kang, Seong-Hee
    Eom, Keun-Yong
    Song, Changhoon
    Kim, In-Ah
    Kim, Jae-Sung
    [J]. PLOS ONE, 2018, 13 (11):
  • [5] Dosimetric Evaluation with Heterogeneity in Acuros XB Advanced Dose Calculation Algorithm and Anisotropic Analytical Algorithm (AAA)
    Rana, S.
    Rogers, K.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3827 - 3827
  • [6] Dosimetric and Radiobiological Variations in Different Dose Calculation Grid Sizes Between Acuros XB and AAA Algorithms for Prostate VMAT Plans
    Kim, K.
    Chung, J.
    Kang, S.
    Kang, S.
    Kim, J.
    Eom, K.
    Song, C.
    Suh, T.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E301 - E301
  • [7] Dosimetric Evaluation of Collapsed Cone Convolution (CCC) Algorithm in Predicting Dose Beyond Air Gap
    Oyewale, S.
    Rana, S.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [8] Dosimetric Verification of Acuros XB Dose Calculation Algorithm On An Air Cavity for 6-MV Flattening Filter-Free Beam.
    Kang, S.
    Chung, J.
    Suh, T.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3411 - 3411