Calibration of a portal imaging device for high-precision dosimetry: A Monte Carlo study

被引:23
|
作者
Keller, H [1 ]
Fix, M [1 ]
Ruegsegger, P [1 ]
机构
[1] ETH Zurich, Inst Biomed Engn, Zurich, Switzerland
关键词
electronic portal imaging; liquid filled ionization chambers; dose calibration; Monte Carlo simulation;
D O I
10.1118/1.598378
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Today electronic portal imaging devices (EPID's) are used primarily to verify patient positioning. They have, however, also the potential as 2D-dosimeters and could be used as such for transit dosimetry or dose reconstruction. It has been proven that such devices, especially liquid filled ionization chambers, have a stable dose response relationship which can be described in terms of the physical properties of the EPID and the pulsed linac radiation. For absolute dosimetry however, an accurate method of calibration to an absolute dose is needed. In this work, we concentrate on calibration against dose in a homogeneous water phantom. Using a Monte Carlo model of the detector we calculated dose spread kernels in units of absolute dose per incident energy fluence and compared them to calculated dose spread kernels in water at different depths. The energy of the incident pencil beams varied between 0.5 and 18 MeV. At the depth of dose maximum in water for a 6 MV beam (1.5 cm) and for a Is MV beam (3.0 cm) we observed large absolute differences between water and detector dose above an incident energy of 4 MeV but only small relative differences in the most frequent energy range of the beam energy spectra. It is shown that for a 6 MV beam the absolute reference dose measured at 1.5 cm water depth differs from the absolute detector dose by 3.8%. At depth 1.2 cm in water, however, the relative dose differences are almost constant between 2 and 6 MeV. The effects of changes in the energy spectrum of the beam on the dose responses in water and in the detector are also investigated. We show that differences larger than 2% can occur for different beam qualities of the incident photon beam behind water slabs of different thicknesses. It is therefore concluded that for high-precision dosimetry such effects have to be taken into account. Nevertheless, the precise information about the dose response of the detector provided in this Monte Carlo study forms the basis of extracting directly the basic radiometric quantities photon fluence and photon energy fluence from the detector's signal using a deconvolution algorithm. The results are therefore promising for future application in absolute transit dosimetry and absolute dose reconstruction. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)02210-X].
引用
收藏
页码:1891 / 1902
页数:12
相关论文
共 50 条
  • [21] ROLE OF ELECTRONIC PORTAL IMAGING IN HIGH-DOSE HIGH-PRECISION RADIOTHERAPY
    VANHERK, M
    GILHUIJS, K
    WILLIAMS, P
    CINQUIN, P
    CIONINI, L
    SWINDELL, W
    RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) : 269 - 270
  • [22] Characterization of a high-elbow, fluoroscopic electronic portal imaging device for portal dosimetry
    de Boer, JCJ
    Heijmen, BJM
    Pasma, KL
    Visser, AG
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (01): : 197 - 216
  • [23] Determination of electron backscattering coefficient of beryllium by a high-precision Monte Carlo simulation
    Hussain, Abrar
    Yang, Lihao
    Mao, Shifeng
    Da, Bo
    Tokesi, Karoly
    Ding, Z. J.
    NUCLEAR MATERIALS AND ENERGY, 2021, 26
  • [24] Monte Carlo simulation of the iView GT portal imager dosimetry
    Juste, B.
    Miro, R.
    Diez, S.
    Campayo, J. M.
    Verdu, G.
    APPLIED RADIATION AND ISOTOPES, 2010, 68 (4-5) : 922 - 925
  • [25] Fast transit portal dosimetry using density-scaled layer modeling of aSi-based electronic portal imaging device and Monte Carlo method
    Jung, Jae Won
    Kim, Jong Oh
    Yeo, Inhwan Jason
    Cho, Young-Bin
    Kim, Sun Mo
    DiBiase, Steven
    MEDICAL PHYSICS, 2012, 39 (12) : 7593 - 7602
  • [26] Dosimetry with a fluoroscopic electronic portal imaging device
    Pasma, KL
    MEDICAL PHYSICS, 2000, 27 (02) : 423 - 423
  • [27] An electronic portal imaging device for transit dosimetry
    SymondsTayler, JRN
    Partridge, M
    Evans, PM
    PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11): : 2273 - 2283
  • [28] Modeling Electronic Portal Imaging Device for Dose Reconstruction using Monte Carlo Method
    Jung, J. W.
    Kim, J. O.
    Yeo, I.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S724 - S724
  • [29] Research on high-precision calibration device for dual-axis tilt sensor
    Zhou, Jin
    Wu, XiaoLi
    Zheng, Jing
    Zhou, Sen
    Yang, Xuemei
    SPIE FUTURE SENSING TECHNOLOGIES 2024, 2024, 13083
  • [30] A High-Precision Calibration Method for Spectrometers
    Perret, Edith
    Balmer, Tobias E.
    AMERICAN LABORATORY, 2011, 43 (05) : 44 - 45