Safety and Efficacy of MRI Scans in Patients With Implanted Microvascular Coupling Devices

被引:3
|
作者
Phalke, Neelam [1 ]
Thomas, W. Walsh [2 ]
Azzi, James [2 ]
Li, Ryan J. [2 ]
Petrisor, Daniel [3 ]
Wax, Mark K. [2 ]
机构
[1] Louisiana State Univ, Dept Otolaryngol, Sch Med, New Orleans, LA USA
[2] Oregon Hlth & Sci Univ, Dept Otolaryngol, 3181 Sw Sam Jackson Pk Rd, Portland, OR 97239 USA
[3] Oregon Hlth & Sci Univ, Dept Oral Maxillofacial Surg, 3181 Sw Sam Jackson Pk Rd, Portland, OR 97239 USA
来源
LARYNGOSCOPE | 2020年 / 130卷 / 06期
关键词
MRI compatibility; microvascular reconstruction; microvascular coupler; MRI safety; head and neck reconstruction; BIOMEDICAL IMPLANTS; INFORMATION;
D O I
10.1002/lary.28275
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective To demonstrate the safety and compatibility of microvascular couplers in patients undergoing postoperative magnetic resonance imaging (MRI) scans. Study Design Retrospective case series. Methods A consecutive review of 1,252 patients undergoing free tissue transfer for head and neck reconstruction at Oregon Health and Sciences University (OHSU) between 2010 and 2017 who had microvascular coupler Synovis, Baxter Deerfield, IL (Synovis Life Technologies, Saint Paul, MN) implantation were reviewed. One hundred fifteen patients had a subsequent MRI scan, which consisted of a variety of Phillips (Amsterdam, Netherlands) MRI machines ranging from 1.0 to 3.0 Tesla (T) in magnetic strength. Results These 115 patients underwent 121 free flaps with 131 couplers (including 32 flow couplers) utilized for venous anastomoses. Couplers ranged in size from 1.5 to 4.0 mm (3.0 mm [42%] followed by 3.5 mm [21%], 2.5 mm [19%], 2.0 mm [10%], 4.0 mm [6%], and 1.5 mm [2%]. Three hundred fifty-nine MRI scans (2 days to 91 months postoperatively) were obtained with 233 MRIs for cancer surveillance, and the remaining were obtained for neurologic disease, injury, or evaluation for metastases. No complications occurred related to the MRI and the metallic components of the coupler or other metal implants, such as reconstruction bars, vascular clips, or metallic surgical mesh. Additionally, no radiology report commented on MRI distortion due to the coupler placement, which contrasts the distortion seen with the other metallic implants. Conclusion Microvascular couplers and their constitutive stainless-steel pins have not been found to cause any complications in a large series of consecutive patients undergoing multiple MRIs with magnetic strength up to 3 T. The U.S. Food and Drug Administration advocates medical alert notification for patients with couplers; however, hesitation regarding potential MRI scanning for surveillance or otherwise is unwarranted. Level of Evidence 4 Laryngoscope, 130:1428-1430, 2020
引用
收藏
页码:1428 / 1430
页数:3
相关论文
共 50 条
  • [41] Update on MRI Safety in Patients with Cardiac Implantable Electronic Devices
    Markman, Timothy M.
    Halperin, Henry R.
    Nazarian, Saman
    RADIOLOGY, 2018, 288 (03) : 656 - 657
  • [42] Safety of intrauterine devices in MRI
    Bussmann, Simon
    Luechinger, Roger
    Froehlich, Johannes M.
    von Weymarn, Constantin
    Reischauer, Carolin
    Koh, Dow Mu
    Gutzeit, Andreas
    PLOS ONE, 2018, 13 (10):
  • [43] Efficacy and safety of new-generation atrial antitachycardia pacing for atrial tachyarrhythmias in patients implanted with cardiac resynchronization therapy devices
    Ueda, Nobuhiko
    Kamakura, Tsukasa
    Noda, Takashi
    Nakajima, Kenzaburo
    Kataoka, Naoya
    Wada, Mitsuru
    Yamagata, Kenichiro
    Ishibashi, Kohei
    Inoue, Yuko
    Miyamoto, Koji
    Nagase, Satoshi
    Aiba, Takeshi
    Izumi, Chisato
    Noguchi, Teruo
    Yasuda, Satoshi
    Kusano, Kengo
    JOURNAL OF CARDIOLOGY, 2020, 75 (05) : 559 - 566
  • [44] SAFETY AND EFFICACY OF VIDEO CAPSULE ENDOSCOPY IN PATIENTS IMPLANTED WITH LEFT VENTRICULAR ASSIST DEVICES PRESENTING WITH GASTROINTESTINAL BLEEDING: A SYSTEMATIC REVIEW
    Baig, Muhammad A.
    Mirwais, Maiwand
    Khan, Zubair
    Attar, Bashar M.
    Demetria, Melchor V.
    Gandhi, Seema
    GASTROINTESTINAL ENDOSCOPY, 2018, 87 (06) : AB301 - AB302
  • [45] Safety of patients fitted with electronic artificial implanted devices in ultraviolet treatment cubicles
    Turner, D.
    Britton, J.
    Tassell, N.
    Bird, D.
    Goulden, V.
    BRITISH JOURNAL OF DERMATOLOGY, 2014, 171 : 63 - 63
  • [46] Safety of enoxaparin in patients implanted with continuous flow left ventricular assist devices
    Klosko, Rachel C.
    Whitson, Bryan
    Lampert, Brent
    Mclaughlin, Eric
    Orzel, Libby
    ARTIFICIAL ORGANS, 2024,
  • [47] Assessment of the Safety Risk of Dermatoscope Magnets in Patients With Cardiovascular Implanted Electronic Devices
    Rishpon, Ayelet
    Braun, Ralph
    Weinstock, Martin A.
    Kulju, Stephen
    Grenga, Andrea
    Navarrete-Dechent, Cristian
    Marghoob, Nadeem G.
    Steffel, Jan
    Marghoob, Ashfaq A.
    JAMA DERMATOLOGY, 2018, 154 (10) : 1204 - 1207
  • [48] FLUOROSCOPY-FREE ABLATION IN PATIENTS WITH IMPLANTED CARDIAC DEVICES: SAFETY AND EFFICIENCY
    Furlan, Tiffany
    Adams, John
    Patel, Uday
    Fudyma, Iga
    Beri, Abhimanyu
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2024, 83 (13) : 229 - 229
  • [49] MR Safety Active Implanted Electronic Devices
    Watson, Robert E., Jr.
    Edmonson, Heidi A.
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2020, 28 (04) : 549 - +
  • [50] MRI and Patients with Non-MRI-conditional Cardiac Devices: Further Evidence of Safety
    Shellock, Frank G.
    RADIOLOGY, 2020, 295 (02) : 314 - 315