Experimental assessment of microwave ablation computational modeling with MR thermometry

被引:18
|
作者
Faridi, Pegah [1 ]
Keselman, Paul [2 ]
Fallahi, Hojjatollah [1 ]
Prakash, Punit [1 ]
机构
[1] Kansas State Univ, Mike Wiegers Dept Elect & Comp Engn, Manhattan, KS 66506 USA
[2] Univ Kansas, Hoglund Brain Imaging Ctr, Med Ctr, Kansas City, KS 66160 USA
基金
美国国家卫生研究院;
关键词
computational modeling; microwave ablation; model validation; MR thermometry; MRI-guided interventions; MAGNETIC-RESONANCE THERMOMETRY; VIVO BOVINE LIVER; EX-VIVO; THERMAL ABLATION; DIELECTRIC-PROPERTIES; TEMPERATURE-DEPENDENCE; RABBIT BRAIN; DESIGN; DEVICE; CONDUCTIVITY;
D O I
10.1002/mp.14318
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Computational models are widely used during the design and characterization of microwave ablation (MWA) devices, and have been proposed for pretreatment planning. Our objective was to assess three-dimensional (3D) transient temperature and ablation profiles predicted by MWA computational models with temperature profiles measured experimentally using magnetic resonance (MR) thermometry inex vivobovine liver. Materials and methods We performed MWA inex vivotissue under MR guidance using a custom, 2.45 GHz water-cooled applicator. MR thermometry data were acquired for 2 min prior to heating, during 5-10 min microwave exposures, and for 3 min following heating. Fiber-optic temperature sensors were used to validate the accuracy of MR temperature measurements. A total of 13 ablation experiments were conducted using 30-50 W applied power at the applicator input. MWA computational models were implemented using the finite element method, and incorporated temperature-dependent changes in tissue physical properties. Model-predicted ablation zone extents were compared against MRI-derived Arrhenius thermal damage maps using the Dice similarity coefficient (DSC). Results Prior to heating, the observed standard deviation of MR temperature data was in the range of 0.3-0.7 degrees C. Mean absolute error between MR temperature measurements and fiber-optic temperature probes during heating was in the range of 0.5-2.8 degrees C. The mean DSC between model-predicted ablation zones and MRI-derived Arrhenius thermal damage maps for 13 experimental set-ups was 0.95. When comparing simulated and experimentally (i.e. using MRI) measured temperatures, the mean absolute error (MAE %) relative to maximum temperature change was in the range 5%-8.5%. Conclusion We developed a system for characterizing 3D transient temperature and ablation profiles with MR thermometry during MWA inex vivoliver tissue, and applied the system for experimental validation of MWA computational models.
引用
收藏
页码:3777 / 3788
页数:12
相关论文
共 50 条
  • [21] In vitro artefact assessment of a new MR-compatible microwave antenna and a standard MR-compatible radiofrequency ablation electrode for tumour ablation
    Rüdiger Hoffmann
    Hansjörg Rempp
    Frank Eibofner
    David-Emanuel Keßler
    Gunnar Blumenstock
    Jakob Weiß
    Philippe L. Pereira
    Konstantin Nikolaou
    Stephan Clasen
    European Radiology, 2016, 26 : 771 - 779
  • [22] In vitro artefact assessment of a new MR-compatible microwave antenna and a standard MR-compatible radiofrequency ablation electrode for tumour ablation
    Hoffmann, Ruediger
    Rempp, Hansjoerg
    Eibofner, Frank
    Kessler, David-Emanuel
    Blumenstock, Gunnar
    Weiss, Jakob
    Pereira, Philippe L.
    Nikolaou, Konstantin
    Clasen, Stephan
    EUROPEAN RADIOLOGY, 2016, 26 (03) : 771 - 779
  • [23] MR-compatible RF ablation system for online treatment monitoring using MR thermometry
    Will, Karl
    Krug, Johannes
    Jungnickel, Kerstin
    Fischbach, Frank
    Ricke, Jens
    Rose, Georg
    Omar, Abbas
    2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 1601 - 1604
  • [24] Adjuvant Thermal Accelerant Gel Use Increases Microwave Ablation Zone updates Temperature in Porcine Liver as Measured by MR Thermometry
    Maxwell, Aaron W. P.
    Park, William K. C.
    Baird, Grayson L.
    Walsh, Edward G.
    Dupuy, Damian E.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2020, 31 (08) : 1357 - 1364
  • [25] MR-guided microwave ablation for malignancies
    Yoshimasa Kurumi
    Tohru Tani
    Shigeyuki Naka
    Hisanori Shiomi
    Tomoharu Shimizu
    Hajime Abe
    Yoshihiro Endo
    Shigehiro Morikawa
    International Journal of Clinical Oncology, 2007, 12 : 85 - 93
  • [26] MR-guided microwave ablation for malignancies
    Kurumi, Yoshimasa
    Tani, Tohru
    Naka, Shigeyuki
    Shiomi, Hisanori
    Shimizu, Tomoharu
    Abe, Hajime
    Endo, Yoshihiro
    Morikawa, Shigehiro
    INTERNATIONAL JOURNAL OF CLINICAL ONCOLOGY, 2007, 12 (02) : 85 - 93
  • [27] Monitoring Thermal Ablation via Microwave Tomography: An Ex Vivo Experimental Assessment
    Scapaticci, Rosa
    Lopresto, Vanni
    Pinto, Rosanna
    Cavagnaro, Marta
    Crocco, Lorenzo
    DIAGNOSTICS, 2018, 8 (04):
  • [28] A Minimally Invasive Antenna for Microwave Ablation Therapies: Design, Performances, and Experimental Assessment
    Cavagnaro, Marta
    Amabile, Claudio
    Bernardi, Paolo
    Pisa, Stefano
    Tosoratti, Nevio
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (04) : 949 - 959
  • [29] Feasibility assessment and analysis of thermal sensitivity of CT-thermometry during microwave ablation of ex vivo porcine kidneys
    Schena, Emiliano
    Massaroni, Carlo
    Giurazza, Francesco
    Park, John
    Park, Jinha
    Fong, Yuman
    Saccomandi, Paola
    2016 NANOTECHNOLOGY FOR INSTRUMENTATION AND MEASUREMENT (NANOFIM), 2016,
  • [30] Non-invasive thermometry for the thermal ablation of liver tumor: A computational methodology
    D'Ambrosio, V.
    Di Barba, P.
    Dughiero, F.
    Mognaschi, M. E.
    Savini, A.
    INTERNATIONAL JOURNAL OF APPLIED ELECTROMAGNETICS AND MECHANICS, 2007, 25 (1-4) : 407 - 412